Skip to content

FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate… recallZ-2756-2026 · Boston Scientific Corporation · initiated Jun 11, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

FDA root cause category: Use error

What the company is doing

An Important Medical Device Product Advisory notification letter dated 6/11/26 was sent to customers. Actions " Follow the revised IFU provided in Appendix 1. " Forward this notice to relevant personnel and any facilities to which the product was transferred. " Complete and return the enclosed Acknowledgement Form. " Report any adverse events or quality concerns associated with use of these devices to Boston Scientific via email at [email protected] or FDA s MedWatch Adverse Event Reporting program [www.fda.gov/MedWatch/report.htm or 1.800.FDA.1088 (332.1088)]. If you require additional assistance or more information regarding this communication, please contact your local Boston Scientific representative.

Recall details

Recall number
Z-2756-2026
Status
Open, Classified
Initiated by the company
Jun 11, 2026
Posted by the FDA
Jul 20, 2026
Recalling firm
Boston Scientific Corporation
Quantity
508,151 units
Product code
DRA
FDA submissions
K233248
Distribution
Global Distribution. US Nationwide.

Product description

FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01

Code information (models, lots, serial numbers)

Show the full code information (3k characters)

UDI-DI: 00191506043131; Lots: CL12695, CL12696, CL12699, CL12700, CL12703, CL12808, CL12816, CL12817, CL12823, CL12834, CL12835, CL12836, CL12853, CL12860, CL12882, CL12883, CL12884, CL12891, CL12907, CL12908, CL12909, CL12919, CL12949, CL12950, CL12969, CL12976, CL12979, CL12980, CL13005, CL13010, CL13011, CL13012, CL13018, CL13032, CL13033, CL13034, CL13037, CL13056, CL13057, CL13058, CL13074, CL13075, CL13076, CL13077, CL13097, CL13098, CL13099, CL13100, CL13104, CL13115, CL13116, CL13117, CL13118, CL13119, CL13135, CL13136, CL13137, CL13138, CL13139, CL13156, CL13157, CL13191, CL13192, CL13193, CL13194, CL13197, CL13208, CL13209, CL13210, CL13215, CL13216, CL13217, CL13224, CL13229, CL13230, CL13257, CL13258, CL13288, CL13289, CL13337, CL13338, CL13365, CL13366, CL13384, CL13398, CL13399, CL13419, CL13420, CL13423, CL13467, CL13469, CL13470, CL13471, CL13472, CL13517, CL13525, CL13526, CL13527, CL13528, CL13535, CL13536, CL13541, CL13550, CL13551, CL13552, CL13553, CL13555, CL13564, CL13565, CL13566, CL13571, CL13574, CL13575, CL13576, CL13598, CL13599, CL13616, CL13637, CL13642, CL13643, CL13644, CL13665, CL13666, CL13667, CL13668, CL13704, CL13705, CL13706, CL13715, CL13716, CL13721, CL13734, CL13735, CL13736, CL13761, CL13762, CL13763, CL13798, CL13799, CL13800, CL13822, CL13825, CL13826, CL13835, CL13836, CL13837, CL13858, CL13859, CL13860, CL13912, CL13913, CL13914, CL13927, CL13928, CL13929, CL13942, CL13944, CL13945, CL13974, CL13975, CL13981, CL13990, CL13991, CL13992, CL14027, CL14028, CL14029, CL14040, CL14041, CL14042, CL14053, CL14054, CL14055, CL14079, CL14080, CL14081, CL14122, CL14123, CL14124, CL14139, CL14140, CL14141, CL14169, CL14172, CL14173, CL14174, CL14188, CL14189, CL14190, CL14227, CL14228, CL14229, CL14250, CL14251, CL14252, CL14288, CL14289, CL14290, CL14347, CL14348, CL14377, CL14378, CL14398, CL14399, CL14400, CL14440, CL14441, CL14442, CL14489, CL14490, CL14492, CL14493, CL14496, CL14502, CL14505, CL14506, CL14520, CL14583, CL14584, CL14585, CL14634, CL14636, CL14637, CL14671, CL14672, CL14673, CL14674, CL14698, CL14699, CL14700, CL14712, CL14713, CL14714, CL14715, CL14751, CL14752, CL14753, CL14771, CL14772, CL14773, CL14774, CL14815, CL14816, CL14817, CL14818, CL14856, CL14857, CL14858, CL14859, CL14872, CL14873, CL14874, CL14930, CL14931, CL14932, CL14936, CL14937, CL14938, CL14970, CL14971, CL14972, CL15006, CL15007, CL15010, CL15011, CL15068, CL15069, CL15070, CL15080, CL15089, CL15110, CL15130, CL15131, CL15132, CL15194, CL15195, CL15196, CL15241, CL15242, CL15243, CL15273, CL15274, CL15275, CL15276, CL15319, CL15320, CL15321, CL15322, CL15359, CL15360, CL15361, CL15362, CL15398, CL15399, CL15400, CL15401, CL15428, CL15429, CL15430, CL15431, CL15451, CL15475, CL15476, CL15477, CL15491, CL15492, CL15493, CL15494, CL15525, CL15540, CL15541, CL15548, CL15549, CL15570, CL15583, CL15584, CL15599, CL15603, CL15617, CL15641, CL15642, CL15718, CL15756, CL15769, CL15770, CL15777

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates