Medical convenience kits recallZ-0174-2026 · American Contract Systems Inc. · initiated Sep 4, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from American Contract Systems Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Firm notified affected distributors who received affected product via email on 9/4/25. An updated letter was provided on 9/9/25, which better clarified product codes and impacted lot numbers and bag serial numbers. Consignees were instructed to immediately discontinue use and discard impacted trays. Distribution partners were provided with a customer letter for further distribution to end users.

Recall details

Recall number
Z-0174-2026
Status
Open, Classified
Initiated by the company
Sep 4, 2025
Posted by the FDA
Oct 16, 2025
Recalling firm
American Contract Systems Inc.
Quantity
32,433 total
Product code
NWU
Distribution
US Nationwide distribution in the states of SD, IA, MN, WA, IL.

Product description

Medical convenience kits Item Number/Description BUCY78E CYSTO PACK CMCP10X CYSTO PACK - 205972 IHCY10Y CYSTO PACK MHCY35AE CYSTO PACK- 212487 PLBY52E CYSTO PACK UDCY31Z CYSTO PACK - 206047 UICY28L MOR CYSTO PACK

Code information (models, lots, serial numbers)

BUCY78E UDI-DI 191072235350 Lot 8531411 Bag serial number 17705421; CMCP10X UDI-DI 191072228215 Lot 8392411 Bag serial number 17622370; IHCY10Y UDI-DI 191072159922 Lot 8026111 Bag serial number 17499185 Lot 8346211 Bag serial number 17692649 Lot 8444211 Bag serial number 17681124; MHCY35AE UDI-DI 191072228864 Lot 7990511 Bag serial number N/A; PLBY52E UDI-DI 191072141415 Lot 8016811 Bag serial number 17435184; UDCY31Z UDI-DI 00191072213990 Lot 8043211 Bag serial number 17483121 17483122 17483143 17483144 17485283; UICY28L UDI-DI 191072205537 Lot 8217311 Bag serial number 17598793 * If Bag serial number is listed as "N/A", then whole lot is affected

Devices named in this recall

7 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 97659).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page