AVANOS 98437Recalls & reported problems
Avanos Medical, Inc. · Introducer Kit for Jejunal/Gastric Jejunal Feeding Tube · Class 2 · Implant
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00350770984377), received between Jun 26, 2020 and Feb 19, 2026 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Device Slipped2
- Premature Separation1
Patient problems
- No Known Impact Or Consequence To Patient3
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
- Foreign Body In Patient1
- No Code Available1
- Pain1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 19, 2026Device Slipped
AVANOS MEDICAL RECEIVED A SINGLE REPORT THAT REFERENCED SIX DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING TWO DIFFERENT EVENTS. THIS IS THE FIRST OF TWO REPORTS. REFER TO 9611594-2026-00078 FOR THE SECOND EVENT. IT WAS REPORTED THAT DURING THE USE OF AN INTRODUCER KIT DURING A PROCEDURE; WHILE PEELING AWAY THE PEEL AWAY SHEATH, A PART OF THE DILATOR SLIPPED OVER THE GUIDEWIRE AND ENTERED THE PATIENT¿S STOMACH.
Patient problems: Foreign Body In Patient
MAUDE report 24392186 ↗MalfunctionFeb 19, 2026Device Slipped
AVANOS MEDICAL RECEIVED A SINGLE REPORT THAT REFERENCED SIX DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING TWO DIFFERENT EVENTS. THIS IS THE SECOND OF TWO REPORTS. REFER TO 9611594-2026-00077 FOR THE FIRST EVENT. IT WAS REPORTED THAT DURING THE USE OF AN INTRODUCER KIT DURING A PROCEDURE; WHILE PEELING AWAY THE PEEL AWAY SHEATH, "THE DILATOR SLIPPED OVER THE GUIDEWIRE DURING REMOVAL OF THE PEEL AWAY SHEATH. FORTUNATELY, THE DOCTOR WAS ABLE TO CATCH THE DILATOR BEFORE IT ENTERED THE STOMACH."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24392173 ↗MalfunctionAug 21, 2023Premature Separation; Premature Separation
IT WAS REPORTED, ALL THREE T-ANCHORS CAME LOOSE AFTER JUST TWO DAYS OF INSTALLATION. PER ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2023, THERE WAS NO NEED FOR MEDICAL INTERVENTIONS. THE PATIENT¿S STATUS WAS REPORTED TO BE ¿OK, NO PROBLEMS.¿
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17586678 ↗MalfunctionSep 15, 2020Break; Break
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE SECOND OF TWO REPORTS. REFER TO 9611594-2020-00171 FOR THE FIRST REPORT. IT WAS REPORTED THAT THE ANCHOR FELL OFF. THE PATIENT'S CONDITION NOW IS FINE, AND A NEW DEVICE WAS USED. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10537529 ↗MalfunctionSep 15, 2020Break; Break
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE FIRST OF TWO REPORTS. REFER TO 9611594-2020-00172 FOR THE SECOND REPORT. IT WAS REPORTED THAT THE ANCHOR FELL OFF. THE PATIENT'S CONDITION NOW IS FINE, AND A NEW DEVICE WAS USED. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10537499 ↗MalfunctionSep 15, 2020Break; Break
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING AN UNKNOWN NUMBER OF PATIENTS. THIS IS THE THIRD OF THREE REPORTS. REFER TO 9611594-2020-00180 FOR THE FIRST REPORT. REFER TO 9611594-2020-00181 FOR THE SECOND REPORT. IT WAS REPORTED THAT THE "BUTTON FELL OFF WHEN THE DEVICE WAS APPLIED." THE BUTTON IS THE ANCHOR. NO PATIENT INJURY WAS REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT YET RECEIVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10535382 ↗MalfunctionSep 15, 2020Break; Break
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING AN UNKNOWN NUMBER OF PATIENTS. THIS IS THE SECOND OF THREE REPORTS. REFER TO 9611594-2020-00180 FOR THE FIRST REPORT. REFER TO 9611594-2020-00182 FOR THE THIRD REPORT. IT WAS REPORTED THAT THE "BUTTON FELL OFF WHEN THE DEVICE WAS APPLIED." THE BUTTON IS THE ANCHOR. NO PATIENT INJURY WAS REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT YET RECEIVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10535379 ↗MalfunctionSep 15, 2020Break; Break
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING AN UNKNOWN NUMBER OF PATIENTS. THIS IS THE FIRST OF THREE REPORTS. REFER TO 9611594-2020-00181 FOR THE SECOND REPORT. REFER TO 9611594-2020-00182 FOR THE THIRD REPORT . IT WAS REPORTED THAT THE "BUTTON FELL OFF WHEN THE DEVICE WAS APPLIED." THE BUTTON IS THE ANCHOR. NO PATIENT INJURY WAS REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT YET RECEIVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10535378 ↗MalfunctionJun 26, 2020Break; Break
IT WAS REPORTED "3 T-FASTENERS IN THE KIT ALL BROKE UPON INSERTION." THERE WAS NO REPORTED INJURY. ADDITIONAL INFORMATION RECEIVED 11-JUN-2020 STATED, "ALL 3 SUTURES THAT WERE INITIALLY PLACED BROKE WHILE ADVANCING THE SERIAL DILATOR, DISPLACING THE STOMACH AND INTRODUCING A LARGE AMOUNT OF FREE AIR." THE PATIENT HAD A POSITRON EMISSION TOMOGRAPHY (PET) SCAN THE FOLLOWING WEEK WHICH SHOWED A LARGE PNEUMOPERITONEUM, WHICH WAS CAUSING SIGNIFICANT PAIN.
Patient problems: No Code Available; Pain
MAUDE report 10201464 ↗
Other TUBE, GASTRO-ENTEROSTOMY devices
- AVANOS 98431Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY93 reports
- AVANOS 98432Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY72 reports
- Gastrointestinal Anchor Set, SAF-T-PEXY* T-Fasteners 98701Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY1 death61 reports
- Halyard 98701Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY1 death40 reports
- Halyard 98431Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY35 reports
- AVANOS 98433Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY34 reports
- AVANOS 98430Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY1 death29 reports
- Halyard 98433Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY27 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.