OBSIDIO™ M0013972001010Recalls & reported problems

Boston Scientific Corporation · Conformable Embolic · Class 2 · Implant

All models of OBSIDIO (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
2
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00191506039332), received between Nov 20, 2023 and Aug 24, 2026 — classified as 2 deaths, 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Migration” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 20, 2024Class IOpen, Classified

    An investigation determined that delivery of the Obsidio embolic using the aliquot technique for lower gastrointestinal bleeding embolization poses a high risk of bowel ischemia. The most serious and the most common adverse health consequence, reasonably foreseeable to occur, is the need to perform major surgery such as bowel resection and/or diverting colostomy. Therefore, Boston Scientific does not recommend that the aliquot technique be used to deliver the Obsidio device for lower GI bleed embolization procedures.

    Z-1374-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Migration3
  • Adverse Event Without Identified Device or Use Problem1
  • Material Separation1
  • Use of Device Problem1

Patient problems

  • Embolism/Embolus5
  • Ischemia2
  • Abdominal Pain1
  • Multiple Organ Dysfunction Syndrome1
  • Necrosis1
  • Obstruction/Occlusion1
  • Pancreatitis1
  • Sepsis1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 24, 2026Migration

    Patient problems: Embolism/Embolus

    MAUDE report 50410408 ↗
  • MalfunctionDec 4, 2025Migration; Migration

    IT WAS REPORTED THAT MIGRATION OCCURRED. OBSIDIO WAS SELECTED FOR USE IN A SPLENIC EMBOLIZATION PROCEDURE. OBSIDIO WAS USED OFF LABEL FOR THE VESSEL SIZE AND TRAVELED OUTSIDE THE DESIRED AREA OCCLUDED. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE DESIRED AMOUNT OF OBSIDIO THAT TRAVELED OUTSIDE THE PARAMETERS WAS INTENDED AND PURPOSEFUL AND THE PATIENT DID NOT EXPERIENCE ANY HARM.

    Patient problems: Embolism/Embolus

    MAUDE report 23722659 ↗
  • DeathMar 26, 2025Migration; Migration

    IT WAS REPORTED VIA JOURNAL ARTICLE, THAT MIGRATION OCCURRED RESULTING IN PATIENT COMPLICATIONS AND DEATH. OBSIDIO WAS SELECTED FOR AN EMBOLIZATION PROCEDURE OF THE POSTERIOR SUPERIOR PANCREATICODUODENAL ARTERY (PSPDA). THE PATIENT WAS ADMITTED FOR HYPOXEMIC RESPIRATORY FAILURE AND SWEETS SYNDROME, AND THE HOSPITAL COURSE WAS COMPLICATED BY MELENA. THE PATIENTS HEMOGLOBIN DECLINED OVERNIGHT FROM 7.1 TO 5.8 G/DL WITH RECURRENT MELENA AND HEMATEMESIS REQUIRING A BLOOD TRANSFUSION. A COMPUTERIZED TOMOGRAPHY (CT) ANGIOGRAM DEMONSTRATED ACTIVE EXTRAVASATION INTO THE DUODENUM. ANGIOGRAPHY OF THE PSPDA WITH A 2.8 FRENCH NON-BOSTON SCIENTIFIC MICROCATHETER DEMONSTRATED A PSEUDOANEURYSM OF A SMALL BRANCH WHERE SUPER SELECTIVE CATHETERIZATION WAS IMPOSSIBLE. EMBOLIZATION OF THE PSPDA USING OBSIDIO WAS PERFORMED, WITH THE GOAL OF OCCLUDING BRANCHES BEYOND THE REACH OF THE MICROCATHETER. THE ALIQUOT TECHNIQUE WAS UTILIZED TO PRESERVE THE MICROCATHETER FOR LATER USE. WHILE APPLYING A SLOW INJECTION FORCE, 0.2 ML ALIQUOTS OF OBSIDIO WAS FLUSHED WITH A 3 ML SYRINGE AND IOHEXOL, WHICH WAS SURMISED TO BETTER RESEMBLE THE VISCOSITY OF OBSIDIO COMPARED TO NORMAL SALINE. THE EMBOLIC WAS ADMINISTERED U

    Patient problems: Abdominal Pain; Embolism/Embolus; Ischemia; Multiple Organ Dysfunction Syndrome; Necrosis; Pancreatitis; Sepsis

    MAUDE report 21696264 ↗
  • DeathFeb 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT DEATH OCCURRED. OBSIDIO WAS SELECTED FOR USE IN A TERMINAL ILEAL BLEED. THE ALIQUOT TECHNIQUE WAS USED WITH CONTRAST AND THE VESSEL SIZE WAS 2MM. THE DEVICE WAS SUCCESSFULLY IMPLANTED INTO TWO BRANCHES OFF OF THE ILEOCOLIC ARTERY. A BOWEL INFARCTION WAS NOTICED ON POST COMPUTERIZED TOMOGRAPHY (CT) IMAGING. APPROXIMATELY FOUR DAYS POST PROCEDURE, THE PATIENT EXPERIENCED PNEUMATOSIS AND UNDERWENT SURGERY FOR A PARTIAL BOWEL RESECTION. THE PATIENT REMAINED HOSPITALIZED UNTIL PASSING AWAY. IT WAS FURTHER REPORTED THAT THE DOCUMENTED CAUSE OF DEATH WAS RESPIRATORY FAILURE AND CARDIAC ARREST.

    Patient problems: Ischemia; Unspecified Tissue Injury

    MAUDE report 18769491 ↗
  • InjuryNov 21, 2023Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT OBSIDIO EMBOLIZED. THE TARGET LOCATION WAS IN THE GASTRODUODENAL ARTERY (GDA). THE VESSEL SIZE WAS 1 - 3MM. OBSIDIO WAS SELECTED FOR USE. AFTER CREATING CASTING OF THE OBSIDIO, THE FELLOW WAS INSTRUCTED BY THE PHYSICIAN TO ADMINISTER A LARGE BOLUS OF SALINE AND OBSIDIO MATERIAL THROUGH ALIQUOT TECHNIQUE. THE REASON FOR THE USE OF THE ALIQUOT METHOD WAS TO RE-USE THE MICROCATHETER. THE ALIQUOT METHOD RESULTED IN A TOTAL OF .6 ML OF OBSIDIO BEING USED AND THE CATHETER, IN USE, BEING PLACED TOO CLOSE TO THE OBSIDIO CASTING. THE OBSIDIO FROM THE SECOND BOLUS/ALIQUOT, MOVED TOWARDS THE BRANCH POINT OF THE RIGHT HEPATIC ARTERY. THE FLOW OF THE VESSEL AND INJECTION OF CONTRAST DURING A HAND INJECTION RUN TOOK THE OBSIDIO EVEN MORE DISTAL. THE OBSIDIO SHUT DOWN THE RIGHT HEPATIC ARTERY SUPPLY. ACTION WAS TAKEN TO BREAK UP THE OBSIDIO INTO PIECES USING A MICROCATHETER. SMALLER PIECES OF OBSIDIO TRAVELED DISTALLY TO THE PANCREAS. MINOR FLOW WAS ACHIEVED BUT NOT ENTIRELY. IMAGING, POST PROCEDURE, REVEALED THE PATIENT HAD PANCREAS FLOW SHUT DOWN RESULTING INTO OTHER COMPLICATIONS LIKE PANCREAS INFARCT.

    Patient problems: Embolism/Embolus; Obstruction/Occlusion

    MAUDE report 18184607 ↗
  • InjuryNov 20, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT THE OBSIDIO BROKE OFF AND EMBOLIZED. THE TARGET LOCATION WAS IN THE GASTRODUODENAL ARTERY (GDA). THE VESSEL SIZE WAS 2.0 - 2.7 MM. OBSIDIO WAS SELECTED FOR USE. AFTER OBSIDIO WAS ADMINISTERED, THE MICROCATHETER WAS LEFT IN PLACE FOR A FEW MINUTES PRIOR TO REMOVING IT. A PIECE OF OBSIDIO BROKE OFF AND TRAVELED TO APPROXIMATELY 1CM AWAY FROM THE OSTIUM. THE TARGET LOCATION WAS BACKFILLED WITH AN EMBOLD FIBERED COIL, AND THE OBSIDIO REMAINED STABLE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: Embolism/Embolus

    MAUDE report 18176969 ↗

FDA clearance and classification

The FDA regulates product code KRD (device, vascular, for promoting embolization) under 21 CFR 870.3300 as a Class II device. 21 CFR 870.3300 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the OBSIDIO™ M0013972001010

Has the OBSIDIO™ M0013972001010 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1374-2024 (Class I), began on Feb 20, 2024: An investigation determined that delivery of the Obsidio embolic using the aliquot technique for lower gastrointestinal bleeding embolization poses a high risk of bowel ischemia. The most serious and the most common… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the OBSIDIO™ M0013972001010?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between Nov 20, 2023 and Aug 24, 2026. In those reports, the most common specific device problems are “Migration” (3), “Material Separation” (1) and “Use of Device Problem” (1); the most common patient problems are “Embolism/Embolus” (5), “Ischemia” (2) and “Abdominal Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the OBSIDIO™ M0013972001010?

2 reports are classified as deaths, 2 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the OBSIDIO™ M0013972001010 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the OBSIDIO™ M0013972001010, and is it still sold?

The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand OBSIDIO™, model M0013972001010). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →