Boston Scientific Corporation · Inflatable Penile Prosthesis, InhibiZone/Preconnected - Penoscrotal · Class 3 · Implant
All models of AMS 700 CXR TENACIO Pump (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00191506021528), received between Jan 2, 2025 and Aug 20, 2026 — classified as 5 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Inflation Problem” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 11, 2025Class IIOpen, Classified
The potential for devices to experience inflation and/or deflation performance issues or difficulties.
Z-0143-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Inflation Problem3
- Adverse Event Without Identified Device or Use Problem2
- Deflation Problem2
- Defective Device1
- Failure to Deflate1
- Inadequacy of Device Shape and/or Size1
- Mechanical Problem1
- Migration1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Pain2
- Unspecified Infection2
- Purulent Discharge1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryAug 20, 2026Inflation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50351070 ↗InjuryJul 17, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT IMPLANTED WITH AN INFLATABLE PENILE PROSTHESIS (IPP) EXPERIENCED SCROTAL PAIN. A SURGICAL PROCEDURE WAS PERFORMED DURING WHICH THE PHYSICIAN IDENTIFIED PURULENT MATERIAL (PUS) WITHIN THE SCROTUM, CONFIRMING THE PRESENCE OF AN INFECTION AT THE IMPLANT SITE. THE AFFECTED AREA WAS THOROUGHLY CLEANED AND THE IPP WAS SUBSEQUENTLY EXPLANTED AND REPLACED WITH A TACTRA DEVICE. NO ADDITIONAL PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Pain; Purulent Discharge; Unspecified Infection
MAUDE report 25856053 ↗InjuryFeb 24, 2026Mechanical Problem; Inadequacy of Device Shape and/or Size; Migration
IT WAS REPORTED THAT THE PATIENT'S INFLATABLE PENILE PROSTHESIS (IPP) EXHIBITED PUMP TUBING BUNCHING. DURING A SUBSEQUENT SURGICAL PROCEDURE, IT WAS DETERMINED THAT THE TUBING HAD BEEN INITIALLY TRIMMED TOO SHORT, CAUSING THE PUMP TO RIDE HIGH WITHIN THE PENOSCROTAL JUNCTION. THE PUMP WAS REPLACED WITH A NEW TENACIO PUMP TO PROVIDE ADEQUATE TUBING LENGTH AND ENSURE PROPER POSITIONING. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24434579 ↗MalfunctionOct 30, 2025Deflation Problem; Inflation Problem; Failure to Deflate; Deflation Problem; Inflation Problem; Failure to Deflate
IT WAS REPORTED THAT THE PATIENT WITH THIS INFLATABLE PENILE PROSTHESIS (IPP) EXPERIENCED DIFFICULTY INFLATING AND DEFLATING THE DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23425729 ↗InjuryMar 3, 2025Deflation Problem; Defective Device; Deflation Problem; Defective Device
IT WAS REPORTED THAT THE INFLATABLE PENILE PROSTHESIS (IPP) WAS TOO HARD TO PUMP AND SOMETIMES, THE DEVICE AUTO DEFLATES. IT WAS ALSO REPORTED PATIENT EXPERIENCED PAIN. A SURGICAL PROCEDURE WAS PERFORMED TO REPLACE THE SYSTEM WITHOUT PATIENT COMPLICATIONS.
Patient problems: Pain
MAUDE report 21510230 ↗MalfunctionFeb 6, 2025Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE SURGEON ATTEMPTED TO ACTIVATE THE PUMP OF THIS INFLATABLE PENILE PROSTHESIS (IPP), AND WHEN THEY REACHED WHAT THEY PERCEIVED AS 75% FULLNESS WITH FULL PUMPS, THE CYLINDERS WOULD NOT FILL ANY FURTHER. MULTIPLE ATTEMPTS WERE MADE TO PUMP AND INFLATE, BUT THE CYLINDERS ONLY REACHED FULL INFLATION WITH FLUTTER PUMPS FOR THE LAST 25%. NO SURGICAL INTERVENTION HAS BEEN INFORMED, AND NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21322790 ↗InjuryJan 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WITH THIS INFLATABLE PENILE PROSTHESIS (IPP) EXPERIENCE A SCROTAL INFECTION. THE EXISTING IPP WAS EXPLANTED, A WASHOUT WAS PERFORMED AND THE DEVICE WAS REPLACED WITH A MALLEABLE IMPLANT. THERE WERE NO FURTHER PATIENT COMPLICATIONS.
Patient problems: Unspecified Infection
MAUDE report 21064054 ↗
Other Prosthesis, penis, inflatable devices
- AMS 700 MS Pump 72404310Boston Scientific Corporation · Prosthesis, Penis, Inflatable992 reports
- AMS 700 LGX MS Pump 72404252Boston Scientific Corporation · Prosthesis, Penis, Inflatable1 death757 reports
- AMS 700 CX MS Pump 72404233-12Boston Scientific Corporation · Prosthesis, Penis, Inflatable656 reports
- AMS 700 CX MS Pump 72404232Boston Scientific Corporation · Prosthesis, Penis, Inflatable639 reports
- AMS 700 LGX MS Pump 72404251Boston Scientific Corporation · Prosthesis, Penis, Inflatable611 reports
- Conceal 720185-01Boston Scientific Corporation · Prosthesis, Penis, Inflatable586 reports
- AMS 700 LGX MS Pump 72404252-10Boston Scientific Corporation · Prosthesis, Penis, Inflatable561 reports
- AMS 700 CX MS Pump 72404232-10Boston Scientific Corporation · Prosthesis, Penis, Inflatable533 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page