Power-PRO 2, High Configuration 650705550001Recalls & reported problems
Stryker Corporation · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Stryker Power-PRO 2 is a wheeled stretcher. Its 99 adverse event reports include 2 deaths and 97 malfunctions. The reports mostly describe the cot tipping or dropping during loading, and difficulty loading it or releasing it from ambulance fasteners. 2 open recalls name this device: one for a bent base leg, the other for weak epoxy that may let the cot separate during transport.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 99
- Deaths reported
- 2
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 99 adverse event reports that list this exact device (DI 07613327559118), received between Jul 26, 2022 and Apr 22, 2026 — classified as 2 deaths, 97 malfunctions.
The most frequently reported specific device problem is “Positioning Failure” (25 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 24, 2023Class IIOpen, Classified
Base leg assembly may bend and result in difficulty engaging/disengaging from fastener, difficulty loading/unloading from ambulance, difficulty raising or lowering the cot, cot tip, or erratic or unintended motion.
Z-2130-2023 · Root cause: Process control · Recall details · FDA record ↗
- Feb 1, 2023Class IIOpen, Classified
Transport Cot has insufficient volume of epoxy to assemblies installed in the head section and foot sectioning system may result in separation of head end or foot end from the cot and result in unexpected disengagement while loading/unloading or transport.
Z-1256-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Positioning Failure25
- Difficult or Delayed Positioning24
- Difficult to Remove18
- Device Fell15
- Device Dislodged or Dislocated9
- Protective Measures Problem9
- Device Tipped Over6
- Device Slipped3
Patient problems
- No Clinical Signs, Symptoms or Conditions88
- Insufficient Information9
- Pain5
- Discomfort2
- Laceration(s)2
- Cardiac Arrest1
- Fall1
- Head Injury1
Recent adverse event reports
The 25 most recent of 99 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 22, 2026Difficult or Delayed Positioning; Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED COT TIP OR DROP (DURING LOAD/UNLOAD) AND DIFFICULT TO LOAD/UNLOAD THE COT IN THE AMBULANCE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24958484 ↗MalfunctionApr 22, 2026Protective Measures Problem
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED INADEQUATE PATIENT RESTRAINT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24958481 ↗MalfunctionApr 22, 2026Difficult to Remove
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 7 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED INABILITY TO DISENGAGE THE COT FROM THE COT FASTENER. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24958479 ↗MalfunctionApr 22, 2026Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED COT TIP OR DROP (DURING LOAD/UNLOAD). NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24958404 ↗MalfunctionApr 22, 2026Difficult or Delayed Positioning; Device Dislodged or Dislocated
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED DIFFICULT TO LOAD/UNLOAD THE COT IN THE AMBULANCE AND COT FALSELY ENGAGES INTO FASTENER. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24958382 ↗MalfunctionApr 22, 2026Device Dislodged or Dislocated
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED COT FALSELY ENGAGES INTO FASTENER. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24958374 ↗DeathFeb 19, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE DEVICE WAS UNABLE TO BE LOADED INTO THE POWER LOAD. THE DEPARTMENT REQUESTED A 2ND RESCUE TO TRANSPORT. IT WAS FURTHER REPORTED THAT THE PATIENT DIED OF A HEART ATTACK.
Patient problems: Cardiac Arrest
MAUDE report 24393943 ↗MalfunctionJan 28, 2026Difficult or Delayed Positioning
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24199653 ↗MalfunctionJan 26, 2026Protective Measures Problem
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1-DECEMBER 31, 2025. THIS REPORT SUMMARIZES 6 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED INADEQUATE PATIENT RESTRAINT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24183348 ↗MalfunctionJan 26, 2026Positioning Failure
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) -(B)(6) 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FALSE LATCH OF SIDERAIL; INADEQUATE SIDERAIL SUPPORT FORCE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24183150 ↗MalfunctionJan 26, 2026Patient Device Interaction Problem
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED PINCH/CRUSH POINT. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT; IT IS UNKNOWN AT THIS TIME WHAT THE PATIENT EXPERIENCED.
Patient problems: Discomfort; Insufficient Information; Pain
MAUDE report 24183141 ↗MalfunctionJan 26, 2026Difficult to Remove
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1-DECEMBER 31, 2025. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED INABILITY TO DISENGAGE THE COT FROM THE COT FASTENER. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24182917 ↗MalfunctionJan 26, 2026Device Tipped Over
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1-DECEMBER 31, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED COT TIP OR DROP (DURING LOAD/UNLOAD). NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24178588 ↗MalfunctionJan 26, 2026Difficult or Delayed Positioning; Difficult to Remove
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1-DECEMBER 31, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED INABILITY TO DISENGAGE THE COT FROM THE COT FASTENER AND DIFFICULT TO LOAD/UNLOAD THE COT IN THE AMBULANCE. THERE WAS A PATIENT FALL, BUT NO SERIOUS ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24178587 ↗MalfunctionJan 26, 2026Difficult or Delayed Positioning; Device Fell
THIS REPORT ORIGINALLY SUMMARIZED 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED DIFFICULT TO LOAD/UNLOAD THE COT IN THE AMBULANCE AND COT TIP OR DROP. AFTER FURTHER EVALUATION, IT WAS FOUND THAT THE DEVICE WAS NOT DIFFICULT TO LOAD/UNLOAD AND ONLY EXPERIENCED UNINTENDED COT LOWERING OR DROPPING; NO TIP (COT DROP).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24178584 ↗DeathDec 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THERE WAS AN AMBULANCE ACCIDENT. THE PATIENT BEING TRANSPORTED PASSED AWAY AS A RESULT OF INJURIES SUSTAINED IN THE CRASH. THE CORONER'S OFFICE BELIEVES THE COT RESTRAINTS MAY HAVE CONTRIBUTED TO THE PATIENT¿S INJURIES, BUT THIS HAS NOT BEEN CONFIRMED.
Patient problems: Insufficient Information
MAUDE report 23881840 ↗MalfunctionOct 24, 2025Protective Measures Problem; Protective Measures Problem
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED INADEQUATE PATIENT RESTRAINT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23384017 ↗MalfunctionOct 24, 2025Difficult to Remove; Difficult to Remove
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 3 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED INABILITY TO DISENGAGE THE COT FROM THE COT FASTENER. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23383863 ↗MalfunctionOct 24, 2025Difficult or Delayed Positioning; Positioning Failure; Difficult or Delayed Positioning; Positioning Failure
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 3 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED DIFFICULT TO LOAD/UNLOAD THE COT IN THE AMBULANCE AND FALSE ENGAGEMENT OF HANDLE IN THE UPRIGHT OR HORIZONTAL POSITION. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23383858 ↗MalfunctionOct 24, 2025Difficult or Delayed Positioning; Difficult or Delayed Positioning
THIS REPORT SUMMARIZES 44 MALFUNCTION EVENT(S), INSTEAD OF 45.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions; Pain
MAUDE report 23383853 ↗MalfunctionOct 24, 2025Device Fell; Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 3 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED COT TIP OR DROP (DURING LOAD/UNLOAD). NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23383845 ↗MalfunctionOct 24, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23383718 ↗MalfunctionJul 25, 2025Difficult or Delayed Positioning; Positioning Failure; Difficult or Delayed Positioning; Positioning Failure
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED DIFFICULT TO LOAD/UNLOAD THE COT IN THE AMBULANCE AND FALSE ENGAGEMENT OF HANDLE IN THE UPRIGHT OR HORIZONTAL POSITION. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22620737 ↗MalfunctionJul 25, 2025Difficult to Remove; Difficult to Remove
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED INABILITY TO DISENGAGE THE COT FROM THE COT FASTENER. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22608307 ↗MalfunctionJul 25, 2025Positioning Failure; Positioning Failure
THIS REPORT SUMMARIZES 125 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FALSE ENGAGEMENT OF HANDLE IN THE UPRIGHT OR HORIZONTAL POSITION. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT; THE PATIENT EXPERIENCED DISCOMFORT.
Patient problems: Discomfort; Muscle/Tendon Damage; No Clinical Signs, Symptoms or Conditions
MAUDE report 22608285 ↗
Other STRETCHER, WHEELED devices
- Power-PRO 6506Stryker Corporation · STRETCHER, WHEELED5 deaths472 reports
- Procedural stretcher P8000HHill-Rom de México, S. de R.L. de C.V. · STRETCHER, WHEELED361 reports
- Prime Series 1115Stryker Corporation · STRETCHER, WHEELED1 death251 reports
- Prime Series 1115Stryker Corporation · STRETCHER, WHEELED1 recall199 reports
- NA 1089Stryker Corporation · STRETCHER, WHEELED175 reports
- NA 5050Stryker Corporation · STRETCHER, WHEELED1 recall99 reports
- Power-PRO 2 650700000000Stryker Corporation · STRETCHER, WHEELED2 recalls1 death42 reports
- NA 0747Stryker Corporation · STRETCHER, WHEELED1 recall87 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page