EMBOLD™ Soft M001394250020040Recalls & reported problems

Boston Scientific Corporation · Detachable Coil System · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00191506026370), received between Oct 3, 2024 and Jan 22, 2026 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Migration2
  • Separation Failure2
  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 22, 2026Migration

    IT WAS REPORTED THAT THE COIL MIGRATED POST-DEPLOYMENT. AN EMBOLD SOFT 2 X 4 COIL WAS SELECTED FOR USE DURING A GASTROINTESTINAL (GI) BLEED PROCEDURE. THE VESSEL WAS MILDLY TORTUOUS, AND THE COIL FULLY DEPLOYED WITHOUT ANY ISSUES. WHILE ATTEMPTING TO ACCESS ANOTHER BRANCH USING A .014 NON-BOSTON SCIENTIFIC WIRE, THE WIRE INADVERTENTLY ENTERED THE WRONG BRANCH AND HOOKED THE DEPLOYED COIL, RESULTING IN COIL DISLODGEMENT. THE ENTIRE SYSTEM, INCLUDING THE MICROCATHETER AND BASE CATHETER, WAS WITHDRAWN AND FLUSHED; HOWEVER, THE COIL COULD NOT INITIALLY BE LOCATED. UPON REINSERTION OF THE BASE CATHETER, THE COIL WAS VISUALIZED UNDER FLUOROSCOPY PROTRUDING FROM THE SIDE HOLE OF THE BASE CATHETER AND WAS SUCCESSFULLY RETRIEVED. PRIOR TO THIS ISSUE, OTHER EMBOLD COILS HAD BEEN SUCCESSFULLY IMPLANTED; THEREFORE, THE PROCEDURE WAS COMPLETE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24151051 ↗
  • MalfunctionJul 30, 2025Difficult to Remove; Migration; Difficult to Remove; Migration

    IT WAS REPORTED THAT THE COIL MIGRATED POST DEPLOYMENT. A 2MM X 4CM EMBOLD SOFT COIL WAS SELECTED FOR USE IN A PROSTATIC ARTERY EMBOLIZATION (PAE) PROCEDURE. THE VESSEL WAS 2-3MM, AND THE TORTUOSITY WAS MODERATE TO SEVERE. DURING THE PROCEDURE, THE COIL COULD NOT BE RETRACTED BACK INTO THE MICROCATHETER. THE COIL WAS DEPLOYED, AND IT WAS CONSIDERED SUCCESSFUL, BUT AFTERWARDS THE COIL MOVED OUT OF POSITION INTO THE SUPERIOR GLUTEAL ARTERY. THE COIL WAS PULLED OUT WITH THE SHEATH. AFTER RETRIEVING THE COIL, THE PHYSICIAN DECIDED TO CLOSE AND COME BACK ANOTHER DAY GIVEN THE EVENTS. ONLY ONE SIDE OF THE PROCEDURE WAS TREATED SUCCESSFULLY; THIS SIDE WAS NOT. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22668553 ↗
  • InjuryOct 3, 2024Break; Separation Failure; Break; Separation Failure

    IT WAS REPORTED THAT THE COIL WAS DIFFICULT TO DEPLOY, REQUIRING A SNARE TO REMOVE. TWO 2X4 EMBOLD SOFT DETACHABLE COIL SYSTEMS WAS SELECTED FOR USE DURING A BLEED COILING PROCEDURE. DURING THE PROCEDURE, THE COIL WAS DIFFICULT TO DEPLOY IN THE MILD TO MODERATELY TORTUOUS TARGET LOCATION. SEVERAL ATTEMPTS WERE MADE TO DEPLOY THE COIL, BUT DIFFICULTY WAS ENCOUNTERED DUE TO SMALL VASCULATURE. IT WAS NOTED THAT THE PULL WIRE/DETACHMENT WIRE WAS BROKEN. A SNARE WAS USED TO RETRIEVE THE COIL, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20359284 ↗
  • InjuryOct 3, 2024Break; Separation Failure; Break; Separation Failure

    IT WAS REPORTED THAT THE COIL WAS DIFFICULT TO DEPLOY, REQUIRING A SNARE TO REMOVE. TWO 2X4 EMBOLD SOFT DETACHABLE COIL SYSTEMS WAS SELECTED FOR USE DURING A BLEED COILING PROCEDURE. DURING THE PROCEDURE, THE COIL WAS DIFFICULT TO DEPLOY IN THE MILD TO MODERATELY TORTUOUS TARGET LOCATION. SEVERAL ATTEMPTS WERE MADE TO DEPLOY THE COIL, BUT DIFFICULTY WAS ENCOUNTERED DUE TO SMALL VASCULATURE. IT WAS NOTED THAT THE PULL WIRE/DETACHMENT WIRE WAS BROKEN. A SNARE WAS USED TO RETRIEVE THE COIL, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20359278 ↗

Other Device, Vascular, for Promoting Embolization devices