HAND PACK 297829--Procedure Kit recallZ-2295-2024 · American Contract Systems Inc · initiated May 15, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from American Contract Systems Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.
FDA root cause category: Under Investigation by firm
What the company is doing
American Contract Systems issued Urgent Voluntary Recall Letter (FA-2024-018) via email on 5/15/24 to Distributors. Distributors to notify end-users. Letter states reason for recall, health risk and action to take: Immediately discontinue use of the cast padding component (DISCARD ITEM 4927009046-CAST PADDING, 6" Cotton Blend Rayon) included in each kit/tray identified in ATTACHMENT 1. Please review your inventory, and segregate and quarantine all affected Trays. If any of the affected trays identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. "Add warning labels to all kits/cases impacted: A labeling template is enclosed for printing warning labels which should be affixed to all affected product in inventory. This label should be affixed to each affected kit and case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing labeling. "Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. "Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. "Complete the enclosed Response Form and return as soon as possible. "If credit is requested, please send an email to [email protected].
Recall details
- Recall number
- Z-2295-2024
- Status
- Open, Classified
- Initiated by the company
- May 15, 2024
- Posted by the FDA
- Jul 5, 2024
- Recalling firm
- American Contract Systems Inc
- Quantity
- 36 units
- Product code
- OJH
- Distribution
- US Nationwide distribution.
Product description
HAND PACK 297829--Procedure Kit Catalog Number: RGHN04I
Code information (models, lots, serial numbers)
UDI-DI: 00191072203175 LOT#"s: 646231 927241
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ACS RGHN04IAmerican Contract Systems, Inc. · Orthopedic tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94765).
- Z-2274-2024 · Open, ClassifiedAHAR50K STANDARD ARTHROSCOPY PACK - 291852. Procedure tray Catalog Number: AHAR50K
- Z-2275-2024 · Open, ClassifiedTOTAL KNEE - 302497- Procedure tray Catalog Number: ANTK12AF
- Z-2276-2024 · Open, ClassifiedMAJOR KNEE - 302497- Procedure tray Catalog Number: ASKN23B
- Z-2277-2024 · Open, ClassifiedLOWER EXTREMITY - 247542- Procedure tray Catalog Number: BFLE56X
- Z-2278-2024 · Open, ClassifiedTOTAL KNEE - 247546 - Procedure tray Catalog Number: BFTK01AA
- Z-2279-2024 · Open, ClassifiedTOTAL KNEE - 247546 Procedure tray Catalog Number: BFTK01Z
- Z-2280-2024 · Open, ClassifiedARTHROSCOPY WOLFSON PACK BHS Procedure tray Catalog Number: BPAR65A
- Z-2281-2024 · Open, ClassifiedARTHROSCOPY SUP JOI PACK Procedure tray Catalog Number: BPAS17C
- Z-2282-2024 · Open, ClassifiedARTHO KNEE BAPTIST SOUTH PACK Procedure tray Catalog Number: BPKA49B
- Z-2283-2024 · Open, ClassifiedARTHRO KNEE BEACHES PACK BHS- Procedure tray Catalog Number: BPKB46B
- Z-2284-2024 · Open, ClassifiedBPKN15C KNEE BASIN PACK BHS- Procedure tray Catalog Number: BPKN15C
- Z-2285-2024 · Open, ClassifiedFOOT ANKLE-LF Procedure tray Catalog Number:CCFA51B
- Z-2286-2024 · Open, ClassifiedHAND WRIST FOREARM-LF Procedure tray Catalog Number: CCHD17B
- Z-2287-2024 · Open, ClassifiedKNEE SCOPE Procedure tray Catalog Number: COKN48L
- Z-2288-2024 · Open, ClassifiedUPPER EXTREMITY PK, STRL F G Catalog Number: IHUE40AG
- Z-2289-2024 · Open, ClassifiedACL, ST LUKES METHODIST 1CS Catalog Number: LMAC10AA
- Z-2290-2024 · Open, ClassifiedTOTAL KNEE, ST LUKES Catalog Number: LMTK17AP
- Z-2291-2024 · Open, ClassifiedKNEE ARTHROSCOPY PACK Catalog Number: MEKA12R
- Z-2292-2024 · Open, ClassifiedKNEE ARTHROSCOPY-Procedure Kit Catalog Number: QPKA35A
- Z-2293-2024 · Open, ClassifiedTOTAL KNEE -Procedure Kit Catalog Number: QPTK95A
- Z-2294-2024 · Open, ClassifiedTOTAL KNEE -Procedure Kit Catalog Number: QPTK95B
- Z-2296-2024 · Open, ClassifiedTOTAL KNEE - 297835-Procedure Kit Catalog Number: RGTK10K
- Z-2297-2024 · Open, ClassifiedTOTAL KNEE - 297835 -Procedure Kit Catalog Number: RGTK10L
- Z-2298-2024 · Open, ClassifiedKNEE ARTHROSCOPY PACK-Procedure Kit Catalog Number: SFAR60A
- Z-2299-2024 · Open, ClassifiedTOTAL KNEE-Procedure Kit Catalog Number: SFTK54B
- Z-2300-2024 · Open, ClassifiedHAND DRAPING SJH-Procedure Kit Catalog Number: SJHD67F
- Z-2301-2024 · Open, ClassifiedKNEE ARTHROSCOPY-Procedure Kit Catalog Number: SLAR14G
- Z-2302-2024 · Open, ClassifiedTOTAL KNEE-Procedure Kit Catalog Number: SMTK21C
- Z-2303-2024 · Open, ClassifiedARTHROSCOPY PACK-Procedure Kit Catalog Number: TNAR22U
- Z-2304-2024 · Open, ClassifiedARTHROSCOPY PACK-Procedure Kit Catalog Number: TNAR22V
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page