FARADRIVE™ M004PF21M402Recalls & reported problems
Boston Scientific Corporation · Steerable Sheath Clear · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Boston Scientific Faradrive steerable sheath has 2,160 reports, which include 13 deaths and 271 injuries. Most describe air in the device, and many describe leaks, including valves leaking fluid and blood. 1 recall is open, from June 2026, over valve damage from how the dilator is inserted during preparation, which can let air in. Reports fell to 284 in the last 12 months from 1,630 in the previous 12.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 2,303
- Deaths reported
- 16
- Injuries reported
- 309
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2,303 adverse event reports that list this exact device (DI 00191506043131), received between Apr 18, 2024 and Sep 30, 2026 — classified as 16 deaths, 309 injuries, 1,978 malfunctions.
The most frequently reported specific device problem is “Air/Gas in Device” (1,383 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jun 11, 2026Class IIOpen, Classified
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Z-2756-2026 · Root cause: Use error · Recall details · FDA record ↗
Most reported problems
Device problems
- Air/Gas in Device1,383
- Leak/Splash450
- Adverse Event Without Identified Device or Use Problem362
- Device Contamination with Body Fluid235
- Material Integrity Problem132
- Nonstandard Device129
- Fluid/Blood Leak114
- Difficult to Advance98
Patient problems
- No Clinical Signs, Symptoms or Conditions1,963
- Low Blood Pressure/ Hypotension90
- Pericardial Effusion90
- Non specific EKG/ECG Changes84
- Thrombosis/Thrombus51
- Air Embolism42
- Cardiac Tamponade36
- Cardiac Perforation34
Recent adverse event reports
The 25 most recent of 2,303 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50941475 ↗InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem
It was reported that a FARADRIVE steerable sheath was selected for use during a procedure. On (b)(6) an ultrasound examination of the puncture site after the operation identified a pseudoaneurysm of the superficial femoral artery on the right side, along with thrombosis within the aneurysm cavity. The suspected cause of the event was vascular access. It was reported that use of a large-diameter sheath may increase local soft tissue tension on the groin following insertion, potentially exacerbating local bleeding. On (b)(6), prothrombin injections were administered. On September 1, the patient underwent right lower limb arterial angiography, followed by the implantation of an artificial covered stent to isolate the pseudoaneurysm and endovascular angioplasty. The patient's hospitalization was prolonged as a result of the adverse event. The patient's outcome was reported as alleviated.\\It was further reported that an ultrasonic testing was performed and that no surgery for the implantation of the artificial covered stent was required.
Patient problems: Pseudoaneurysm; Thrombosis/Thrombus
MAUDE report 50926269 ↗DeathSep 28, 2026Adverse Event Without Identified Device or Use Problem
It was reported that a patient death occurred.\ \A FARADRIVE Steerable Sheath was selected for use. During the transseptal access procedure, while performing the Brockenbrough puncture, the device inadvertently traversed the space between the aorta and the left atrium rather than entering the intended puncture site. Subsequently, cardiac perforation occurred, resulting in cardiac tamponade. The physician reported an approximately 5 mm perforation on the roof side of the anterior wall of the left atrium. Pericardial drainage was performed to manage the cardiac tamponade. Due to deterioration of the patient's hemodynamic condition, percutaneous cardiopulmonary support (PCPS) was initiated. During PCPS management, an approximately 2 cm laceration of the inferior vena cava (IVC) occurred secondary to vascular perforation. The patient's condition became extremely critical, and open chest surgery was performed in an attempt to treat the injuries. No catheter was reportedly present within the heart at the time the health consequences were recognized. The physician reported that no resistance was encountered during guidewire manipulation and therefore it was believed that the sheath had be
Patient problems: Cardiac Perforation; Cardiac Tamponade; Perforation of Vessels
MAUDE report 50917781 ↗InjurySep 28, 2026Air/Gas in Device
FDA MedSun for Medwatch Notification:It was reported that that patient developed ST elevations during PFA "[Pulsed Field Ablation]" application of the RSPV that was sustained and associated with hypotension and inferior wall motion abnormalities. Suspect air bubbles from PFA sheath as cause of ST elevation. Post procedure, patient was found to have an air embolism in the brain causing ischemic injury (stroke).A FARADRIVE Sheath was selected for use to Atrial fibrillation. It was noted that the patient had preexisting Coronary heart disease. During the procedure, the Torflex catheter was exchanged for the FaraDrive sheath and inserted the Farawave catheter. The FaraDrive sheath was attached to continuous flush through a drip line that was carefully checked and prepped. The J-wire was positioned in each vein during the application. The patient developed transient ST elevations that self-resolved within 2 minutes after the basket pulses were delivered in the LSPV "[ Left Superior Pulmonary Vein]". No evidence of bubbles were seen in the cardiac chambers on ICE imaging and there appeared to be no bubbles in the sheaths. The patients hemodynamics were stable and it elected to move forwa
Patient problems: Air Embolism; Heart Block; Ischemia Stroke; Low Blood Pressure/ Hypotension; Non specific EKG/ECG Changes; Tachycardia
MAUDE report 50916734 ↗MalfunctionSep 24, 2026Leak/Splash; Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50888909 ↗MalfunctionSep 24, 2026Air/Gas in Device
It was reported that during a Pulse Field Ablation (PFA) procedure for the treatment of atrial fibrillation (AF), a FARADRIVE Steerable Sheath was selected for use. During withdrawal of the sheath from the left atrium, aspiration was performed as part of the removal process. At that time, a large quantity of air bubbles was observed mid-sheath while the sheath was being withdrawn from the body. All observed air bubbles were successfully aspirated, and the patient was assessed and monitored for signs and symptoms of air embolism and stroke. No air was observed in the patient, and the patient did not experience a stroke. The procedure was completed successfully using the device. No adverse patient effects or complications occurred, and the patient remained stable throughout and following the procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50888710 ↗MalfunctionSep 23, 2026Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50876289 ↗MalfunctionSep 23, 2026Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50874513 ↗MalfunctionSep 21, 2026Nonstandard Device; Device-Device Incompatibility
It was reported that during an atrial fibrillation ablation a FARADRIVE was selected for use. Groin was accessed with needle, starter guidewire was inserted, 12F coons dilator was inserted over the wire to pre-dilate, then it was removed and Faradrive with Versacross Connect for Faradrive was inserted over the wire. When the physician tried to insert the sheath and dilator through the groin it was stated that the sheath bunched up at the tip on the dilator because the transition was too abrupt. The dilator would not properly snap into the sheath. No abnormal force was used. No significant resistance was noticed when inserting the dilator. It occurred when attempting to insert the sheath through the access point. A new sheath and dilator were requested. While that was being prepped, a 14F coons dilator was used to further pre-dilate. The second sheath and dilator were inserted into the groin without incident. The procedure was completed without patient complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50839232 ↗MalfunctionSep 18, 2026Break
Faradrive steerable sheath broke no pt Harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50818320 ↗InjurySep 18, 2026Difficult to Advance
It was reported that patient experience Pericardial effusion and perforation.\\FARADRIVE Steerable Sheath Clear was selected for use. It was mentioned that once the transseptal was completed using versacross wire, the wire was sent into the left atrium. The Patient had thick septum, so was hard for the physician to push the sheath across. As per the physician opinion, Pericardial effusion occurred before ablation, when the physician pushed the Faradrive sheath too hard to cross, enabling the wire to puncture the LA. Perforation was noted in left atrium. Pericardiocentesis/TAP was performed and patient was transferred for Surgery into the OR (Operating Room). The Procedure was cancelled. Patient current status was unknown.
Patient problems: Cardiac Perforation; Low Blood Pressure/ Hypotension; Pericardial Effusion
MAUDE report 50818191 ↗MalfunctionSep 18, 2026Air/Gas in Device
It was reported that the sheath exhibited air bubbles in the irrigation tubing during aspiration.During a pulse field ablation (PFA) procedure to treat paroxysmal atrial fibrillation a FARADRIVE sheath was used. After insertion and introduction of the FARAWAVE catheter into the sheath aspiration was performed. Air bubbles were observed in the irrigation tubing. The sheath was replaced and the procedure was completed. No patient complications were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50817425 ↗MalfunctionSep 17, 2026Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50809241 ↗InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem
It was reported that Pericardial effusion occurred. FARADRIVE Steerable Sheath was selected for use during farapulse pulsed field ablation. It has been mentioned that patient arrived in flutter during an Atrial fibrillation ablation under deep sedation. The ICE Catheter was introduced into right atrium and then right ventricle and scanned noting a baseline effusion around the basal aspect of the right ventricle. Transeptal was performed without issue noted and visualized under ICE. The ICE Catheter was introduced into the left atrium. A decapolar Coronary Sinus catheter was then placed into the Coronary sinus. An eccentric activation pattern was noted and suspected roof dependent flutter. All 4 veins were isolated and the flutter persisted. A posterior wall isolation was performed. Termination occurred during an application in flower. Pressure dropped following conversion and anesthesia noted they were not responding to pressors. The Farapulse catheter and Faradrive sheath were removed from the LA. ICE was then pulled back to the Right atrium and scanned the pericardium. An increase in the pericardial effusion was noted to approximately 1cm. Hepranization was reversed and anesthesi
Patient problems: Atrial Flutter; Low Blood Pressure/ Hypotension; Pericardial Effusion
MAUDE report 50803247 ↗MalfunctionSep 17, 2026Fluid/Blood Leak
It was reported that during preparations for a FARAPULSE procedure to treat paroxysmal atrial fibrillation, upon unpacking and flushing a FARADRIVE sheath. a drip like leak of saline was observed coming from the hemostatic valve. Nothing had been inserted inside the sheath at this point. The catheter was replaced with a similar device, and the procedure was completed. No patient complications were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50802764 ↗InjurySep 16, 2026Adverse Event Without Identified Device or Use Problem
It was reported that a pericardial effusion occurred during a PVI ablation procedure in which a Faradrive sheath was being used. Following transseptal puncture, the Faradrive was positioned in the left atrium, and the physician was using a non-Boston Scientific catheter through the Faradrive to map the left atrium. During mapping, the anesthesiologist made a comment about the patient blood pressure. Intracardiac echocardiography (ICE) was subsequently performed and revealed a large pericardial effusion. No ablation had been performed at the time the effusion was identified. It appeared that the physician was initially unaware of when the perforation occurred. The procedure was canceled, and the patient was transferred to the operating room, where a cardiothoracic surgeon intervened. \\It was further reported that the perforation site had not been identified prior to the patient's transfer to the operating room. The transseptal puncture was performed using a non-Boston Scientific device.
Patient problems: Pericardial Effusion
MAUDE report 50796511 ↗MalfunctionSep 16, 2026Nonstandard Device; Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50788142 ↗MalfunctionSep 15, 2026Nonstandard Device; Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50777883 ↗InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem
During a Pulmonary Vein Isolation (PVI) Redo and Posterior Wall Isolation procedure, a FARADRIVE Steerable Sheath and FARAWAVE NAV Catheter were selected for use. All devices were prepared and used in accordance with the Instructions for Use (IFU), and no device malfunctions, system errors, OPAL alarms, air ingress, audible suction sounds, or other procedural abnormalities were reported. The procedure began with left atrial mapping using the FARAWAVE NAV catheter to assess the patients substrate, which demonstrated reconnection of the right pulmonary veins (RPVs). Pulsed Field Ablation (PFA) was subsequently delivered to the RPVs and electrically active posterior wall substrate. Following completion of the ablation lesions, a post-ablation left atrial map was performed to confirm lesion effectiveness and define the mitral isthmus anatomy for the planned mitral isthmus (MI) ablation. A total of 20 PFA applications were delivered, with the final application occurring at 12:06 PM. While performing post-ablation mapping of the left atrium, approximately four minutes after the last PFA application, ST-segment elevation was observed at 12:10 PM, most prominently in leads DII and DIII. Th
Patient problems: Non specific EKG/ECG Changes
MAUDE report 50765226 ↗MalfunctionSep 14, 2026Fluid/Blood Leak; Device Contamination with Body Fluid
It was reported that FARADRIVE Steerable Sheath was selected for use during paroxysmal atrial fibrillation ablation. It has been mentioned that there was a damage to hemostatic valve. During the catheter insertion, blood leakage was observed from the hemostatic valve. No air ingress was seen. No patient complications occurred. The procedure was completed using different device (same model).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50752815 ↗InjurySep 11, 2026Adverse Event Without Identified Device or Use Problem
It was reported that a patient experienced Blood Loss and Back Pain. \ \During a Pulsed Field Ablation (PFA) procedure to treat Paroxysmal Atrial Fibrillation, a Faradrive was selected for use. Ablation was completed successfully, and the physician was closing and started to pull the Faradrive sheath when he noticed additional bleeding. The physician held pressure, but bleeding continued. Additional protamine was given to return Activated Clotting Time (ACT) to non-therapeutic levels. The sheath was re-inserted into the vein, and the bleed subsided. ACT was returned to a sub-200 level, and a non-BSC device was used. Additional protamine was given to help return the patient to non-therapeutic ACT levels to help slow bleeding. The access site complication occurred on the same side where the BSC device was inserted. The patient was moved to the Intensive Care Unit (ICU) while intubated due to back pain and inability to lay flat for an extended period of time. No additional medication was given for the pain, and no known intervention was performed due to the pain. The procedure was successfully completed with the same device. The patient is expected to fully recover from the event.
Patient problems: Back pain; Hemorrhage/Blood Loss/Bleeding
MAUDE report 50741736 ↗MalfunctionSep 10, 2026Leak/Splash; Device Contamination with Body Fluid
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50724640 ↗MalfunctionSep 10, 2026Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50722155 ↗MalfunctionSep 10, 2026Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50718405 ↗InjurySep 10, 2026Adverse Event Without Identified Device or Use Problem
It was reported that a patient experienced Cerebral infarction (Ischemia, Stroke) and thrombus.\\It was reported that during procedure for a Pulse Field Ablation (PFA) procedure a Farastar Recording System Module NAM, Farastar Mapping System Module, Farastar PFA Generator NAM a Faradrive steerable sheath, a Farawave Ablation Catheter and two Farawave NAV Ablation Catheter were selected for use. It was noted that a Bad Head error (BAD HDR error) occurred immediately after connecting the catheter. As troubleshooting, it was attempted to disconnect and restart the MSM and RSM multiple times. Non-Nav catheter was replaced for a Nav catheter, which resolved the error, nonetheless with energization error 301 started happening frequently. The catheter was removed from the body to adjust shape and reinsert, error still occurred. As troubleshooting, it was attempted to restart MSM and RSM, and to replace catheter and AUX cable; no success. Second Nav catheter went through the same 301 error. Second NAV catheter was replaced for the first Non-Nav catheter used during the procedure. When this catheter Non-NAV was plugged directly into the console and energized, it energized without any issues
Patient problems: Ischemia Stroke; Thrombosis/Thrombus
MAUDE report 50706308 ↗
FDA clearance and classification
The FDA regulates product code DRA (catheter, steerable) under 21 CFR 870.1280 as a Class II device. 21 CFR 870.1280 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the FARADRIVE™ M004PF21M402
Has the FARADRIVE™ M004PF21M402 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2756-2026 (Class II), began on Jun 11, 2026: Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation.… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the FARADRIVE™ M004PF21M402?
The FDA's MAUDE database lists 2,303 adverse event reports for this device, received between Apr 18, 2024 and Sep 30, 2026. In those reports, the most common specific device problems are “Air/Gas in Device” (1,383), “Leak/Splash” (450) and “Device Contamination with Body Fluid” (235); the most common patient problems are “Low Blood Pressure/ Hypotension” (90), “Pericardial Effusion” (90) and “Non specific EKG/ECG Changes” (84). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the FARADRIVE™ M004PF21M402?
16 reports are classified as deaths, 309 as injuries and 1,978 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the FARADRIVE™ M004PF21M402, and is it still sold?
The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand FARADRIVE™, model M004PF21M402). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, STEERABLE devices
- MITRACLIP SGC0701Abbott Vascular Inc. · CATHETER, STEERABLE17 deaths962 reports
- STEERABLE GUIDE CATHETER SGC0301Abbott Vascular Inc. · CATHETER, STEERABLE5 deaths390 reports
- FlexCath Advance™ 4FC12Medtronic, Inc. · CATHETER, STEERABLE6 deaths167 reports
- FlexCath Contour™ 10FCC13Medtronic, Inc. · CATHETER, STEERABLE2 deaths115 reports
- FlexCath Advance™ 4FC12Medtronic, Inc. · CATHETER, STEERABLE4 deaths61 reports
- FlexCath Advance™ 4FC12Medtronic, Inc. · CATHETER, STEERABLE4 deaths60 reports
- FlexCath Advance™ 4FC12Medtronic, Inc. · CATHETER, STEERABLE59 reports
- DLP® 12116Medtronic, Inc. · CATHETER, STEERABLE1 recall8 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page