AVANOS 40778Recalls & reported problems

Avanos Medical, Inc. · ON-Q* Pain Relief System T-bloc* Echogenic Continuous Stimulating Nerve Block Tray · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00193494407780), received between Mar 16, 2022 and Sep 13, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Break3
  • Difficult to Remove1

Patient problems

  • Foreign Body In Patient2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 13, 2023Break; Break

    IT WAS REPORTED "AFTER PLACING CATHETER AND CONNECTING THE NEXUS CONNECTOR, THE CATHETER SHEARED AT NEXUS CONNECTION SITE. THE EXTERIOR OF CATHETER STRETCHED FROM THE INTERIOR SPIROL (SPIRAL) OF CATHETER. THE DEFECTIVE CATHETER WAS REMOVED FROM PATIENT." THERE WAS NO REPORTED INJURY.

    Patient problems: Foreign Body In Patient

    MAUDE report 17741014 ↗
  • MalfunctionMar 16, 2022Break; Break

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH TWO SEPARATE PATIENTS. THIS IS THE SECOND OF TWO REPORTS. REFER TO 2026095-2022-00013 FOR THE FIRST REPORT. IT WAS REPORTED THERE WAS AN INCIDENT OF POSSIBLE CATHETER SHEARING UPON REMOVAL. THERE WAS NO REPORT OF MEDICAL INTERVENTION. THE DATE OF SURGERY (B)(6) 2021. ADDITIONAL INFORMATION RECEIVED 08MAR2022 REPORTED THAT AT SOME POINT CATHETER A REMOVAL WAS ATTEMPTED AND APPROXIMATELY 4 CENTIMETERS OF THE CATHETER WAS LEFT INSIDE THE PATIENT. THERE WAS NO REPORTED INJURY.

    Patient problems: Foreign Body In Patient

    MAUDE report 13777209 ↗
  • MalfunctionMar 16, 2022Break; Difficult to Remove; Break; Difficult to Remove

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS. THIS IS THE FIRST OF TWO REPORTS. REFER TO 2026095-2022-00014 FOR THE SECOND REPORT. IT WAS REPORTED THERE WAS AN INCIDENT OF POSSIBLE CATHETER SHEARING UPON REMOVAL. THERE WAS NO REPORT OF MEDICAL INTERVENTION. THE DATE OF SURGERY (B)(6) 2021. ADDITIONAL INFORMATION RECEIVED 08MAR2022 REPORTED THE PATIENT RETURNED TO THE HOSPITAL TO HAVE THE CATHETER REMOVED AND IT APPEARED TO BE NOT REMOVEABLE; IT WAS EXTRACTED WITH A PIECE OF THE RETAINED CATHETER LEFT BEHIND. THE PATIENT RETURNED FOR A FOLLOW UP APPOINTMENT AND IT WAS NOTED THAT THE PATIENT DID NOT EXPERIENCE ANY KNOWN COMPLICATIONS OR INFECTION; THE PATIENT WILL CONTINUE TO FOLLOW-UP AS PLANNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13777204 ↗

FDA clearance and classification

The FDA regulates product code CAZ (anesthesia conduction kit) under 21 CFR 868.5140 as a Class II device. 21 CFR 868.5140 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the AVANOS 40778

Has the AVANOS 40778 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00193494407780), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the AVANOS 40778?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Mar 16, 2022 and Sep 13, 2023. In those reports, the most common specific device problems are “Break” (3) and “Difficult to Remove” (1); the most common patient problem is “Foreign Body In Patient” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the AVANOS 40778?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the AVANOS 40778 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the AVANOS 40778, and is it still sold?

The FDA's device database lists Avanos Medical, Inc. as the company responsible for it (brand AVANOS, model 40778). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Anesthesia conduction kit devices

All Anesthesia conduction kit devices →