EMBOLD™ Soft M001394250080600Recalls & reported problems
Boston Scientific Corporation · Detachable Coil System · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00191506026578), received between Apr 29, 2024 and Apr 28, 2025 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Difficult to Remove2
- Adverse Event Without Identified Device or Use Problem1
- Difficult to Insert1
- Material Integrity Problem1
- Premature Activation1
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Foreign Body In Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 28, 2025Break; Difficult to Insert; Difficult to Remove; Material Integrity Problem; Break; Difficult to Insert; Difficult to Remove; Material Integrity Problem
IT WAS REPORTED THAT THE TIP BROKE. AN 8MM X 60 CM EMBOLD SOFT COIL WAS SELECTED FOR USE IN A SPLENIC EMBOLIZATION PROCEDURE. DURING THE PROCEDURE, THE TIP BROKE DUE TO DIFFICULTY RE-CAPTURING INTO THE INTRODUCER SHEATH. THE DEVICE WAS EXCHANGED FOR ANOTHER ONE, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21922542 ↗InjuryDec 19, 2024Break; Difficult to Remove; Stretched; Adverse Event Without Identified Device or Use Problem; Break; Difficult to Remove; Stretched; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE COIL BROKE, AND A PORTION REMAINED IN THE VESSEL. AN EMBOLD SOFT 8X60 WAS SELECTED FOR A PROCEDURE. DURING THE RETRACTION OF THE COIL, HOWEVER, RESISTANCE WAS FELT, AND THE COIL STRETCHED AND BROKE. A PORTION OF THE COIL WAS REMOVED, WHILE THE BROKEN PORTION WAS LEFT IN THE TARGET VESSEL. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 20979221 ↗MalfunctionApr 29, 2024Premature Activation; Premature Activation
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(6) 2024. IT WAS REPORTED THAT THE COIL PREMATURELY DEPLOYED INSIDE THE MICROCATHETER. AN 8MM X 60CM EMBOLD SOFT DETACHABLE COIL SYSTEM AND A 60CM EMBOLD PACKING DETACHABLE COIL SYSTEM WERE SELECTED FOR USE IN A COIL EMBOLIZATION PROCEDURE TO TREAT A PSEUDOANEURYSM. THE ANATOMY WAS TORTUOUS. A TRUSELECT MICROCATHETER WAS USED TO DELIVER THE 8MM X 60 CM EMBOLD SOFT COIL FIRST. HOWEVER, IT WAS DETERMINED THAT THE COIL WAS UNDERSIZED, SO IT WAS RETRACTED BACK INTO THE MICROCATHETER. THE COIL PREMATURELY DEPLOYED FROM THE DELIVERY WIRE. THE MICROCATHETER WAS PULLED OUT WITH THE DEPLOYED COIL INSIDE, AND ACCESS WAS LOST. AFTER ACCESS WAS RE-GAINED, AND EMBOLD FIBERED COIL WAS SUCCESSFULLY DEPLOYED. FOLLOWING THAT, THE 60CM EMBOLD PACKING COIL WAS ADVANCED. THE COIL WAS NOT SITTING LIKE THE PHYSICIAN WANTED, SO THE COIL WAS RETRACTED. THE COIL PREMATURELY DEPLOYED, AND THE MICROCATHETER WAS PULLED TO REMOVE THE COIL. AFTER ACCESS WAS RE-GAINED, THE CASE WAS COMPLETED WITH NON-BOSTON SCIENTIFIC COILS. ONE OF THE NON-BOSTON SCIENTIFIC COILS ALSO PREMATURELY DEPLOYED DURING THIS CASE. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, DEVICE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19209484 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page