Henry Schein Maxima 5702660Recalls & reported problems

Henry Schein, Inc. · Maxima ELITE 2 1:5 Contra Angle · Class 1

All models of Henry Schein Maxima (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
14
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00304040153857), received between Jul 27, 2023 and Jan 30, 2025 — classified as 14 injuries, 1 malfunction.

The most frequently reported specific device problem is “Use of Device Problem” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem6
  • Excessive Heating5
  • Temperature Problem3
  • Insufficient Device Problem Information2
  • Detachment of Device or Device Component1
  • Mechanical Problem1

Patient problems

  • Burn(s)9
  • Superficial (First Degree) Burn3
  • Foreign Body In Patient2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 30, 2025Temperature Problem; Temperature Problem

    TWO FEMALE PATIENTS WERE BURNED FROM THE MAXIMA ELITE 2 CONTRA ANGLE ON TWO DIFFERENT OCCASIONS. ONE TIME WAS ON (B)(6) 2024 WHILE THE OTHER WAS ON (B)(6) 2024. ONE PATIENT IS 43 YEARS OLD WHILE THE OTHER IS 67 YEARS OLD. IT IS NOT CLEAR WHICH HAD THE BURN WHERE BUT ONE WAS ON THE LIP WHILE THE OTHER WAS ON THE BUCCAL MUSCOSA. THE TYPE OF BURN WAS A LARGE ULCER. VITAMIN E WAS PLACED ON THE AREA FOR THE BURN AND NO FURTHER TREATMENT NEEDED. IT WAS THE HEAD PART THAT GOT HOT. ONE PATIENT IS DOING FINE AS OF TODAY WHILE THE OTHER THEY HAVE NOT SEEN SINCE. ADDITIONAL CORRESPONDENCE FROM THE DOCTOR WAS RECEIVED ON (B)(6) 2024: IT WAS STATED THAT BOTH A PATIENT AND STAFF MEMBER RECEIVED A BURN FROM THE HANDPIECE IS THIS CORRECT? NO, 2 PATIENTS WERE BURNED. WE WERE ALSO MADE AWARE THAT IT WAS A MALE AND FEMALE. WHICH WAS THE PATIENT AND WHICH WAS THE STAFF MEMBER? NO, 2 FEMALE PATIENTS. COULD YOU SHARE THE AGE OF BOTH THE PATIENT AND STAFF MEMBER? 43 YEARS OLD AND 67 YEARS OLD. WHERE WAS THE BURN RECEIVED ON EACH? ONE ON THE LIP AND THE OTHER ON THE BUCCAL MUCOSA. WHAT TYPE OF BURN WAS RECEIVED? IT WAS A LARGE ULCER. WHAT WAS DONE TO TREAT THE BURN RECEIVED? VITAMIN E WAS PLACED. WHAT PAR

    Patient problems: Burn(s); Superficial (First Degree) Burn

    MAUDE report 21270988 ↗
  • InjuryJan 30, 2025Temperature Problem; Temperature Problem

    TWO FEMALE PATIENTS WERE BURNED FROM THE MAXIMA ELITE 2 CONTRA ANGLE ON TWO DIFFERENT OCCASIONS. ONE TIME WAS ON (B)(6)2024 WHILE THE OTHER WAS (B)(6)2024. ONE PATIENT IS 43 YEARS OLD WHILE THE OTHER IS 67 YEARS OLD. IT IS NOT CLEAR WHICH HAD THE BURN WHERE BUT ONE WAS ON THE LIP WHILE THE OTHER WAS ON THE BUCCAL MUSCOSA. THE TYPE OF BURN WAS A LARGE ULCER. VITAMIN E WAS PLACED ON THE AREA FOR THE BURN AND NO FURTHER TREATMENT NEEDED. IT WAS THE HEAD PART THAT GOT HOT. ONE PATIENT IS DOING FINE AS OF TODAY WHILE THE OTHER THEY HAVE NOT SEEN SINCE. ADDITIONAL CORRESPONDENCE FROM THE DOCTOR WAS RECEIVED ON 12/18/2024: IT WAS STATED THAT BOTH A PATIENT AND STAFF MEMBER RECEIVED A BURN FROM THE HANDPIECE IS THIS CORRECT? NO, 2 PATIENTS WERE BURNED. WE WERE ALSO MADE AWARE THAT IT WAS A MALE AND FEMALE. WHICH WAS THE PATIENT AND WHICH WAS THE STAFF MEMBER? NO, 2 FEMALE PATIENTS. COULD YOU SHARE THE AGE OF BOTH THE PATIENT AND STAFF MEMBER? 43 YEARS OLD AND 67 YEARS OLD. WHERE WAS THE BURN RECEIVED ON EACH? ONE ON THE LIP AND THE OTHER ON THE BUCCAL MUCOSA. WHAT TYPE OF BURN WAS RECEIVED? IT WAS A LARGE ULCER. WHAT WAS DONE TO TREAT THE BURN RECEIVED? VITAMIN E WAS PLACED. WHAT PART OF T

    Patient problems: Superficial (First Degree) Burn

    MAUDE report 21270923 ↗
  • InjuryJan 8, 2025Excessive Heating; Excessive Heating

    TWO FEMALE PATIENTS WERE BURNED FROM THE MAXIMA ELITE 2 CONTRA ANGLE ON TWO DIFFERENT OCCASIONS. ONE TIME WAS ON (B)(6) 2024 WHILE THE OTHER WAS (B)(6) 2024. ONE PATIENT IS 43 YEARS OLD WHILE THE OTHER IS 67 YEARS OLD. IT IS NOT CLEAR WHICH HAD THE BURN WHERE BUT ONE WAS ON THE LIP WHILE THE OTHER WAS ON THE BUCCAL MUSCOSA. THE TYPE OF BURN WAS A LARGE ULCER. VITAMIN E WAS PLACED ON THE AREA FOR THE BURN AND NO FURTHER TREATMENT NEEDED. IT WAS THE HEAD PART THAT GOT HOT. ONE PATIENT IS DOING FINE AS OF TODAY WHILE THE OTHER THEY HAVE NOT SEEN SINCE. ADDITIONAL CORRESPONDENCE FROM THE DOCTOR WAS RECEIVED ON 12/18/24: IT WAS STATED THAT BOTH A PATIENT AND STAFF MEMBER RECEIVED A BURN FROM THE HANDPIECE IS THIS CORRECT? NO, 2 PATIENTS WERE BURNED . WE WERE ALSO MADE AWARE THAT IT WAS A MALE AND FEMALE. WHICH WAS THE PATIENT AND WHICH WAS THE STAFF MEMBER? NO, 2 FEMALE PATIENTS. COULD YOU SHARE THE AGE OF BOTH THE PATIENT AND STAFF MEMBER? 43 YEARS OLD AND 67 YEARS OLD. WHERE WAS THE BURN RECEIVED ON EACH? ONE ON THE LIP AND THE OTHER ON THE BUCCAL MUCOSA. WHAT TYPE OF BURN WAS RECEIVED? IT WAS A LARGE ULCER. WHAT WAS DONE TO TREAT THE BURN RECEIVED? VITAMIN E WAS PLACED. WHAT PART OF

    Patient problems: Superficial (First Degree) Burn

    MAUDE report 21108022 ↗
  • MalfunctionDec 4, 2024Temperature Problem; Temperature Problem

    DEVICE WAS OVER HEATING DURING TREATMENT OF PATIENT. NO INJURIES OR BURNS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20843640 ↗
  • InjuryNov 13, 2024Excessive Heating; Excessive Heating

    DEVICE WAS OVER HEATING DURING TREATMENT OF PATIENT. NO INJURIES OR BURNS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20676070 ↗
  • InjuryOct 17, 2024Mechanical Problem; Mechanical Problem

    DOCTOR WAS USING THE HANDPIECE ON THE PATIENT WHEN THE BUR DISLODGED FROM THE CHUCK AND WAS SWALLOWED BY THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 20470114 ↗
  • InjurySep 20, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    DOCTOR WAS USING THE HANDPIECE ON THE PATIENT WHEN THE BUR DISLODGED FROM THE CHUCK AND WAS SWALLOWED BY THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 20267963 ↗
  • InjuryJul 17, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    THE CUSTOMER DESCRIBED THE HANDPIECES ARE STILL HAVING OVERHEATING ISSUES.

    Patient problems: Burn(s)

    MAUDE report 19759695 ↗
  • InjuryJun 19, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    THE CUSTOMER DESCRIBED THE HANDPIECES ARE STILL HAVING OVERHEATING ISSUES.

    Patient problems: Burn(s)

    MAUDE report 19570635 ↗
  • InjuryAug 23, 2023Use of Device Problem; Use of Device Problem

    AFTER SEVERAL USES, THE PRODUCT HEATED UP AND BURNED THREE (3) PATIENTS.

    Patient problems: Burn(s)

    MAUDE report 17600814 ↗
  • InjuryAug 23, 2023Use of Device Problem; Use of Device Problem

    AFTER SEVERAL USES, THE PRODUCT HEATED UP AND BURNED THREE (3) PATIENTS.

    Patient problems: Burn(s)

    MAUDE report 17600813 ↗
  • InjuryAug 23, 2023Use of Device Problem; Use of Device Problem

    AFTER SEVERAL USES, THE PRODUCT HEATED UP AND BURNED THREE (3) PATIENTS.

    Patient problems: Burn(s)

    MAUDE report 17600811 ↗
  • InjuryJul 27, 2023Use of Device Problem; Excessive Heating; Use of Device Problem; Excessive Heating

    AFTER SEVERAL USES, THE PRODUCT HEATED UP AND BURNED THREE (3) PATIENTS.

    Patient problems: Burn(s)

    MAUDE report 17405968 ↗
  • InjuryJul 27, 2023Use of Device Problem; Excessive Heating; Use of Device Problem; Excessive Heating

    AFTER SEVERAL USES, THE PRODUCT HEATED UP AND BURNED THREE (3) PATIENTS.

    Patient problems: Burn(s)

    MAUDE report 17405948 ↗
  • InjuryJul 27, 2023Use of Device Problem; Excessive Heating; Use of Device Problem; Excessive Heating

    AFTER SEVERAL USES, THE PRODUCT HEATED UP AND BURNED THREE (3) PATIENTS.

    Patient problems: Burn(s)

    MAUDE report 17405862 ↗

Other Handpiece, air-powered, dental devices

All Handpiece, air-powered, dental devices →