CONTOUR® NEXT ONE 9763Recalls & reported problems

Ascensia Diabetes Care US Inc. · Blood Glucose Monitoring System. 1 kit. For self-testing, monitoring and management of blood glucose. Kit contains: CONTOUR® NEXT ONE Meter; 10 CONTOUR® NEXT Blood Glucose Test Strips; Lancing Device; 10 MICROLET® Colored Lancets (provided sterile); User Guide (including CONTOUR DIABETES app download instructions); Quick Reference Guide; Carrying Case; Web Registration / Warranty Card; CONTOUR® NEXT Control Solution. · Class 2

All models of CONTOUR NEXT ONE (6)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Contour Next One blood glucose meter appears in 120 reports: no deaths, 5 injuries and 115 malfunctions. Most reports describe test results that were not saved in the meter's memory, and a few describe readings that seemed high or incorrect. No recalls name this device. Reports fell to 1 in the last 12 months from 8 in the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
120
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 120 adverse event reports that list this exact device (DI 00301939763012), received between Dec 6, 2017 and Nov 4, 2025 — classified as 5 injuries, 115 malfunctions.

The most frequently reported specific device problem is “Missing Test Results” (88 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 4 reportsJan 2024: 1 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Missing Test Results88
  • Incorrect, Inadequate or Imprecise Result or Readings9
  • Nonstandard Device9
  • High Readings7
  • Image Display Error/Artifact4
  • Display Difficult to Read2
  • No Device Output2

Patient problems

  • No Clinical Signs, Symptoms or Conditions69
  • No Known Impact Or Consequence To Patient41
  • Dizziness4
  • Hypoglycemia3
  • Cognitive Changes1
  • Complaint, Ill-Defined1
  • Fever1
  • Headache1

Recent adverse event reports

The 25 most recent of 120 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 4, 2025Missing Test Results; Missing Test Results

    THE CUSTOMER CONTACTED ASCENSIA CUSTOMER SERVICE TO SEEK ASSISTANCE WITH HIS CONTOUR® NEXT ONE METER. DURING TROUBLESHOOTING, IT WAS FOUND THAT THE CUSTOMER'S BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. REPLACEMENT METER KIT AND TEST STRIPS WERE SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23467282 ↗
  • MalfunctionJul 15, 2025Missing Test Results; Missing Test Results

    THE CUSTOMER CONTACTED ASCENSIA CUSTOMER SERVICE TO SEEK ASSISTANCE WITH HER CONTOUR® NEXT ONE METER. DURING TROUBLESHOOTING, IT WAS FOUND THAT THE CUSTOMER'S BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE METER'S MEMORY. THE CUSTOMER STATED THAT SHE WAS FEELING DIZZY. ADDITIONALLY, THE CUSTOMER STATED THAT SHE HAD RECENTLY REPLACEMENT THE METER'S BATTERIES. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: Dizziness

    MAUDE report 22509350 ↗
  • MalfunctionMay 1, 2025Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR® NEXT ONE METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21942041 ↗
  • MalfunctionApr 25, 2025Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR® NEXT ONE METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21909264 ↗
  • MalfunctionApr 4, 2025Nonstandard Device; Nonstandard Device

    A CUSTOMER FROM UNITED KINGDOM (UK) REPORTED THAT THE CONTOUR® NEXT ONE METER PURCHASED IN UK FROM AMAZON WEBSITE WAS DISPLAYING THE UNITS OF MEASUREMENT ACCOMPANIED WITH THE BLOOD GLUCOSE READINGS IN MG/DL INSTEAD OF MMOL/L. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER RETURNED THE SUSPECTED DEVICE TO AMAZON. THEREFORE, THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21777428 ↗
  • InjuryMar 3, 2025High Readings; High Readings

    AN ADVOCATE CALLED TO REPORT THAT THE CUSTOMER OBTAINED A BLOOD GLUCOSE READING OF 300 MG/DL WITH THE CONTOUR® NEXT ONE METER. THE CUSTOMER TOOK 5 UNITS OF INSULIN BASED ON THE METER'S READINGS. THE CUSTOMER WAS EXPERIENCING SYMPTOMS SUCH AS BODY ACHE AND FEVER. SINCE THE CUSTOMER WAS NOT FEELING WELL EVEN AFTER TAKING INSULIN, SHE WENT TO THE HOSPITAL WHERE A READING OF 153 MG/DL WAS OBTAINED WITH THE HOSPITAL'S METER. THE CUSTOMER WAS GIVEN FLUIDS IN THE HOSPITAL. A FEW HOURS LATER, A LABORATORY TESTING WAS PERFORMED AND A READING OF 163 MG/DL WAS OBTAINED. NO FURTHER EVENT DETAILS WERE PROVIDED. THE CUSTOMER DISCONNECTED THE CALL. THEREFORE, A REPLACEMENT DEVICE COULD NOT BE OFFERED AND THE DEVICE COULD NOT BE REQUESTED BACK FOR EVALUATION. SINCE THE STRIP INFORMATION WAS NOT PROVIDED, THIS REPORT WILL BE SUBMITTED UNDER THE METER INFORMATION.

    Patient problems: Fever; Pain

    MAUDE report 21504834 ↗
  • MalfunctionJan 2, 2025Missing Test Results; Missing Test Results

    THE CUSTOMER CALLED ASCENSIA CUSTOMER SERVICE TO SEEK ASSISTANCE WITH THE CONTOUR® NEXT ONE METER. DURING THE TROUBLESHOOTING, IT WAS FOUND THAT THE CUSTOMER'S BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21057246 ↗
  • MalfunctionJan 2, 2025Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR® NEXT ONE METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21057238 ↗
  • MalfunctionDec 11, 2024Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HE REPLACED THE BATTERY OF THE CONTOUR NEXT ONE METER ON (B)(6) 2024. THE CUSTOMER'S BLOOD GLUCOSE READINGS PRIOR TO (B)(6) 2024 WERE NOT STORED IN THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20903590 ↗
  • MalfunctionJul 11, 2024Missing Test Results; Missing Test Results

    AN ADVOCATE CALLED ON BEHALF OF THE CUSTOMER TO REPORT THAT THEIR BLOOD GLUCOSE READINGS WERE NOT STORED IN THE MEMORY OF THE CONTOUR NEXT ONE METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE ADVOCATE WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19719304 ↗
  • MalfunctionJul 1, 2024Missing Test Results; Missing Test Results

    THE CUSTOMER CALLED ASCENSIA CUSTOMER SERVICE TO SEEK ASSISTANCE WITH HIS CONTOUR NEXT ONE METER. DURING TROUBLESHOOTING, IT WAS FOUND THAT ONE OF HIS BLOOD GLUCOSE READINGS WAS NOT STORED IN THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19643532 ↗
  • MalfunctionJun 28, 2024Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT ONE METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19632350 ↗
  • MalfunctionApr 15, 2024Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT ONE METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19105251 ↗
  • MalfunctionFeb 6, 2024High Readings; High Readings

    THE CUSTOMER REPORTED THAT THE BLOOD GLUCOSE READINGS OBTAINED WITH THE CONTOUR NEXT ONE METER WERE APPROXIMATELY 50 MG/DL HIGHER THAN THE LABORATORY RESULTS. NO SPECIFIC READINGS WERE PROVIDED. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR EVALUATION. SINCE THE STRIP INFORMATION WAS NOT PROVIDED, THIS REPORT WILL BE SUBMITTED UNDER THE METER INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18650168 ↗
  • MalfunctionFeb 1, 2024Missing Test Results; Missing Test Results

    AN ADVOCATE CALLED ON BEHALF OF THE CUSTOMER TO REPORT THAT THEIR BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT ONE METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE ADVOCATE WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18622495 ↗
  • MalfunctionJan 2, 2024Missing Test Results; Missing Test Results

    AN ADVOCATE CALLED ON BEHALF OF THE CUSTOMER, TO REPORT THAT ONE OF THEIR BLOOD GLUCOSE READINGS WAS NOT STORED IN THE MEMORY OF THE CONTOUR NEXT ONE METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE ADVOCATE WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18425918 ↗
  • MalfunctionDec 11, 2023Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT ONE METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18301070 ↗
  • MalfunctionDec 11, 2023Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT ONE METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18301056 ↗
  • MalfunctionDec 4, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED THAT THE BLOOD GLUCOSE RESULTS OBTAINED WITH THE CONTOUR NEXT ONE METER AND LABORATORY TESTING WITHIN 10 MINUTES WERE DIFFERENT BY 30 MG/DL TO 45 MG/DL. NO SPECIFIC READINGS WERE PROVIDED. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER. SINCE THE STRIP INFORMATION WAS NOT PROVIDED, THIS REPORT WILL BE SUBMITTED UNDER THE METER INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18256333 ↗
  • MalfunctionDec 1, 2023Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT ONE METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18247501 ↗
  • MalfunctionAug 7, 2023Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT ONE METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17470427 ↗
  • MalfunctionAug 1, 2023Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT ONE METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17435141 ↗
  • MalfunctionJul 25, 2023Incorrect, Inadequate or Imprecise Result or Readings; Missing Test Results; Incorrect, Inadequate or Imprecise Result or Readings; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT ONE METER. ADDITIONALLY, ON SOME DAYS TWO READINGS WERE STORED IN THE METER'S MEMORY WHEN ONLY ONE TEST WAS PERFORMED. AFTER THE CUSTOMER CHANGED THE BATTERIES, THE READINGS WERE PROPERLY STORED IN THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17390858 ↗
  • MalfunctionJul 1, 2023Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT ONE METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17244206 ↗
  • MalfunctionJul 1, 2023Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT ONE OF HIS BLOOD GLUCOSE READINGS WAS NOT STORED IN THE MEMORY OF THE CONTOUR NEXT ONE METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17244202 ↗

FDA clearance and classification

Cleared through 510(k) K160682 on Nov 17, 2016 (decision: Substantially Equivalent). Submitted by Ascensia Diabetes Care as “CONTOUR NEXT ONE Blood Glucose Monitoring System”. The FDA review took 252 days. FDA record ↗

The FDA regulates product code LFR (glucose dehydrogenase, glucose) under 21 CFR 862.1345 as a Class II device. 21 CFR 862.1345 ↗

Common questions about the CONTOUR® NEXT ONE 9763

Has the CONTOUR® NEXT ONE 9763 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00301939763012), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the CONTOUR® NEXT ONE 9763?

The FDA's MAUDE database lists 120 adverse event reports for this device, received between Dec 6, 2017 and Nov 4, 2025. In those reports, the most common specific device problems are “Missing Test Results” (88), “Incorrect, Inadequate or Imprecise Result or Readings” (9) and “Nonstandard Device” (9); the most common patient problems are “Dizziness” (4), “Hypoglycemia” (3) and “Cognitive Changes” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CONTOUR® NEXT ONE 9763?

None of the reports are classified as deaths, 5 as injuries and 115 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the CONTOUR® NEXT ONE 9763 FDA approved?

It was cleared through 510(k) K160682 on Nov 17, 2016. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the CONTOUR® NEXT ONE 9763, and is it still sold?

The FDA's device database lists Ascensia Diabetes Care US Inc. as the company responsible for it (brand CONTOUR® NEXT ONE, model 9763). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Glucose Dehydrogenase, Glucose devices

All Glucose Dehydrogenase, Glucose devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page