Skip to content

AMS 700™ CXR TENACIO™ Pump 72404467Recalls & reported problems

Boston Scientific Corporation · Inflatable Penile Prosthesis, InhibiZone/Preconnected - Infrapubic · Class 3 · Implant

All models of AMS 700 CXR TENACIO Pump (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00191506021566), received between Jan 2, 2025 and Mar 31, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Mechanical Problem” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 11, 2025Class IIOpen, Classified

    The potential for devices to experience inflation and/or deflation performance issues or difficulties.

    Z-0143-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Mechanical Problem2
  • Air/Gas in Device1
  • Collapse1
  • Fluid/Blood Leak1
  • Inadequacy of Device Shape and/or Size1
  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 31, 2026Collapse; Fluid/Blood Leak; Mechanical Problem; Migration; Air/Gas in Device

    IT WAS REPORTED THAT THE PUMP OF THIS INFLATABLE PENILE PROSTHESIS (IPP) MIGRATED. UPON REVISION, AIR WAS NOTED WITHIN THE PUMP, AND THE PUMP WOULD NOT REFILL WHEN CONNECTED TO THE RESERVOIR. THE PUMP WAS ALSO FOUND TO BE FLAT. THE RESERVOIR WAS FLAT AND EMPTY UPON REMOVAL. NO LEAKS IN THE SYSTEM WERE APPARENT WHEN EXAMINED AND TESTED. AS A RESULT, THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24737968 ↗
  • InjuryJan 2, 2025Mechanical Problem; Inadequacy of Device Shape and/or Size; Mechanical Problem; Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THAT THE PUMP OF THIS INFLATABLE PENILE PROSTHESIS (IPP) WAS NOT WORKING. THE PATIENT WAS DISSATISFIED, AND IT WAS NOTICED THAT THE DEVICE WAS MIDSIZED. THE IPP WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21064057 ↗

Other Prosthesis, penis, inflatable devices

All Prosthesis, penis, inflatable devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page