Skip to content

ACS RGHN04IRecalls & reported problems

American Contract Systems, Inc. · HAND PACK - 297829 · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. May 15, 2024Open, Classified

    Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

    Z-2295-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Other Orthopedic tray devices

All Orthopedic tray devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page