Skip to content

Z890LRecalls & reported problems

Nakanishi Inc. · Class 1

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04589551385348), received between Jun 21, 2024 and Dec 11, 2025 — classified as 5 injuries.

The most frequently reported specific device problem is “Overheating of Device” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device4
  • Detachment of Device or Device Component1

Patient problems

  • Partial thickness (Second Degree) Burn4
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 11, 2025Overheating of Device; Overheating of Device

    ON (B)(6) 2025, NAKANISHI BECAME AWARE OF A HANDPIECE OVERHEATING THROUGH A COMPLAINT INPUT INTO THE COMPLAINT DATABASE BY A DISTRIBUTOR (NSK AMERICA). ACCORDING TO THE DISTRIBUTOR, TWO PATIENTS WERE INVOLVED WITH THE DEVICE. THEREFORE, NAKANISHI IS SUBMITTING TWO MDRS FOR THE TWO PATIENTS. THE DETAILS NAKANISHI OBTAINED ABOUT THE SECOND PATIENT ARE AS FOLLOWS. - THE DENTIST WAS PERFORMING A DENTAL PROCEDURE ON A PATIENT USING THE Z890L HANDPIECE (SERIAL NO. (B)(6). - DURING THE PROCEDURE, THE BACK OF THE HANDPIECE HEAD OVERHEATED, AND THE PATIENT RECEIVED A THERMAL BURN INJURY TO THEIR INNER CHEEK.

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 23785511 ↗
  • InjuryDec 11, 2025Overheating of Device; Overheating of Device

    ON (B)(6) , 2025, NAKANISHI BECAME AWARE OF A HANDPIECE OVERHEATING THROUGH A COMPLAINT INPUT INTO THE COMPLAINT DATABASE BY A DISTRIBUTOR (NSK AMERICA). ACCORDING TO THE DISTRIBUTOR, TWO PATIENTS WERE INVOLVED WITH THE DEVICE. THEREFORE, NAKANISHI IS SUBMITTING TWO MDRS FOR THE TWO PATIENTS. THE DETAILS NAKANISHI OBTAINED ABOUT THE FIRST PATIENT ARE AS FOLLOWS. - THE DENTIST WAS PERFORMING A DENTAL PROCEDURE ON A PATIENT USING THE Z890L HANDPIECE (SERIAL NO. (B)(6) ) - DURING THE PROCEDURE, THE BACK OF THE HANDPIECE HEAD OVERHEATED, AND THE PATIENT RECEIVED A THERMAL BURN INJURY TO THEIR INNER CHEEK.

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 23785469 ↗
  • InjuryNov 21, 2025Overheating of Device; Overheating of Device

    DENTAL OFFICE REPORTED THAT THEIR PATIENT RECEIVED A THERMAL BURN INJURY TO THEIR INNER CHEEK FROM THE BACK OF THE HANDPIECE HEAD DURING A PROCEDURE. PATIENT INFORMATION AND DATE OF THE ADVERSE EVENT HAVE NOT BEEN PROVIDED BY THE DENTAL PRACTICE AT THE TIME OF THIS REPORT. A FOLLOW-UP REPORT WILL BE MADE IF THIS INFORMATION IS MADE AVAILABLE BY THE DENTAL OFFICE.

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 23623586 ↗
  • InjuryNov 21, 2025Overheating of Device; Overheating of Device

    DENTAL OFFICE REPORTED THAT THEIR PATIENT RECEIVED A THERMAL BURN INJURY TO THEIR INNER CHEEK FROM THE BACK OF THE HANDPIECE HEAD DURING A PROCEDURE. PATIENT INFORMATION AND DATE OF THE ADVERSE EVENT HAVE NOT BEEN PROVIDED BY THE DENTAL PRACTICE AT THE TIME OF THIS REPORT. A FOLLOW-UP REPORT WILL BE MADE IF THIS INFORMATION IS MADE AVAILABLE BY THE DENTAL OFFICE.

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 23623535 ↗
  • InjuryJun 21, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE DOCTOR WAS PERFORMING A CROWN PREPARATION WHEN HE NOTICED THAT THE HANDPIECE BEGAN TO LOSE POWER AND WAS NOT CUTTING WELL, IT THEN BECAME NOTICEABLY LOUDER AND THE BUR ALONG WITH THE INTERNAL PARTS JUST SHOT OUT INTO THE PATIENT'S MOUTH. THE DOCTOR WAS ABLE TO QUICKLY RETRIEVE THE PARTS AND THE PATIENT WAS NOT AFFECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19580550 ↗

Other Handpiece, air-powered, dental devices

All Handpiece, air-powered, dental devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.