Skip to content

AUTOBAHN 1257.5036SRecalls & reported problems

Globus Medical, Inc. · AUTOBAHN EVO Retrograde Femoral Nail, 10x360mm, 5°, Ti · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00193982498399), received between Sep 25, 2026 and Sep 25, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Migration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration1

Patient problems

  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 25, 2026Migration

    It was reported that during patient's follow-up visit on (b)(6) 2026, imagine showed that the distal screw on the proximal end of the nail and the most distal screw on the distal end of the nail wasbacking out. Patient reported symptomatic pain and could "feel it shifting." Both screws were removed during the revision surgery. The surgeon implanted 2 new screws, one in the medial oblique slot and a proximal screw in the lateral to medial slot.

    Patient problems: Pain

    MAUDE report 50900594 ↗

Other NAIL, FIXATION, BONE devices

All NAIL, FIXATION, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page