B125M B125M Patient MonitorRecalls & reported problems
Ge Medical Systems Information Technologies, Inc. · B125M VSP4 ATO MODEL · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00195278490247), received between Oct 30, 2025 and Oct 30, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Low Readings” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 26, 2025Class IIOpen, Classified
Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.
Z-0455-2026 · Root cause: Component change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Low Readings1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryOct 30, 2025Low Readings; Low Readings
IT WAS REPORTED THAT THE PATIENT MONITOR PROVIDED INCORRECT NIBP READINGS FOR WHICH THE PATIENT WAS UNNECESSARILY ADMINISTERED MEDICATION. THE PATIENT DID NOT SUSTAIN ANY INJURY DUE TO THE ALLEGED UNNECESSARY MEDICATION.
Patient problems: Insufficient Information
MAUDE report 23429615 ↗
Other Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase devices
- Alaris 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall2 deaths8,444 reports
- Alaris 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall607 reports
- BD Alaris™ ETCO2 Module, Model 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE485 reports
- BD Alaris™ ETCO2 Module, Model 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE471 reports
- NA GF-210RNihon Kohden Corporation · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE386 reports
- Trilogy IT2110X21BRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls30 reports
- Trilogy Trilogy Evo Ventilator, GermanyRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls26 reports
- Trilogy Evo Trilogy Evo Ventilator with oxygen blending model, GermanyRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.