EMBOLD™ Packing M001394260000450Recalls & reported problems
Boston Scientific Corporation · Detachable Coil System · Class 2 · Implant
All models of EMBOLD Packing (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00191506026608), received between Nov 15, 2024 and May 5, 2026 — classified as 4 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Stretched4
- Difficult to Remove3
- Device-Device Incompatibility1
- Difficult to Advance1
- Difficult to Insert1
- Failure to Advance1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Foreign Body In Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2026Break
IT WAS REPORTED THAT A BREAK OCCURRED. AN 45CM EMBOLD PACKING WAS SELECTED FOR USE. DURING THE PROCEDURE, THE COIL PARTS WERE BROKEN/SEPARATED. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25084756 ↗InjuryJun 11, 2025Break; Difficult to Remove; Stretched; Positioning Problem; Break; Difficult to Remove; Stretched; Positioning Problem
IT WAS REPORTED THAT THE CATHETER KICKED-BACK UPON COIL PACKING, AND THE COIL BROKE UPON REMOVAL. THIS PATIENT WAS RECEIVING EMBOLIZATION TREATMENT FOR BLEEDING IN THE GASTRIC ARTERY. DURING THE PROCEDURE, A 3 MM EMBOLD FIBERED COIL DELIVERED THEN THE REST WAS FILLED IN WITH THIS 45CM EMBOLD PACKING COIL. THE COIL WAS ADVANCED EASILY INTO THE ARTERY, BUT DURING PACKING, THE COIL PUSHED THE MICROCATHETER OUT OF THE VESSEL. THE COIL WAS ATTEMPTED TO BE RETRACTED, BUT THE COIL UNRAVELED/STRETCHED AND WAS UNABLE TO BE PULLED ALL THE WAY BACK INTO THE MICROCATHETER AND EVENTUALLY BROKE. THE COIL WAS PARTIALLY EXPLANTED, AND A PORTION STILL REMAINS INSIDE THE PATIENT. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION INDICATED THAT A SNARE WAS ATTEMPTED TO BE USED TO REMOVE THE PORTION THAT REMAINED INSIDE THE PATIENT, AND IT ENDED UP PUSHING COIL INTO PARENT VESSEL.
Patient problems: Foreign Body In Patient
MAUDE report 22188713 ↗InjuryJun 5, 2025Break; Difficult to Remove; Stretched; Break; Difficult to Remove; Stretched
IT WAS REPORTED THAT THE COIL BROKE AND WAS DIFFICULT TO REMOVE, REQUIRING AN ADDITIONAL DEVICE. THIS EMBOLD PACKING 45 WAS SELECTED FOR USE DURING AN EPIGASTRIC EMBOLIZATION PROCEDURE IN MILDLY TORTUOUS ANATOMY. DURING THE PROCEDURE WHEN THIS EMBOLD PACKING COIL WAS BEING ADVANCED, THE PUSHER WIRE DETACHED, AND THE INTERNAL WIRE STAYED ATTACHED TO THE COUPLERS. AN ATTEMPT WAS MADE TO REMOVE THE ENTIRE SYSTEM; HOWEVER, ONLY THE DELIVERY WIRE CAME OUT WITH BOTH COUPLERS REMAINING ON THE DELIVERY WIRE. A SNARE WAS USED TO PULL OUT THE COIL WHICH EXPERIENCED STRETCHING AND BREAKING IN SEVERAL AREAS. THREE DIFFERENT SNARES WERE USED TO EXTRACT THE COIL, AND THE PROCEDURE WAS COMPLETED USING PUSHABLE COILS. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THIS DID NOT OCCUR DURING AN ATTEMPT TO DEPLOY THE COILS, AS THE PROXIMAL RELEASE PERFORATION WAS NOT BROKEN. THERE WAS A DETACHMENT AT THE COUPLERS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22155100 ↗InjuryMay 21, 2025Stretched; Material Protrusion/Extrusion; Stretched; Material Protrusion/Extrusion
IT WAS REPORTED VIA THAT THE COIL WAS STRETCHED AND PROTRUDING FROM THE TARGET LOCATION, REQUIRING EXPLANTATION. A 45CM EMBOLD PACKING COIL WAS SELECTED FOR USE IN AN EMBOLIZATION PROCEDURE IN THE 4MM GASTRODUODENAL ARTERY (GDA). A 4MM FIBERED COIL WAS PLACED DISTALLY PRIOR TO THE PLACEMENT OF THIS COIL. THIS EMBOLD PACKING COIL WAS PLACED IN THE GDA WITHOUT INCIDENT. HOWEVER, UPON FOLLOW UP USING CT, THE COIL WAS FOUND TO BE STRETCHED AND EXTENDING OUT OF THE TARGET VESSEL INTO THE ORIGIN OF THE CELIAC ARTERY INTO THE PROXIMAL AORTA AND THEN DESCENDING TO THE RIGHT COMMON FEMORAL ARTERY. THE COIL WAS SCHEDULED TO BE EXPLANTED ON (B)(6) 2025.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22066771 ↗InjuryMar 27, 2025Break; Stretched; Difficult to Advance; Break; Stretched; Difficult to Advance
IT WAS REPORTED THAT THE COIL BROKE AND REQUIRED A SNARE FOR RETRIEVAL. AN EMBOLD PACKING 45 WAS SELECTED FOR A HYPO EMBOLIZATION PROCEDURE IN THE MILDLY TORTUOUS INTERNAL ILIAC ARTERY. WHILE ADVANCING THE COIL USING INTERMITTENT FLUSHING, IT BECAME DIFFICULT TO PUSH FROM THE DISTAL PART OF THE 2.4 FRENCH NON-BOSTON SCIENTIFIC MICROCATHETER. IT WAS ATTEMPTED TO RETRACT THE COIL, BUT THE COIL STRETCHED TO THE POINT OF BREAKING. THE COIL BROKE OFF WITH A PORTION IN THE CATHETER AND A PORTION IN THE COIL NEST. THE MICROCATHETER WAS REMOVED WITH ONE FRACTURED PORTION OF THE COIL INSIDE, AND THE FRACTURED PORTION IN THE COIL NEST WAS REMOVED WITH A SNARE. NO COIL WAS LEFT IN THE PATIENT. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21713059 ↗MalfunctionFeb 5, 2025Difficult to Insert; Device-Device Incompatibility; Material Deformation; Difficult to Insert; Device-Device Incompatibility; Material Deformation
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 10JAN2025. AN EMBOLD PACKING 45 COIL WAS SELECTED FOR USE. THE PHYSICIAN HAD DELIVERED A COMBINATION OF EMBOLD FIBERED, SOFT, AND PACKING COILS VIA A NON-BOSTON SCIENTIFIC MICROCATHETER WITHOUT DIFFICULTY. THERE WAS A DIFFICULT BRANCH TO ENGAGE WITH THE FIRST MICROCATHETER, SO THE DEVICE WAS EXCHANGED FOR A SHORTER NON-BOSTON SCIENTIFIC MICROCATHETER. ONCE THE VESSEL WAS ENGAGED, THE PHYSICIAN WANTED TO DEPLOY AN EMBOLD PACKING COIL. THE PHYSICIAN MET RESISTANCE WHEN TRYING TO ADVANCE THE COIL INTO THE MICROCATHETER AND NOTICED THE DISTAL TIP OF THE COIL WAS KINKED. THE COIL WAS REMOVED AND NOT IMPLANTED. THE PHYSICIAN ASKED FOR A SECOND EMBOLD PACKING COIL AND HAD SOME DIFFICULTY ADVANCING AGAIN BUT WAS ABLE TO GET THE COIL WITHIN THE MICROCATHETER. HOWEVER, WITH THE INCREASED RESISTANCE WITH ADVANCING THE COIL, THE PHYSICIAN DECIDED TO REMOVE THE SECOND COIL. UPON RETRACTING THE COIL, THE COIL BEGAN TO STRETCH. THE ENTIRE COIL WAS REMOVED FROM THE MICROCATHETER. AT THIS POINT, THE PHYSICIAN DECIDED TO STOP THE CASE AND REEVALUATE THE IMAGES. THE PHYSICIAN THEN REALIZED WHAT WAS NEEDED TO BE DONE WAS ACCOMPLISHED AND THE PROCEDURE W
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21314904 ↗MalfunctionDec 6, 2024Break; Difficult to Remove; Failure to Advance; Break; Difficult to Remove; Failure to Advance
IT WAS REPORTED THAT A COIL BREAK OCCURRED. AN EMBOLD PACKING COIL WAS SELECTED FOR USE IN A BILATERAL RENAL ARTERY EMBOLIZATION PROCEDURE. THE VESSEL WAS MILDLY TORTUOUS. AFTER IMPLANTING A FIBERED COIL WITH NO ISSUE, THE EMBOLD PACKING COIL WAS DELIVERED WITH A PULSATION METHOD. IT WAS NOTED THAT THE COIL WAS NOT ADVANCING AND THE COIL BECAME DETACHED INTO TWO PIECES. DURING AN ATTEMPT TO PULL THE COIL BACK, IT MAINTAINED INSIDE THE 2.4 FRENCH NON-BOSTON SCIENTIFIC MICROCATHETER. SEVERAL WIRES WERE USED TO PUSH THE COIL FORWARD AND A WIRE WAS PLACED ALONG SIDE OF THE COIL AND ANCHORED IT AGAINST THE MICRO CATHETER. EVERYTHING WAS ABLE TO BE RETRIEVED. THE PROCEDURE WAS COMPLETED WITH ANOTHER BASE AND MICRO CATHETER. THERE WERE NO PATIENT COMPLICATIONS AND THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20866255 ↗MalfunctionNov 15, 2024Break; Separation Failure; Break; Separation Failure
IT WAS REPORTED THAT A COIL BREAK OCCURRED. AN EMBOLD PACKING 45 WAS SELECTED FOR AN EMBOLIZATION PROCEDURE IN THE PORTAL VEIN. DURING THE PROCEDURE, THE COIL DID NOT DETACH CORRECTLY AND BROKE OFF IN THE CATHETER. THE COIL WAS REMOVED FROM THE CATHETER. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20694685 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page