EXCELSIUS GPS 6143.2692Recalls & reported problems

Globus Medical, Inc. · CREO MIS Driver Outer Sleeve, 30mm, GPS · Class 2

All models of EXCELSIUS GPS (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00193982278724), received between Oct 15, 2025 and Oct 15, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Human-Device Interface Problem” (1 report).

1 other recall cover devices cleared under the same FDA submission (K171651); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Human-Device Interface Problem1

Patient problems

  • Spinal Column Injury1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 15, 2025Human-Device Interface Problem; Human-Device Interface Problem

    IT WAS REPORTED THAT DURING A CASE USING THE 30 MM MIS GPS DRIVER AND WHILE PUTTING IN SCREWS THE SURGEON FIGURED OUT THAT ON THE NEW 2 PIECE STYLE GPS DRIVERS, THEY CAN SLIDE DOWN THE ARRAY WITHOUT KNOWING ON NAVIGATION. THEY ENDED UP PUTTING IN A SCREW FURTHER THAN WE KNEW ON THE SCREEN AND DIDN'T CATCH IT TILL A CONFIRMATION SPIN OF THE SCREWS. THEY ALSO MEASURED THIS AFTER THE CASE AND THE DRIVER CAN SLIDE 8 MM FURTHER DOWN THE TUBE IF YOU ACCIDENTALLY HIT THE BUTTON IN THE MIDDLE OF THE ARRAY.

    Patient problems: Spinal Column Injury

    MAUDE report 23294891 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K171651. One submission often covers many models, so check the product and lot details in each record.

  • Z-1007-2024 · Dec 15, 2023 · Open, Classified
    ExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes, Model No. 6143.2844

Other Orthopedic stereotaxic instrument devices

All Orthopedic stereotaxic instrument devices →