TRIGEN 71676627Recalls & reported problems
Smith & Nephew, Inc. · TRIGEN INTERTAN 1.5 13MM X 42CM 130DEGREE LEFT · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885556131633), received between Apr 11, 2022 and May 11, 2022 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
9 other recalls cover devices cleared under the same FDA submission (K040212, K210980); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Apr 22, 2022Open, Classified
Right nails were anodized, marked, and labelled as left nails and vice versa
Z-1133-2022 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Activation, Positioning or Separation Problem1
- Device Markings/Labelling Problem1
- Packaging Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Perforation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 11, 2022Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT, DURING INTERNAL FIXATION, AN INTERTAN 1.5 13X42 130D LT WAS OPENED, AND IT APPEARED NORMAL. WHEN ATTACHED TO THE TARGETER, IT WOULD ONLY ATTACH IN THE ¿REVERSE¿ AS A RIGHT NAIL WOULD. FURTHER, INSPECTION SHOWED A LIME GREEN NAIL, CLEARLY MARKED LEFT THAT APPEARS TO ACTUALLY BE A RIGHT NAIL INCORRECTLY COLORED AND LABELED. REPLACEMENT/ALTERNATE HAD TO BE RETRIEVED. SURGERY WAS RESUMED AFTER A NON-SIGNIFICANT DELAY THE PROCEDURE WAS FINISHED WITH A SMITH AND NEPHEW BACK UP DEVICE. THE PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14355662 ↗InjuryApr 11, 2022Activation, Positioning or Separation Problem; Packaging Problem
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY IN WHICH THE PATIENT NEEDED A LEFT NAIL, AN INTERTAN 1.5 13X42 130D LEFT NAIL WAS SELECTED AND IMPLANTED, BUT IT DID NOT FIT CORRECTLY. THE NAIL PENETRATED THROUGH THE ANTERIOR CORTEX SO IT WAS EXPLANTED. UPON INSPECTION, IT WAS NOTICED THAT THE NAIL CONFIGURATION WAS RIGHT, BUT IT HAD BEEN COLOURED, MARKED AND BOXED AS LEFT. THE PROCEDURE WAS COMPLETED, AFTER 30 MIN, BY USING AN INTERTAN 1.5 13X44 130D LEFT NAIL.
Patient problems: Perforation
MAUDE report 14079904 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K040212, K210980. One submission often covers many models, so check the product and lot details in each record.
- Z-2456-2026 · Mar 16, 2026 · Open, ClassifiedTRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
- Z-1806-2025 · Mar 18, 2025 · Open, ClassifiedMETA-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
- Z-2444-2023 · Jul 24, 2023 · Open, ClassifiedTRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT, REF 71647340, Femur nail
- Z-2443-2023 · Jul 24, 2023 · Open, ClassifiedTRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT, REF 71647336, Femur nail
- Z-0772-2021 · Nov 2, 2020 · TerminatedBone fixation nail. TRIGEN INTERTAN - Product Usage: indicated for fractures of the femur including simple shaft fractures, comminuted shaft fractures, spiral shaft fractures, long
- Z-0097-2018 · Oct 24, 2017 · Open, Classifiedsmith&nephew TRIGEN INTERTAN 10S, REF 71675384, 10 MM X 18 CM NAIL, 130o (degrees), SET SCREW PREASSEMBLED IN NAIL
- Z-2258-2014 · Jul 2, 2014 · TerminatedTRIGEN(R) INTERTAN(TM) 10MM X 18 CM NAIL, 125o, REF 71675201, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 18 CM NAIL, 125o, REF 71
- Z-1636-2014 · Apr 4, 2014 · TerminatedTRIGEN INTERTAN NAIL: (1) REF 71675201, 10 MM X 18 CM, 125 degree, (2) REF 71675202, 11.5 MM X 18 CM, 125 degree, (3) REF 71675203, 13 MM X 18 CM, 125 degree, (4) REF 71675204,
- Z-1637-2014 · Apr 4, 2014 · TerminatedTRIGEN INTERTAN 1.5 NAIL: (1) REF 71676515, 10 MM X 10 CM, 125 degree, LEFT, (2) REF 71676556, 11.5 MM X 36 CM, 125 degree, RIGHT, (3) REF 71676558, 11.5 MM X 38 CM, 125 degree
Other NAIL, FIXATION, BONE devices
- NA 03.043.024Synthes GmbH · NAIL, FIXATION, BONE234 reports
- NA 03.010.440Synthes GmbH · NAIL, FIXATION, BONE70 reports
- TRIGEN 71676628Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall0 reports
- NA 03.043.018Synthes GmbH · NAIL, FIXATION, BONE33 reports
- TRIGEN 71675383Smith & Nephew, Inc. · NAIL, FIXATION, BONE29 reports
- NA 03.010.441Synthes GmbH · NAIL, FIXATION, BONE27 reports
- NA 03.010.435Synthes GmbH · NAIL, FIXATION, BONE24 reports
- NA 04.043.240SSynthes GmbH · NAIL, FIXATION, BONE24 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.