ENFit PU Feeding Tube PFTS5.0P-NCRecalls & reported problems

Avanos Medical, Inc. · ENFit PU Feeding Tube Purp 5Fr 40cm · Class 2

All models of ENFit PU Feeding Tube (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00350770001753), received between May 17, 2024 and Sep 17, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

1 other recall cover devices cleared under the same FDA submission (K082238); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Material Puncture/Hole1
  • Migration or Expulsion of Device1

Patient problems

  • Bradycardia1
  • Foreign Body In Patient1
  • Insufficient Information1
  • Low Oxygen Saturation1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2026Break

    It was reported the patient's mucous fistula was being "refed taking ostomy content and refeeding through mucous fistula with a gastric tube at 9cm. The gastric tube "[was]" tied with a suture, and the suture strings are placed under the Hollister appliance, and the tube "[was]" also secured by Duoderm and ETT "[endotracheal tube]" tape." The clinician changed mucous fistula appliance at 0750-hours due to leakage and was secure after changed. At 1020-hours the patient was crying and the clinician noted ostomy output on the blanket outside mucous fistula. The clinician noted that the gastric tube was "broken" off from the gastric tube hub. The gastric tube was noted to be stretched out and broken down due continuous acidic stool content moving through the tube. This resulted in the ostomy output draining into the bed and not the gastric tube. It is unknown when the gastric tube broke off. The Hollister pouch remained intact. The Charge clinician and team were notified, and the gastric tube was removed and area cleansed and dressed. The healthcare provider came to reinsert the gastric tube at 1210-hours, there were no complications and refeeding was resumed."

    Patient problems: Insufficient Information

    MAUDE report 50806201 ↗
  • MalfunctionJun 9, 2026Break; Migration or Expulsion of Device

    THE CUSTOMER REPORTED THAT A NASOGASTRIC FEEDING TUBE BROKE DURING USE. THE PATIENT REPORTEDLY BEGAN VOMITING, DEVELOPED BRADYCARDIA AND OXYGEN DESATURATION, AND THE FEEDING TUBE WAS OBSERVED PROTRUDING FROM THE MOUTH. THE SECUREMENT DEVICE REMAINED IN PLACE AND THE BROKEN PORTION OF THE TUBE WAS REMOVED. ADDITIONAL INFORMATION REPORTED THAT THE TUBE PHYSICALLY SNAPPED AT APPROXIMATELY THE 26-CENTIMETER MARKING. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. ADDITIONAL INFORMATION RECEIVED ON MAY 26, 2026, REPORTED THAT THE TUBE PHYSICALLY SNAPPED AT THE PROXIMAL END WHERE THE SYRINGE CONNECTS.

    Patient problems: Bradycardia; Low Oxygen Saturation

    MAUDE report 25443305 ↗
  • MalfunctionMay 6, 2026Material Puncture/Hole

    IT WAS REPORTED THAT A NASOGASTRIC (NG) FEEDING TUBE WAS SUSPECTED TO HAVE A HOLE AFTER MULTIPLE PLACEMENT ATTEMPTS FAILED TO CONFIRM POSITION VIA AUSCULTATION. A SECOND FEEDING TUBE WAS PLACED AT THE SAME DEPTH AND PLACEMENT WAS SUCCESSFULLY CONFIRMED, INDICATING A LIKELY DEFECT IN THE INITIAL DEVICE. NO PATIENT HARM OR INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25104769 ↗
  • MalfunctionMay 17, 2024Break; Break

    IT WAS REPORTED, ¿WHEN CHANGING THE BABY¿S DIAPER 1.5 CM OF A TUBE WAS FOUND IN THE STOOL. ON INSPECTION IT LOOKED LIKE THE TIP OF A GASTRIC TUBE. I SHOWED THE FATHER AND CALLED DR TO THE BEDSIDE. WE REMOVED THE NEOMED ENFIT 5 FR NASOGASTRIC (NG) TUBE AND THE TIP WAS MISSING. THE PIECE IN THE DIAPER WAS THE TIP OF THE NG TUBE. A NEW NEOMED ENFIT 5 FR NG TUBE WAS PLACED. ON REVIEW, IT WAS FOUND THAT THE NG TUBE IN QUESTION WAS PLACED ON (B)(6) 2024. THE MANUFACTURE RECOMMENDATION IS THE TUBE SHOULD BE CHANGED OUT ON OR BEFORE 30 DAYS.¿ THE TUBE WAS REMOVED AND REPLACED.

    Patient problems: Foreign Body In Patient

    MAUDE report 19343541 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K082238. One submission often covers many models, so check the product and lot details in each record.

  • Z-0093-2019 · Jun 4, 2018 · Terminated
    NeoMed Enteral Feeding Tubes with Enteral Only Connectors, French sizes 4.0Fr, 5.0Fr, 6.5Fr, 8.0Fr, and lengths 40cm, 60cm, 90cm. The product consists of a polyurethane feeding tu

Other TUBE, NASOGASTRIC devices