ACS UIPC89DRecalls & reported problems
American Contract Systems, Inc. · PROCUREMENT PACK · Class 2
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Sep 4, 2025Open, Classified
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Z-0172-2026 · Root cause: Nonconforming Material/Component · FDA record ↗
- Jul 14, 2025Open, Classified
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
Z-2467-2025 · Root cause: Under Investigation by firm · FDA record ↗
Other General surgery tray devices
- Halyard MSBM036-03_0012Avid Medical, Inc. · General surgery tray5 recalls0 reports
- Medline Industries, Inc. DYNJ58534Medline Industries, Inc. · General surgery tray5 recalls0 reports
- Medline Industries, Inc. DYNJ82007AMedline Industries, Inc. · General surgery tray4 recalls0 reports
- Halyard KRIV26-02_0003Avid Medical, Inc. · General surgery tray4 recalls0 reports
- Medline Industries, Inc. DYNJ63410Medline Industries, Inc. · General surgery tray4 recalls0 reports
- Medline Industries, Inc. DYNJ908124AMedline Industries, Inc. · General surgery tray4 recalls0 reports
- Medline Industries, Inc. CDS983467JMedline Industries, Inc. · General surgery tray4 recalls0 reports
- Medline Industries, Inc. ACC010690Medline Industries, Inc. · General surgery tray4 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.