Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00191506026363), received between May 5, 2026 and May 5, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Difficult or Delayed Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult or Delayed Separation1
  • Malposition of Device1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 5, 2026Malposition of Device; Difficult or Delayed Separation

    IT WAS REPORTED THAT MALPOSITION OF THE DEVICE OCCURRED. AN EMBOLD SOFT 2X2 COIL WAS SELECTED FOR AN EMBOLIZATION PROCEDURE FOR A LOWER GASTROINTESTINAL BLEED. THE COIL WAS DEPLOYED; HOWEVER, AFTER ATTEMPTED DETACHMENT, THE COIL WAS ADHERENT TO THE PUSHER WIRE. MULTIPLE MANEUVERS WERE ATTEMPTED TO RETRACT THE COIL INSIDE THE MICROCATHETER (TRUSELECT 130CM) BUT WERE NOT SUCCESSFUL. THE DECISION WAS MADE TO ATTEMPT TO RETRACT THE COIL AND MICROCATHETER TOGETHER UNTIL EXTERNALIZED. ONCE THE MICROCATHETER AND COIL CONTACTED THE BASE CATHETER, THE COIL DETACHED AND MIGRATED DOWN TO THE ILEOCOLIC ARTERY. A SUPERIOR MESENTERIC ARTERIOGRAM WAS PERFORMED VIA THE BASE CATHETER, SHOWING THE COIL IN A DISTAL BRANCH OF THE ILEOCOLIC ARTERY; HOWEVER, NO EVIDENCE OF DECREASED FLOW TO THE ILEOCOLIC SEGMENT WAS NOTED. NO FURTHER PRODUCTS WERE USED AFTER THE FACT. THE PATIENT BLOOD PRESSURE WAS STABILIZED AND PRESSOR SUPPORT WAS DISCONTINUED. THE PATIENT LEFT THE DEPARTMENT IN STABLE CONDITION WITH NO COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 25088275 ↗

Other Device, Vascular, for Promoting Embolization devices