POWER F2 0567093Recalls & reported problems

Ferno-Washington, Inc. · POWER F2 W/PWR DRIVE · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
1
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00190790006570), received between Mar 25, 2026 and Apr 29, 2026 — classified as 1 death, 2 malfunctions.

The most frequently reported specific device problem is “Communication or Transmission Problem” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 19, 2026Class IIOpen, Classified

    Product may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may become lodged during operation and movement of the stretcher along the rail may no longer be possible through either powered or manual operation. May cause delays in transport of patient to healthcare facility.

    Z-2760-2026 · Root cause: Process control · Recall details · FDA record ↗

  2. Apr 28, 2026Open, Classified

    Firmware issue may result in intermittent loss or degradation of powered functionality when paired with compatible stretcher. The issue may lead to the system not operating in powered mode, or exhibiting reduced powered performance. Issue may result in interruption or delay in patient transport.

    Z-0145-2027 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Communication or Transmission Problem2
  • Intermittent Communication Failure1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2026Communication or Transmission Problem

    THE END USER REPORTED UPON ARRIVAL AT A PATIENT'S HOME, THE FASTENING SYSTEM INSTALLED IN THE AMBULANCE WOULD NOT ALLOW THE STRETCHER TO BE UNLOADED FROM THE AMBULANCE. THE END USER CONFIRMED THE FASTENING SYSTEM WAS OPERATING PROPERLY PRIOR TO ARRIVAL ON SCENE. THE END USER ADVISED THAT THE CREW ELECTED TO NOT TRANSITION TO MANUAL OPERATION OF THE SYSTEM BUT RATHER THEY CALLED FOR A BACKUP AMBULANCE TO COMPLETE THE PATIENT TRANSFER. THIS RESULTED IN A REPORTED 30 MIN DELAY. THE PATIENT'S CONDITION WAS REPORTED AS "HAVING TROUBLE BREATHING". THERE WERE NO ALLEGATIONS THAT THE DELAY RESULTED IN AN ADVERSE HEALTH CONSEQUENCE FOR THE PATIENT. A FOLLOW UP CALL TO THE END USER WAS MADE AND IT WAS DETERMINED THE FASTENING SYSTEM HAD BECOME UNPAIRED WITH THE STRETCHER WHICH RESULTED IN THE DISRUPTION OF THE COMMUNICATION NECESSARY FOR OPERATION OF THE POWERED FEATURES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25033098 ↗
  • MalfunctionApr 29, 2026Communication or Transmission Problem

    THE END USER REPORTED THE POWERED FASTENING SYSTEM WOULD NOT FULLY RETRACT THE CARRIAGE AND STRETCHER WHEN LOADING A PATIENT INTO THE AMBULANCE. THE MEDIC CREW TRANSITIONED INTO MANUAL OPERATION OF THE FASTENING SYSTEM TO COMPLETE THE TRANSFER TO THE HOSPITAL. UPON UNLOADING AT THE HOSPITAL, THE FASTENING SYSTEM WOULD THEN NOT RELEASE THE STRETCHER FROM THE FASTENER. THE PATIENT WAS LIFTED FROM THE STRETCHER AND ONTO A HOSPITAL STRETCHER THAT WAS BROUGHT OUT TO THE AMBULANCE. THERE WERE NO ALLEGATIONS OF ADVERSE HEALTH CONSEQUENCES TO THE PATIENT AS A RESULT OF THE DELAY IN LOADING AND UNLOADING. AFTER THE TRANSPORT, THE END USER WAS ABLE TO RELEASE THE STRETCHER FROM THE FASTENER AND REPAIR THE FASTENING SYSTEM AND STRETCHER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25033071 ↗
  • DeathMar 25, 2026Mechanical Problem; Intermittent Communication Failure

    THE END USER REPORTED AFTER ARRIVAL AT THE HOSPITAL WITH AN ACTIVE CARDIAC PATIENT; WHILE UNLOADING THE STRETCHER FROM THE AMBULANCE, THE LEGS OF THE STRETCHER HAD TO BE MANUALLY LOWERED BUT THEN THE CARRIAGE ARMS WOULD NOT LOWER DOWNWARD TO RELEASE THE STRETCHER FROM THE LOADER. THE CREW WAS ABLE TO MANUALLY LOWER THE CARRIAGE ARMS AND COMPLETE THE TRANSPORT INTO THE HOSPITAL. THE END USER ADVISED AN ESTIMATED TWO-MINUTE DELAY WAS EXPERIENCED DURING THE UNLOADING PROCESS. PATIENT DETAILS: MALE, APPROXIMATELY 220 LBS, PATIENT WAS EXPERIENCING AND BEING TREATED FOR CARDIAC ARREST ON SCENE AND THROUGHOUT THE TRANSPORT. CPR WAS BEING ADMINISTERED VIA LIFELINE ARM. THE PATIENT WAS LATER PRONOUNCED DECEASED AT THE HOSPITAL. THE END USER ADVISED THEY WERE UNABLE TO DUPLICATE THE ISSUE AFTER THE INCIDENT AND MADE THE DECISION TO KEEP THE POWER F2 IN SERVICE, IN MANUAL MODE ONLY, UNTIL THE EVALUATION. THIS CONTINUES TO BE AN ACTIVE INVESTIGATION AND ADDITIONAL INFORMATION WILL BE PROVIDED IN A FOLLOW UP REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24689746 ↗

Other Stretcher, Wheeled devices