ENFit PU Feeding Tube PFTS6.5P-NCRecalls & reported problems

Avanos Medical, Inc. · ENFit PU Feeding Tube Purp 6.5Fr 40cm · Class 2

All models of ENFit PU Feeding Tube (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00350770001784), received between Mar 17, 2026 and Jul 1, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K082238); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash2
  • Obstruction of Flow2
  • Pumping Problem2
  • Complete Blockage1

Patient problems

  • Insufficient Information3
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 1, 2026Leak/Splash

    THE FAILURE ANALYSIS LABORATORY RECEIVED AN EXTRA SAMPLE. THE RETURNED SAMPLE EXHIBITED DAMAGE (TEAR/HOLE). NO ADDITIONAL INCIDENT INFORMATION, PATIENT INFORMATION, OR LOT NUMBER WAS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 25678208 ↗
  • MalfunctionJun 8, 2026Complete Blockage

    IT WAS REPORTED THAT AN NG TUBE BECAME OCCLUDED AND HARDENED ALONG A PORTION OF THE TUBE DURING USE, PREVENTING FEEDING AND REQUIRING REMOVAL AND REPLACEMENT. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED. EVENT DATE UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25428014 ↗
  • MalfunctionMay 20, 2026Leak/Splash

    THE CUSTOMER REPORTED THAT A NASOGASTRIC (NG) FEEDING TUBE WAS OBSERVED LEAKING ENTERAL FEEDING FORMULA FROM THE PATIENT¿S NARE DURING FEEDING ADMINISTRATION. THE PUMP FEEDING WAS STOPPED, THE NG TUBE WAS REMOVED AND REINSERTED, AND LEAKAGE WAS AGAIN OBSERVED FROM THE NARE UPON RESUMPTION OF FEEDING. A NEW NG TUBE WAS THEN INSERTED, PLACEMENT WAS VERIFIED, AND FEEDING RESUMED WITHOUT FURTHER LEAKAGE. THE REMOVED NG TUBE WAS REPORTEDLY EVALUATED AND APPEARED TO BE LEAKING FROM APPROXIMATELY THE 18¿19 CM MARKING, NEAR THE AREA WHERE THE TUBE HAD BEEN SECURED AT THE NARE. ADDITIONAL INFORMATION WAS RECEIVED ON 11-MAY-2026, REPORTING NO INJURY AND NO HISTORY OF TUBE CLOGGING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25229322 ↗
  • MalfunctionMar 18, 2026Obstruction of Flow; Pumping Problem

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED DIFFERENT INCIDENTS, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING DIFFERENT PATIENTS. THIS IS THE FIRST OF THREE REPORTS. REFER TO 3011270181-2026-00037 FOR THE SECOND REPORT. REFER TO 3011270181-2026-00038 FOR THE THIRD REPORT. THE CUSTOMER REPORTED THAT CLINICIANS ON THE INFANT TODDLER FLOOR EXPERIENCED OCCLUSION ALARMS WHEN USING KANGAROO PUMPS WITH AVANOS 6.5 FR NEOMED FEEDING TUBES (REF: PFTS6.5P-NC) DURING INFUSION. THE PUMP ALARMED ¿OCCLUDED¿ AND DID NOT RESPOND TO TROUBLESHOOTING. THE ISSUE OCCURRED DURING USE WITH PEDIALYTE (THIN LIQUID), AND CLINICIANS SUSPECTED A POSSIBLE PSI OR PRESSURE SETTING ISSUE RELATED TO THE FEEDING TUBE. MULTIPLE EVENTS WERE REPORTED WITH THIS TUBE SIZE OVER THE COURSE OF A WEEK. THE EXACT QUANTITY IS UNKNOWN. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 24630567 ↗
  • MalfunctionMar 17, 2026Obstruction of Flow; Pumping Problem

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING DIFFERENT PATIENTS. THIS IS THE SECOND OF THREE REPORTS. REFER TO 3011270181-2026-00036 FOR THE FIRST REPORT, REFER TO 3011270181-2026-00038 FOR THE THIRD REPORT. THE CUSTOMER REPORTED THAT CLINICIANS ON THE INFANT TODDLER FLOOR EXPERIENCED OCCLUSION ALARMS WHEN USING KANGAROO PUMPS WITH AVANOS 6.5 FR NEOMED FEEDING TUBES (REF: PFTS6.5P-NC) DURING INFUSION. THE PUMP ALARMED ¿OCCLUDED¿ AND DID NOT RESPOND TO TROUBLESHOOTING. THE ISSUE OCCURRED DURING USE WITH PEDIALYTE (THIN LIQUID), AND CLINICIANS SUSPECTED A POSSIBLE PSI OR PRESSURE SETTING ISSUE RELATED TO THE FEEDING TUBE. MULTIPLE EVENTS WERE REPORTED WITH THIS TUBE SIZE OVER THE COURSE OF A WEEK. THE EXACT QUANTITY IS UNKNOWN. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 24621348 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K082238. One submission often covers many models, so check the product and lot details in each record.

  • Z-0093-2019 · Jun 4, 2018 · Terminated
    NeoMed Enteral Feeding Tubes with Enteral Only Connectors, French sizes 4.0Fr, 5.0Fr, 6.5Fr, 8.0Fr, and lengths 40cm, 60cm, 90cm. The product consists of a polyurethane feeding tu

Other TUBE, NASOGASTRIC devices

All TUBE, NASOGASTRIC devices →