Braun IRT6500USRecalls & reported problems
Kaz USA, Inc. · Braun Infrared ThermoScan Ear Thermometer - for intermittent measurement of temperature for patients of all ages, this thermometer is used with lens filters. · Class 2
All models of Braun (13)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00328785000054), received between Jun 3, 2020 and Apr 16, 2025 — classified as 10 malfunctions.
The most frequently reported specific device problem is “False Negative Result” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- False Negative Result10
- Insufficient Device Problem Information2
Patient problems
- Fever10
- Pneumonia1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 16, 2025False Negative Result; False Negative Result
A CONSUMER REPORTED THAT HIS THERMOMETER HAD ALLEGEDLY GIVEN FALSE NEGATIVE READINGS ON HIS SON. THE DEVICE ALLEGEDLY GAVE CONSISTENT READINGS OF 96.7°F, AND A FEVER OF 101.0°F WAS LATER CONFIRMED BY A DOCTOR. THERE WERE NO COMPLICATIONS FROM THIS INCIDENT. KAZ USA, INC. HAS REQUESTED THAT THE PRODUCT BE RETURNED TO OUR COMPANY FOR TESTING.
Patient problems: Fever
MAUDE report 21852780 ↗MalfunctionMar 29, 2024False Negative Result; False Negative Result
A CONSUMER REPORTED THAT THEIR THERMOMETER HAD ALLEGEDLY GIVEN FALSE NEGATIVE READINGS ON HIMSELF. THE DEVICE ALLEGEDLY GAVE A READING OF 95.0°F, AND A FEVER OF 100.9°F WAS LATER MEASURED AT A DOCTOR'S OFFICE. THERE WERE NO COMPLICATIONS FROM THIS INCIDENT, AND THE PATIENT IS DOING WELL NOW.
Patient problems: Fever
MAUDE report 19002877 ↗MalfunctionMar 16, 2023False Negative Result; False Negative Result
A CONSUMER REPORTED THAT HER THERMOMETER HAD ALLEGEDLY GIVEN FALSE NEGATIVE READINGS ON HER DAUGHTER. THE DEVICE ALLEGEDLY GAVE A READING OF 97.6°F, AND A FEVER OF 101.0°F WAS LATER CONFIRMED BY A DOCTOR. THERE WERE NO COMPLICATIONS FROM THIS INCIDENT, AND THE PATIENT IS DOING WELL NOW. KAZ USA, INC. HAS REQUESTED THAT THE PRODUCT BE RETURNED TO OUR COMPANY FOR TESTING.
Patient problems: Fever
MAUDE report 16552802 ↗MalfunctionOct 20, 2022False Negative Result; False Negative Result
A CONSUMER REPORTED THAT HER THERMOMETER HAD ALLEGEDLY GIVEN FALSE NEGATIVE READINGS ON HER FRIENDS' FATHER. THE DEVICE ALLEGEDLY GAVE READINGS AROUND 101°F, AND THE PATIENT WAS LATER TAKEN BY AMBULANCE WHERE A FEVER OF 107°F WAS CONFIRMED BY PARAMEDICS. THE PATIENT WAS TREATED AT A HOSPITAL FOR A FEW DAYS, AND WAS LATER RELEASED. THERE WERE NO COMPLICATIONS FROM THIS INCIDENT, AND THE PATIENT IS DOING WELL NOW. KAZ USA, INC. HAS REQUESTED THAT THE PRODUCT BE RETURNED TO OUR COMPANY FOR TESTING.
Patient problems: Fever
MAUDE report 15639239 ↗MalfunctionOct 13, 2022False Negative Result; False Negative Result
A CONSUMER REPORTED THAT HIS THERMOMETER HAD ALLEGEDLY GIVEN FALSE NEGATIVE READINGS ON HIS SON. THE DEVICE ALLEGEDLY GAVE READINGS THAT WERE 6-7°F LOWER THAN WAS LATER MEASURED AT A DOCTOR'S OFFICE. THERE WERE NO COMPLICATIONS FROM THIS INCIDENT, AND THE PATIENT IS DOING WELL NOW. KAZ USA, INC. HAD REQUESTED THAT THE PRODUCT BE RETURNED TO OUR COMPANY FOR TESTING.
Patient problems: Fever
MAUDE report 15597516 ↗MalfunctionMar 8, 2022False Negative Result; False Negative Result
A CONSUMER REPORTED THAT HIS THERMOMETER HAD ALLEGEDLY GIVEN FALSE NEGATIVE READINGS ON HER SON. THE DEVICE ALLEGEDLY GAVE READINGS THAT WERE 3-4°F LOWER THAN WAS LATER MEASURED AT A DOCTOR'S OFFICE. THERE WERE NO COMPLICATIONS FROM THIS INCIDENT, AND THE PATIENT IS DOING WELL NOW. KAZ USA, INC. HAD REQUESTED THAT THE PRODUCT BE RETURNED TO OUR COMPANY FOR TESTING.
Patient problems: Fever
MAUDE report 13693217 ↗MalfunctionDec 19, 2021False Negative Result; False Negative Result
A CONSUMER REPORTED THAT A THERMOMETER HAD ALLEGEDLY GIVEN FALSE NEGATIVE READINGS ON HER INFANT. THE DEVICE ALLEGEDLY GAVE READINGS THAT WERE 4°F LOWER THAN WAS LATER MEASURED AT A DOCTOR'S OFFICE. THERE WERE NO COMPLICATIONS FROM THIS INCIDENT, AND THE PATIENT IS DOING WELL NOW. KAZ USA, INC. HAD REQUESTED THAT THE PRODUCT BE RETURNED TO OUR COMPANY FOR TESTING.
Patient problems: Fever
MAUDE report 13041129 ↗MalfunctionNov 4, 2021False Negative Result; False Negative Result
A CONSUMER REPORTED THAT THEIR THERMOMETER HAD GIVEN FALSE NEGATIVE READINGS ON HERSELF. THE DEVICE ALLEGEDLY WAS READING 5.2°F LOWER THAN WHAT WAS MEASURED BY A DOCTOR, WHERE A READING OF 102.8°F WAS OBTAINED. THE PATIENT WAS TRANSPORTED TO A HOSPITAL, AND TREATED FOR A FEVER AND PNEUMONIA. THERE WERE NO COMPLICATIONS FROM THIS INCIDENT, AND THE PATIENT IS DOING WELL NOW. KAZ USA, INC. HAS REQUESTED THAT THE DEVICE BE RETURNED TO OUR COMPANY FOR TESTING.
Patient problems: Fever; Pneumonia
MAUDE report 12750652 ↗MalfunctionJun 10, 2020False Negative Result; Insufficient Device Problem Information; False Negative Result; Insufficient Device Problem Information
A CONSUMER REPORTED THAT HER THERMOMETER HAD ALLEGEDLY GIVEN FALSE NEGATIVE READINGS ON HERSELF. THE DEVICE ALLEGEDLY GAVE A READING THAT WAS 3.2°F LOWER THAN WHAT WAS LATER MEASURED AT THE HOSPITAL, WHERE A FEVER WAS RECORDED. THE CONSUMER TESTED POSITIVE FOR COVID-19 AND WAS ADMITTED FOR TREATMENT FOR NINE DAYS. THERE WERE NO COMPLICATIONS FROM THIS INCIDENT, AND THE PATIENT IS DOING WELL NOW. KAZ USA, INC. HAD REQUESTED THAT THE PRODUCT BE RETURNED TO OUR COMPANY FOR TESTING.
Patient problems: Fever
MAUDE report 10137595 ↗MalfunctionJun 3, 2020False Negative Result; Insufficient Device Problem Information; False Negative Result; Insufficient Device Problem Information
A CONSUMER REPORTED THAT HER THERMOMETER HAD ALLEGEDLY GIVEN FALSE NEGATIVE READINGS ON HER DAUGHTER. THE DEVICE ALLEGEDLY GAVE A READING THAT WAS 5.8 - 6.8°F LOWER THAN WHAT WAS LATER MEASURED AT THE HOSPITAL, WHERE A HIGH FEVER WAS RECORDED. THERE WERE NO COMPLICATIONS FROM THIS INCIDENT, AND THE PATIENT IS DOING WELL NOW. KAZ USA, INC. HAD REQUESTED THAT THE PRODUCT BE RETURNED TO OUR COMPANY FOR TESTING, BUT THE CONSUMER STATED THAT THEY HAD ALREADY DISCARDED THE DEVICE.
Patient problems: Fever
MAUDE report 10114324 ↗
FDA clearance and classification
The FDA regulates product code FLL (thermometer, electronic, clinical) under 21 CFR 880.2910 as a Class II device. 21 CFR 880.2910 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Braun IRT6500US
Has the Braun IRT6500US been recalled?
No FDA recall record in our data names this exact device identifier (DI 00328785000054), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Braun IRT6500US?
The FDA's MAUDE database lists 10 adverse event reports for this device, received between Jun 3, 2020 and Apr 16, 2025. In those reports, the most common specific device problem is “False Negative Result” (10); the most common patient problems are “Fever” (10) and “Pneumonia” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Braun IRT6500US?
None of the reports are classified as deaths, 0 as injuries and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Braun IRT6500US, and is it still sold?
The FDA's device database lists Kaz USA, Inc. as the company responsible for it (brand Braun, model IRT6500US). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Thermometer, Electronic, Clinical devices
- Welch Allyn, Inc. 901054Welch Allyn, Inc. · Thermometer, electronic, clinical1 recall221 reports
- Welch Allyn, Inc. 901054Welch Allyn, Inc. · Thermometer, electronic, clinical2 recalls13 reports
- NA B105 VSP3.0 Patient MonitorGE Medical Systems Information Technologies, Inc. · Thermometer, electronic, clinical2 recalls0 reports
- Mon-a-Therm 90050Covidien LP · Thermometer, electronic, clinical1 recall3 reports
- Mon-a-Therm 90044Covidien LP · Thermometer, electronic, clinical1 recall2 reports
- CARESCAPE B450 MBA313GE Healthcare Finland Oy · Thermometer, electronic, clinical1 recall1 report
- Temperature Probes Esophageal/Rectal Temperature ProbePhilips Medical Systems Hsg · Thermometer, electronic, clinical1 recall1 report
- NA 989803162621Philips Medical Systems Hsg · Thermometer, electronic, clinical1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.