Trilogy Evo Trilogy Evo Ventilator with oxygen blending model, GermanyRecalls & reported problems
Respironics, Inc. · Trilogy Evo Ventilator system with oxygen blending module, Germany · Class 2
All models of Philips Trilogy Ventilator (90) · Trilogy Evo (20)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
5 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 5
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00606959054103), received between Sep 21, 2020 and Jan 31, 2025 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (4 reports).
5 FDA recall records name this device's identifier, 5 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 7, 2024Class IOpen, Classified
Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
Z-0290-2025 · Root cause: Device Design · Recall details · FDA record ↗
- Jul 16, 2024Class IOpen, Classified
Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls.
Z-2944-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Mar 19, 2024Class IOpen, Classified
Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.
Z-1503-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Nov 27, 2023Class IIOpen, Classified
The following was missing from the Contraindications Statement: The AVAPS-AE therapy mode is contraindicated for invasive use and patients less than 10 kg. The AVAPS feature is contraindicated for patients less than 10 kg. Potential harms if devices are used on contraindicated patients include barotrauma, hypoventilation/hypercapnia, and rebreathing of excessive CO2.
Z-0638-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Mar 1, 2023Class IOpen, Classified
The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually delivering. This may lead to under delivery of oxygen.
Z-1287-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Mechanical Problem4
- Circuit Failure2
- Computer Software Problem2
- Contamination1
- Device Sensing Problem1
- Electrical /Electronic Property Problem1
- Failure to Calibrate1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 31, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE CONDITION AND/OR VENT SERVICE REQUIRED CONDITION HAD OCCURRED ON THE DEVICE. THERE WAS NO HARM OR INJURY REPORTED. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21286636 ↗MalfunctionJan 28, 2025Device Sensing Problem; Device Sensing Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A PRESSURE SENSOR MISMATCH AND VENTILATOR INOPERATIVE CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21252381 ↗MalfunctionNov 26, 2024Failure to Calibrate; Failure to Calibrate
THE MANUFACTURER RECEIVED INFORMATION ALLEGING THE DEVICE FAILED A PNEUMATIC TEST. THERE WAS NO HARM OR INJURY REPORTED. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20779807 ↗MalfunctionJul 19, 2023Circuit Failure; Contamination; Mechanical Problem; Circuit Failure; Contamination; Mechanical Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR WAS HAVING ERROR IN THE O2 SENSOR. THE DEVICE WAS NOT IN PATIENT USE. THE VENTILATOR WAS EVALUATED BY A THIRD-PARTY SERVICE CENTER AND THE CUSTOMER¿S COMPLAINT WAS CONFIRMED. THE DEVICE'S OXYGEN BLENDING MODULE BOARD, IN-LINE FILTER AND MACHINE FLOW SENSOR NEEDS TO BE REPLACED TO ADDRESS THE ISSUES. THERE WAS EVIDENCE OF DUST AND DIRT CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17356184 ↗MalfunctionFeb 15, 2022Circuit Failure; Mechanical Problem; Circuit Failure; Mechanical Problem
A VENTILATOR WAS RETURNED TO A THIRD PARTY SERVICE CENTER FOR EVALUATION. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S OXYGEN BLENDING MODULE ASSEMBLY WAS REPLACED TO ADDRESS THE ISSUE. A "VENTILATOR INOPERATIVE" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SYSTEM BOARD WAS REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13535170 ↗MalfunctionMar 9, 2021Computer Software Problem; Computer Software Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED. THE DEVICE'S SOFTWARE WAS RELOADED AT THE REPORTING FACILITY TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11446952 ↗MalfunctionMar 9, 2021Computer Software Problem; Computer Software Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED. THE PATIENT WAS PLACED ON ANOTHER DEVICE WITHOUT INCIDENT. IT WAS INITIALLY REPORTED, THE DEVICE DID NOT ALARM WHEN THE VENTILATOR INOPERATIVE CONDITION OCCURRED. ADDITIONAL INFORMATION WAS RECEIVED THAT INDICATES THE DEVICE DID ALARM AT THE TIME OF THE EVENT. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT COULD NOT BE DUPLICATED. REVIEW OF THE DEVICE'S DOWNLOADED EVENT LOGS INDICATED A VENTILATOR INOPERATIVE CONDITION DID OCCUR ON (B)(6) 2021. THE DEVICE'S SOFTWARE WAS RELOADED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11446333 ↗MalfunctionFeb 23, 2021Mechanical Problem; Mechanical Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S MACHINE FLOW SENSOR WAS REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11367150 ↗MalfunctionSep 21, 2020Mechanical Problem; Mechanical Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO THE ACTIVE EXAHALATION CONTROL MODULE. THERE WAS NO HARM OR INJURY REPORTED. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S PROPORTIONAL VALVE AND SOLENOID VALVE WERE REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10560615 ↗
Other ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE devices
- Alaris 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall2 deaths8,444 reports
- Alaris 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall607 reports
- BD Alaris™ ETCO2 Module, Model 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE485 reports
- BD Alaris™ ETCO2 Module, Model 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE471 reports
- NA GF-210RNihon Kohden Corporation · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE386 reports
- Trilogy IT2110X21BRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls30 reports
- Trilogy Trilogy Evo Ventilator, GermanyRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls26 reports
- Trilogy Evo EU2100X19Respironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page