AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays… recallZ-2464-2025 · American Contract Systems Inc · initiated Jul 14, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from American Contract Systems Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

FDA root cause category: Under Investigation by firm

What the company is doing

Owens & Minor issued an "URGENT: MEDICAL DEVICE CORRECTION" notice to its consignees on 07/15/2025 via email. The notice explained the problem, potential risk to health, and requested the following: ACTIONS TO TAKE: " Discontinue use of all affected components. " Add warning labels to affected procedure trays to ensure the components are discarded. A labeling template is enclosed for printing labels which can be affixed on all affected kits in inventory. This label should be applied to each affected procedure tray and product case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. o Customers (end-users): If you receive a product from a distributor, please contact the distributor or your sales representative with any questions you may have. o Distributors: If any of the affected procedure trays identified above were further distributed, please ensure all end-users are appropriately notified of this product correction and maintain records of effectiveness for your consignees. " Complete the attached Recall Response Form and return as soon as possible to [email protected]. Report any quality problems or adverse events associated with the products listed in this notification to [email protected]. If you have any questions, contact the Manager, Regulatory Affairs at 470-280-4277 or email to: [email protected].

Recall details

Recall number
Z-2464-2025
Status
Open, Classified
Initiated by the company
Jul 14, 2025
Posted by the FDA
Aug 28, 2025
Recalling firm
American Contract Systems Inc
Quantity
31506 units
Product code
OEZ
Distribution
US distribution to states of: IL and LA

Product description

AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) HEART PACK - 205947, MODEL/ITEM NUMBER ANCV78BC; 2) VASCULAR PACK - 232024, MODEL/ITEM NUMBER ANVA82AB; 3) OFF PUMP CABG PACK - 170316, MODEL/ITEM NUMBER BHOP48O; 4) OFF PUMP CABG PACK - 170316, MODEL/ITEM NUMBER BHOP48P; 5) PEDIATRIC OPEN HEART PACK - 170212, MODEL/ITEM NUMBER BHPH18F; 6) PEDIATRIC OPEN HEART PACK - 170212, MODEL/ITEM NUMBER BHPH18G; 7) CABG ACCESSORY PACK, MODEL/ITEM NUMBER HGCB98F; 8) CV SET UP PACK, MODEL/ITEM NUMBER HGCV05K; 9) OPEN HEART PACK B - COMPONENTS, MODEL/ITEM NUMBER HGOH99H; 10) STRADLE TABLE PACK - 301659, MODEL/ITEM NUMBER MHCV10AM; 11) MINOR VASCULAR PACK, MODEL/ITEM NUMBER MTVS73D; 12) AV GRAFT RCH 0246952, MODEL/ITEM NUMBER RCAV40C; 13) OPEN HEART PART 1 RCH 0247159, MODEL/ITEM NUMBER RCOH61G; 14) CARDIO 612968, MODEL/ITEM NUMBER TNCA08T; 15) CV B PACK, MODEL/ITEM NUMBER TNCB10AM; 16) A-V FISTULA - 206042, MODEL/ITEM NUMBER UDAV16AD; 17) CAROTID PACK - 206045, MODEL/ITEM NUMBER UDCP24AF; 18) FEM POP PACK - 206052, MODEL/ITEM NUMBER UDFP36AH; 19) CARDIOVASCULAR SUPPLY (PS 907285), MODEL/ITEM NUMBER UICD62AN; 20) SFCH PEDS CARDIAC SUPPLY PACK (PS131653), MODEL/ITEM NUMBER UICS88S; 21) SFCH ECMO PACK (PS 107116), MODEL/ITEM NUMBER UIPD03Q; 22) THORACIC SETUP (PS 035700), MODEL/ITEM NUMBER UITC53AL; 23) VASCULAR PACK, MODEL/ITEM NUMBER UIVA75J; 24) CABG ACCESSORY PACK 175619, MODEL/ITEM NUMBER UMCA14E; 25) OPEN HEART ACCESS PACK 175628, MODEL/ITEM NUMBER UMOH42E; 26) OPEN HEART ACCESS PACK 175628, MODEL/ITEM NUMBER UMOH42F; 27) OPEN HEART ACCESS PACK 175628, MODEL/ITEM NUMBER UMOH42G.

Code information (models, lots, serial numbers)

Show the full code information (3k characters)

1) MODEL/ITEM NUMBER ANCV78BC: UDI/DI 00191072235169, Kit Lot Numbers: 42-7844511, 42-8133811, 42-8037211, 42-7952411, 42-7882011; 2) MODEL/ITEM NUMBER ANVA82AB: UDI/DI 00191072235183, Kit Lot Numbers: 42-8025011, 42-8108511, 42-7928511; 3) MODEL/ITEM NUMBER BHOP48O: UDI/DI 00191072221490, Kit Lot Numbers: 9207467, 8862838, 9140871, 9049946, 9007346, 8934280; 4) MODEL/ITEM NUMBER BHOP48P: UDI/DI 00191072234544, Kit Lot Numbers: 9268782, 9345847; 5) MODEL/ITEM NUMBER BHPH18F: UDI/DI 00191072224651, Kit Lot Numbers: 8993526, 9098058, 8899297; 6) MODEL/ITEM NUMBER BHPH18G: UDI/DI 00191072234551, Kit Lot Numbers: 9238544; 7) MODEL/ITEM NUMBER HGCB98F: UDI/DI 00191072205285, Kit Lot Numbers: 42-7813711; 8) MODEL/ITEM NUMBER HGCV05K: UDI/DI 00191072212474, Kit Lot Numbers: 42-8097611, 42-8021411, 42-7870011, 42-7831911, 42-8093211; 9) MODEL/ITEM NUMBER HGOH99H: UDI/DI 00191072235336, Kit Lot Numbers: 42-8119511, 42-8044311, 42-7874411; 10) MODEL/ITEM NUMBER MHCV10AM: UDI/DI 00191072235619. Kit Lot Numbers: 42-7848211; 11) MODEL/ITEM NUMBER MTVS73D: UDI/DI 00191072237583, Kit Lot Numbers: 12-7397911, 12-7335111, 12-7072111, 12-7142511, 12-7224911; 12) MODEL/ITEM NUMBER RCAV40C: UDI/DI 00191072235381, Kit Lot Numbers: 42-8053511, 42-8151011, 42-7978811, 42-7853911; 13) MODEL/ITEM NUMBER RCOH61G: UDI/DI 00191072229823, Kit Lot Numbers: 42-7803611, 42-7941711, 42-7997711, 42-8075711; 14) MODEL/ITEM NUMBER TNCA08T: UDI/DI 00191072235534, Kit Lot Numbers: 42-7827011, 42-8158811; 15) MODEL/ITEM NUMBER TNCB10AM: UDI/DI 00191072232991, Kit Lot Numbers: 42-7864811, 42-7971311, 42-8065311; 16) MODEL/ITEM NUMBER UDAV16AD: UDI/DI 00191072216229, Kit Lot Numbers: 42-7919811; 17) MODEL/ITEM NUMBER UDCP24AF: UDI/DI 00191072235305, Kit Lot numbers: 42-8010311, 42-8109811; 18) MODEL/ITEM NUMBER UDFP36AH: UDI/DI 00191072235329, Kit Lot Numbers: 42-7835011, 42-7902811, 42-8014011, 42-8091211; 19) MODEL/ITEM NUMBER UICD62AN: UDI/DI 00191072200495, Kit Lot Numbers: 42-7874211, 42-7934511, 42-7999011, 42-8072211, 42-8132211; 20) MODEL/ITEM NUMBER UICS88S: UDI/DI 00191072235428, Kit Lot Numbers: 42-7821511, 42-7913211, 42-7979011, 42-8032611, 42-8106011; 21) MODEL/ITEM NUMBER UIPD03Q: UDI/DI 00191072235435, Kit Lot Numbers: 42-8083011, 42-8154511; 22) MODEL/ITEM NUMBER UITC53AL: UDI/DI 00191072235572, Kit Lot Numbers: 42-7887011, 42-7993911, 42-8071411; 23) MODEL/ITEM NUMBER UIVA75J: UDI/DI 00191072235602, Kit Lot Numbers: 42-7994011, 42-8056011, 42-8114011, 42-7827411, 42-7916411; 24) MODEL/ITEM NUMBER UMCA14E: UDI/DI 00191072222510, Kit Lot Numbers: 334022, 9238536; 25) MODEL/ITEM NUMBER UMOH42E: UDI/DI 00191072222534, Kit Lot Numbers: 8860283; 26) MODEL/ITEM NUMBER UMOH42F: UDI/DI 00191072224712, Kit Lot Numbers: 9125072, 9044687, 8977422, 8925203, 9167965; 27) MODEL/ITEM NUMBER UMOH42G: UDI/DI 00191072234896, Kit Lot Numbers: 9202772, 9334223.

Devices named in this recall

27 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 97358).

Sources and updates