AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays… recallZ-2467-2025 · American Contract Systems Inc · initiated Jul 14, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from American Contract Systems Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
FDA root cause category: Under Investigation by firm
What the company is doing
Owens & Minor issued an "URGENT: MEDICAL DEVICE CORRECTION" notice to its consignees on 07/15/2025 via email. The notice explained the problem, potential risk to health, and requested the following: ACTIONS TO TAKE: " Discontinue use of all affected components. " Add warning labels to affected procedure trays to ensure the components are discarded. A labeling template is enclosed for printing labels which can be affixed on all affected kits in inventory. This label should be applied to each affected procedure tray and product case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. o Customers (end-users): If you receive a product from a distributor, please contact the distributor or your sales representative with any questions you may have. o Distributors: If any of the affected procedure trays identified above were further distributed, please ensure all end-users are appropriately notified of this product correction and maintain records of effectiveness for your consignees. " Complete the attached Recall Response Form and return as soon as possible to [email protected]. Report any quality problems or adverse events associated with the products listed in this notification to [email protected]. If you have any questions, contact the Manager, Regulatory Affairs at 470-280-4277 or email to: [email protected].
Recall details
- Recall number
- Z-2467-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 14, 2025
- Posted by the FDA
- Aug 28, 2025
- Recalling firm
- American Contract Systems Inc
- Quantity
- 6617 units
- Product code
- LRO
- Distribution
- US distribution to states of: IL and LA
Product description
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) A-C SUPPLEMENT PACK - 205946, Model/Item Number ANCV67X; 2) KIDNEY PERFUSION PACK - 208439, Model/Item Number ANKP15AD; 3) LAPCHOLAPPY PACK, Model/Item Number FYLA38G; 4) CYSTO PACK, Model/Item Number HSCY13M; 5) PICC DRESSING CHANGE KIT, Model/Item Number HSDR10K; 6) MIDLINE CHANGE KIT, Model/Item Number HSMD20H; 7) SHOULDER ARTHOSCOPY PACK, Model/Item Number PDSA12D; 8) PDA LF SJW, Model/Item Number SWPD12K; 9) AAA PACK - 206038, Model/Item Number UDAA63AH; 10) BASIC CAB PACK - 206043, Model/Item Number UDBC35AF; 11) BASIC HEART PACK - 206044, Model/Item Number UDBH37AX; 12) FREE FLAP DRAPE PACK (PS 907864), Model/Item Number UIFF72AD; 13) PROCUREMENT PACK, Model/Item Number UIPC89D; 14) THYROID PACK (PS 096552), Model/Item Number UITR45W; 15) TRAUMA PACK 175634, Model/Item Number UMTR17B; 16) TRAUMA PACK 175634, Model/Item Number UMTR17C.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
1) Model/Item Number ANCV67X: UDI/DI 00191072203021, Kit Lot Numbers: 42-7789611, 42-8074911, 42-7925811, 42-8005111; 2) Model/Item Number ANKP15AD: UDI/DI 00191072235176, Kit Lot Numbers: 42-7916611, 42-8047711; 3) Model/Item Number FYLA38G: UDI/DI 00191072003706, Kit Lot Numbers: 42-7905711, 42-8134111; 4) Model/Item Number HSCY13M: UDI/DI 00191072005571, Kit Lot Numbers: 42-8125311, 42-7959411; 5) Model/Item Number HSDR10K: UDI/DI 00191072005588, Kit Lot Numbers: 42-7984111, 42-7883711; 6) Model/Item Number HSMD20H: UDI/DI 00191072237262, Kit Lot Numbers: 42-7937911; 7) Model/Item Number PDSA12D: UDI/DI 00191072137111, Kit Lot Numbers: 70-053496; 8) Model/Item Number SWPD12K: UDI/DI 00191072228512, Kit Lot Numbers: 5437920, 5436824, 5436860, 5437058, 5437174, 5437197, 5437211, 5437326, 5437390, 5437696, 5437720, 5437752, 5437800, 5437859, 5438008, 5438056, 5438066, 5438331, 5438336, 5438446, 5438033 , 5438535; 9) Model/Item Number UDAA63AH: UDI/DI 00191072235275, Kit Lot Numbers: 42-8052111, 42-8104211; 10) Model/Item Number UDBC35AF: UDI/DI 00191072103420, Kit Lot Numbers: 42-7867111, 42-7940911, 42-7981511; 11) Model/Item Number UDBH37AX: UDI/DI 00191072012098, Kit Lot Numbers: 42-7867811, 42-7947511, 42-8007611, 42-8032411; 12) Model/Item Number UIFF72AD: UDI/DI 00191072235541, Kit Lot Numbers: 42-7849111, 42-7947711, 42-8024111, 42-8095211; 13) Model/Item Number UIPC89D: UDI/DI 00191072205742, Kit Lot Numbers: 42-8024311, 42-8113911; 14) Model/Item Number UITR45W: UDI/DI 00191072235589, Kit Lot Numbers: 42-7875311, 42-7980811, 42-8067111; 15) Model/Item Number UMTR17B: UDI/DI 00191072222596, Kit Lot Numbers: 9124674, 8925186, 9003138, 9051991 16) Model/Item Number UMTR17C: UDI/DI 00191072233806, Kit Lot Numbers: " 9202780, 9292532.
Devices named in this recall
16 devices in the FDA device registry whose identifier appears in the recall's code information.
- ACS UIPC89DAmerican Contract Systems, Inc. · General surgery tray2 recalls0 reports
- ACS FYLA38AAmerican Contract Systems, Inc. · General surgery tray1 recall0 reports
- ACS HSCY13CAmerican Contract Systems, Inc. · General surgery tray1 recall0 reports
- ACS HSDR10CAmerican Contract Systems, Inc. · General surgery tray1 recall0 reports
- ACS UDBH37AGAmerican Contract Systems, Inc. · General surgery tray1 recall0 reports
- ACS UDBC35AAAmerican Contract Systems, Inc. · General surgery tray1 recall0 reports
- ACS PDSA12AAmerican Contract Systems, Inc. · General surgery tray1 recall0 reports
- ACS ANCV67XAmerican Contract Systems, Inc. · General surgery tray1 recall0 reports
- ACS UMTR17BAmerican Contract Systems, Inc. · General surgery tray1 recall0 reports
- ACS SWPD12KAmerican Contract Systems, Inc. · General surgery tray1 recall0 reports
- ACS UMTR17CAmerican Contract Systems, Inc. · General surgery tray1 recall0 reports
- ACS ANKP15ADAmerican Contract Systems, Inc. · General surgery tray1 recall0 reports
- ACS UDAA63AHAmerican Contract Systems, Inc. · General surgery tray1 recall0 reports
- ACS UIFF72ADAmerican Contract Systems, Inc. · General surgery tray1 recall0 reports
- ACS UITR45WAmerican Contract Systems, Inc. · General surgery tray1 recall0 reports
- ACS HSMD20HAmerican Contract Systems, Inc. · General surgery tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97358).
- Z-2464-2025 · Open, ClassifiedAMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) HEART PACK - 205947, MODEL/ITEM NUMBER ANCV78BC; 2) VASCULAR PACK - 232024, MODEL/ITEM NUMBER ANVA82AB…
- Z-2465-2025 · Open, ClassifiedAMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAP CHOLE PACK, Model/Item Number CILC69L; 2) LAP CHOLE PACK, Model/Item Number HNLC80J; 3) LAP CHOLE…
- Z-2466-2025 · Open, ClassifiedAMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) CYSTO, Model/Item Number ASCY24A; 2) CYSTO PACK, Model/Item Number BUCY78E; 3) CYSTO PACK, Model/Item…
- Z-2468-2025 · Open, ClassifiedAMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) SFCH CRANI PACK (PS 132432), MODEL/ITEM NUMBER UICR10W; 2) SFCH SHUNT PACK (PS 139798), MODEL/ITEM NUM…
- Z-2469-2025 · Open, ClassifiedAMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) BORN ON ARRIVAL KIT, Model/Item Number LLBN11C; 2) LABOR & DELIVERY PACK, Model/Item Number LLLD19I.
- Z-2470-2025 · Open, ClassifiedAMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAPAROSCOPY PACK, MODEL/ITEM NUMBER EVLP61I; 2) PELVISCOPY, MODEL/ITEM NUMBER EVPV29G.
- Z-2471-2025 · Open, ClassifiedAMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ROBOT HYST PACK, MODEL/ITEM NUMBER GRRB10E; 2) VAG HYST PACK, MODEL/ITEM NUMBER SPVH62P.
- Z-2472-2025 · Open, ClassifiedAMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ANTERIOR HIP PACK, MODEL/ITEM NUMBER AMAH40N; 2)TOTAL KNEE DAN MYER, MODEL/ITEM NUMBER CCKM67F; 3)HIP…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page