ACS HGCV05KRecalls & reported problems
American Contract Systems, Inc. · CV SET UP PACK · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Sep 4, 2025Open, Classified
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Z-0176-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Jul 14, 2025Open, Classified
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
Z-2464-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Other Cardiovascular procedure kit devices
- Medline Industries, Inc. CDS840396YMedline Industries, Inc. · Cardiovascular procedure kit6 recalls0 reports
- Medline Industries, Inc. CDS983513FMedline Industries, Inc. · Cardiovascular procedure kit6 recalls0 reports
- Medline Industries, Inc. CDS840396XMedline Industries, Inc. · Cardiovascular procedure kit5 recalls0 reports
- Medline Industries, Inc. CDS984355GMedline Industries, Inc. · Cardiovascular procedure kit5 recalls0 reports
- Medline Industries, Inc. DYNJ51038RMedline Industries, Inc. · Cardiovascular procedure kit4 recalls2 reports
- Medline Industries, Inc. CDS984355IMedline Industries, Inc. · Cardiovascular procedure kit4 recalls0 reports
- NAMIC 60033123Medline Industries, Inc. · Cardiovascular procedure kit4 recalls0 reports
- Medline Industries, Inc. CDS840261ABMedline Industries, Inc. · Cardiovascular procedure kit4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page