ACS UICY28LRecalls & reported problems
American Contract Systems, Inc. · MOR CYSTO PACK · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Sep 4, 2025Open, Classified
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Z-0174-2026 · Root cause: Nonconforming Material/Component · FDA record ↗
Other Endoscope introducer kit devices
- ACS BUCY78EAmerican Contract Systems, Inc. · Endoscope introducer kit2 recalls0 reports
- Medline Industries, Inc. DYNJ44673GMedline Industries, Inc. · Endoscope introducer kit2 recalls0 reports
- Medline Industries, Inc. DYNJ66447Medline Industries, Inc. · Endoscope introducer kit2 recalls0 reports
- Medline Industries, Inc. DYNJ58648JMedline Industries, Inc. · Endoscope introducer kit2 recalls0 reports
- CARDINAL HEALTH SBA73CLKFDCardinal Health 200, LLC · Endoscope introducer kit1 recall0 reports
- Medline Industries, Inc. DYKE1355Medline Industries, Inc. · Endoscope introducer kit1 recall0 reports
- Medline Industries, Inc. DYNJ60126BMedline Industries, Inc. · Endoscope introducer kit1 recall0 reports
- Medline Industries, Inc. DYK1012268E1Medline Industries, Inc. · Endoscope introducer kit1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.