Braun IRT6520BUSRecalls & reported problems

Kaz USA, Inc. · Braun ThermoScan 7 Ear Therm Black · Class 2

All models of Braun (13)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00328785001914), received between Mar 17, 2022 and Jan 20, 2026 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result6

Patient problems

  • Fever6
  • Convulsion/Seizure1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 20, 2026False Negative Result

    A HOSPITAL REPORTED THAT THE EAR THERMOMETER ALLEGEDLY PROVIDED A LOW TEMPERATURE READING FOR A PEDIATRIC PATIENT. THE CHILD'S TEMPERATURE WAS REPORTEDLY MEASURED AT 37.2°C. A SUBSEQUENT MEASUREMENT USING A MERCURY THERMOMETER ALLEGEDLY SHOWED A TEMPERATURE OF 39.7°C, INDICATING THE PRESENCE OF A FEVER. NO ADDITIONAL INFORMATION WAS PROVIDED REGARDING THE PATIENT'S DIAGNOSIS OR CLINICAL OUTCOME. THE ONLY INFORMATION AVAILABLE IS THAT A TEMPERATURE DISCREPANCY OCCURRED, AND THE DEVICE WAS REPORTED TO HAVE GIVEN A LOW READING. KAZ USA, INC. HAS REQUESTED THAT THE PRODUCT BE RETURNED TO OUR COMPANY FOR TESTING.

    Patient problems: Fever

    MAUDE report 24126613 ↗
  • MalfunctionDec 31, 2025False Negative Result; False Negative Result

    A CONSUMER REPORTED THAT THE THERMOMETER ALLEGEDLY PROVIDED FALSE NEGATIVE READINGS ON THEIR 7 YEAR OLD DAUGHTER. THE DEVICE ALLEGEDLY DISPLAYED A MEASUREMENT OF 37°C, WHILE A FEVER OF 40°C WAS LATER CONFIRMED BY A DOCTOR WHO ADMINISTERED MEDICATION. THERE WERE NO COMPLICATIONS FROM THE INCIDENT, AND THE CHILD IS DOING WELL NOW. KAZ USA, INC. HAD REQUESTED THAT THE PRODUCT BE RETURNED TO OUR COMPANY FOR TESTING.

    Patient problems: Fever

    MAUDE report 23938613 ↗
  • MalfunctionMay 11, 2025False Negative Result; False Negative Result

    A HOSPITAL REPORTED THAT THEIR THERMOMETER HAD ALLEGEDLY GIVEN A FALSE NEGATIVE READING ON A PATIENT WHO HAD ALREADY BEEN ADMITTED WITH A LUNG INFECTION. THE DEVICE ALLEGEDLY GAVE A READING OF 36.9°C, AND A FEVER OF 38.9°C WAS CONFIRMED BY A MEDICAL PROFESSIONAL USING A DIFFERENT EAR THERMOMETER. NO INFORMATION WAS PROVIDED ON THE DIAGNOSIS OR CURRENT STATUS OF THE PATIENT. THE ONLY KNOWN INFORMATION PROVIDED WAS THAT A FEVER WAS CONFIRMED BY A MEDICAL PROFESSIONAL, AND OUR DEVICE WAS GIVING FALSE NEGATIVE READINGS. KAZ USA, INC. HAS REQUESTED THAT THE PRODUCT BE RETURNED TO OUR COMPANY FOR TESTING.

    Patient problems: Fever

    MAUDE report 21999806 ↗
  • InjuryJan 10, 2025False Negative Result; False Negative Result

    A CONSUMER REPORTED THAT THEIR THERMOMETER HAD ALLEGEDLY GIVEN FALSE NEGATIVE READINGS ON THEIR INFANT SON. THE DEVICE ALLEGEDLY GAVE READINGS RANGING FROM 36.5 TO 38.9°C. THE CHILD LATER HAD A FEBRILE SEIZURE AND WAS TAKEN TO THE HOSPITAL WHERE A FEVER OF 38.0°C WAS CONFIRMED. THE CHILD WAS ADMITTED TO THE HOSPITAL FOR THREE DAYS. THERE WERE NO COMPLICATIONS FROM THIS INCIDENT, AND THE PATIENT IS DOING WELL NOW. KAZ USA, INC. HAS REQUESTED THAT THE PRODUCT BE RETURNED TO OUR COMPANY FOR TESTING.

    Patient problems: Convulsion/Seizure; Fever

    MAUDE report 21129508 ↗
  • MalfunctionNov 30, 2024False Negative Result; False Negative Result

    A HOSPITAL REPORTED THAT THEIR THERMOMETER HAD ALLEGEDLY GIVEN A FALSE NEGATIVE READING ON A PATIENT WHO WAS FEBRILE DUE TO HIGH LEUKOCYTES CAUSED BY AN INFECTION. THE DEVICE ALLEGEDLY GAVE A READING OF 34°C, AND A FEVER OF 39°C WAS CONFIRMED BY A MEDICAL PROFESSIONAL USING A MERCURY THERMOMETER. NO INFORMATION WAS PROVIDED ON THE DIAGNOSIS OR CURRENT STATUS OF THE PATIENT. THE ONLY KNOWN INFORMATION PROVIDED WAS THAT A FEVER WAS CONFIRMED BY A MEDICAL PROFESSIONAL, AND OUR DEVICE WAS GIVING FALSE NEGATIVE READINGS. KAZ USA, INC. HAS REQUESTED THAT THE PRODUCT BE RETURNED TO OUR COMPANY FOR TESTING.

    Patient problems: Fever

    MAUDE report 20814091 ↗
  • MalfunctionMar 17, 2022False Negative Result; False Negative Result

    A CONSUMER REPORTED THAT HIS THERMOMETER HAD ALLEGEDLY GIVEN FALSE NEGATIVE READINGS ON HER SON. THE DEVICE ALLEGEDLY GAVE READINGS THAT WERE 5-6°F LOWER THAN WAS LATER MEASURED AT A DOCTOR'S OFFICE. THERE WERE NO COMPLICATIONS FROM THIS INCIDENT, AND THE PATIENT IS DOING WELL NOW. KAZ USA, INC. HAD REQUESTED THAT THE PRODUCT BE RETURNED TO OUR COMPANY FOR TESTING.

    Patient problems: Fever

    MAUDE report 13798576 ↗

FDA clearance and classification

The FDA regulates product code FLL (thermometer, electronic, clinical) under 21 CFR 880.2910 as a Class II device. 21 CFR 880.2910 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Braun IRT6520BUS

Has the Braun IRT6520BUS been recalled?

No FDA recall record in our data names this exact device identifier (DI 00328785001914), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Braun IRT6520BUS?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between Mar 17, 2022 and Jan 20, 2026. In those reports, the most common specific device problem is “False Negative Result” (6); the most common patient problems are “Fever” (6) and “Convulsion/Seizure” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Braun IRT6520BUS?

None of the reports are classified as deaths, 1 as an injury and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Braun IRT6520BUS, and is it still sold?

The FDA's device database lists Kaz USA, Inc. as the company responsible for it (brand Braun, model IRT6520BUS). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Thermometer, Electronic, Clinical devices

All Thermometer, Electronic, Clinical devices →