DJO SURGICAL S-201106Recalls & reported problems
Encore Medical, L.P. · Special, Shortened RSP Baseplate Drill_Screw Guide · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00190446266068), received between Sep 15, 2021 and Dec 9, 2025 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Device Handling Problem1
- Material Fragmentation1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Foreign Body In Patient1
- Insufficient Information1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryDec 9, 2025Material Fragmentation; Material Integrity Problem; Material Fragmentation; Material Integrity Problem
PATIENT PRESENTED WITH ARTHRITIS OF THE GLENOHUMERAL JOINT, IT WAS DECIDED TO INITIATE SURGERY FOR A REVERSE TOTAL SHOULDER REPLACEMENT. DURING SURGERY, THE BASEPLATE WAS IMPLANTED AND THE SHORT 4 BARREL DRILL GUIDE WAS PLACED ONTO THE BASEPLATE DURING INSERTION OF THAT INSTRUMENT (8-201106), A PORTION OF CENTRAL SCREW OF THE DRILL GUIDE BROKE OFF IN THE BASEPLATE. AFTER SEVERAL ATTEMPTS TO REMOVE THE SCREW AND THE BASEPLATE (ABOUT 25-30 MINUTES) IT WAS DETERMINED TO RELY ON THE MORSE TAPERS ON THE GLENOSPHERE AND FORGO THE RETAINING SCREW.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 23763846 ↗MalfunctionSep 12, 2022Device Handling Problem; Device Handling Problem
INSTRUMENT FAILURE - DUE TO THE SURGEON APPLIED TOO MUCH FORCE AND THE SCREW BROKE OFF. THE BROKEN PIECE TO THE SCREW WAS LEFT IN THE BASEPLATE TO PREVENT FURTHER COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15403284 ↗MalfunctionApr 13, 2022Break
INSTRUMENT FAILURE - SURGEON WAS ATTACHING THE TWO PIECE DRILL GUIDE TO THE BASEPLATE AND WAS NOT ABLE TO BECAUSE THE TIP OF THE SCREW BROKE OFF IN THE BASEPLATE. WHEN HE WAS PUTTING ON THE GLENOSPHERE HE WAS UNABLE TO PUT IN THE SET SCREW BECAUSE OF THE METAL PIECE. EVERYTHING FINISHED WITH NO ISSUE.
Patient problems: Insufficient Information
MAUDE report 14105064 ↗MalfunctionSep 15, 2021Break
PRIMARY SURGERY COMPLAINT - A HEALTH PROFESSIONAL REPORTED THAT THE CENTRAL SCREW IN THE 4-HOLE DRILL GUIDE BROKE INSIDE THE BASEPLATE MAKING IT IMPOSSIBLE TO DRILL OR INTRODUCE THE RETAINING SCREW TO SECURE THE GLENOID HEAD. THE SURGERY WAS COMPLETED WITH ANOTHER DEVICE AFTER A 10 MINUTE DELAY. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12476392 ↗
Other Guide, surgical, instrument devices
- NA 03.037.017Synthes GmbH · Guide, surgical, instrument346 reports
- EVOS 71174934Smith & Nephew, Inc. · Guide, surgical, instrument149 reports
- LCP 323.042Synthes GmbH · Guide, surgical, instrument129 reports
- DJO SURGICAL 342-10-705Encore Medical, L.P. · Guide, Surgical, Instrument1 recall51 reports
- DJO SURGICAL 341-10-707Encore Medical, L.P. · Guide, Surgical, Instrument81 reports
- NA 03.010.404Synthes GmbH · Guide, surgical, instrument73 reports
- DJO SURGICAL 341-14-708Encore Medical, L.P. · Guide, Surgical, Instrument1 recall22 reports
- DJO SURGICAL 341-10-711Encore Medical, L.P. · Guide, Surgical, Instrument1 recall16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page