Skip to content

35X-NM ProFlexx® 0015788Recalls & reported problems

Ferno-Washington, Inc. · 35X-NM ProFlexx®, Non Magnetic · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00190790000349), received between Dec 28, 2016 and Jul 18, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Unintended Movement” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unintended Movement1
  • Use of Device Problem1

Patient problems

  • Abrasion1
  • Injury1
  • Pain1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 18, 2025Unintended Movement; Unintended Movement

    A PRODUCT ISSUE ON THE SUBJECT STRETCHER WAS REPORTED VIA MEDSUN REPORT #(B)(4) ALLEGING A WHEEL LOCK ISSUE BUT NO SPECIFIC PATIENT INCIDENT OR INJURY WAS ASSOCIATED WITH THE ISSUE. UPON FURTHER INVESTIGATION WITH THE END USER, IT WAS DISCOVERED THE EVENT ASSOCIATED WITH THE SUBJECT STRETCHER INVOLVED RESTRAINTS AND NOT A WHEEL LOCK ISSUE. IT WAS REPORTED WHILE TRANSPORTING A PATIENT ON THE STRETCHER TO AN MRI TRAILER, THE PATIENT'S WEIGHT SHIFTED WHILE TRAVERSING A THRESHOLD AND THE STRETCHER TIPPED OVER. REPORTED PATIENT INJURY INCLUDED ABRASIONS TO NOSE, ARMS, KNEES AND TOES. THE PATIENT WAS NOT RESTRAINED TO THE STRETCHER AT THE TIME OF INCIDENT NOR WERE THE PROVIDED RESTRAINTS INSTALLED ON THE STRETCHER AT THE TIME OF INCIDENT. THERE WAS NO ALLEGATION OF PRODUCT MALFUNCTION CONTRIBUTING TO THE INCIDENT AND THERE HAS BEEN NO ADDITIONAL INFORMATION PROVIDED ON THE PATIENT'S INJURIES. THE END USER CONDUCTED AN INTERNAL INVESTIGATION WHICH INCLUDED A CORRECTIVE ACTION TO INSTALL THE PROVIDED RESTRAINTS ONTO THE STRETCHER BEFORE PLACING IT BACK IN SERVICE.

    Patient problems: Abrasion

    MAUDE report 22542838 ↗
  • Dec 28, 2016Use of Device Problem; Use of Device Problem

    IT WAS REPORTED WHILE LIFTING THE FRAME OF THE COT TO RELEASE THE LOCKING MECHANISM, A MEDIC ALLEGEDLY INJURED THEIR BACK WHEN THE MECHANISM WOULD NOT RELEASE. THERE WAS A PATIENT ON THE COT AT THE TIME OF INCIDENT BUT THERE WERE NO REPORTS OF INJURY OR ADVERSE CONSEQUENCE TO THE PATIENT. THE INITIAL REPORT OF THE INCIDENT INDICATED THE MEDIC DID NOT SEEK MEDICAL INTERVENTION FOR THE ALLEGED INJURY BUT RATHER USED ICE AND REST. IT WAS LATER REPORTED ON 12/23/2016 THAT THE MEDIC WAS GOING TO SEEK MEDICAL TREATMENT FOR THE ALLEGED INJURY DUE TO ONGOING PAIN.

    Patient problems: Injury; Pain

    MAUDE report 6208287 ↗

Other Stretcher, Wheeled devices

All Stretcher, Wheeled devices →

Sources and updates