AVANOS 98705Recalls & reported problems
Avanos Medical, Inc. · Enteral Access Dilation System · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00350770987057), received between Nov 9, 2021 and Nov 9, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 9, 2021Break; Break
IT WAS REPORTED A PIECE OF A TELESCOPING DILATOR BROKE OFF INSIDE OF A PATIENT. ADDITIONAL INFORMATION RECEIVED 14-OCT-2021 STATED THE PHYSICIAN "NOTICED WHEN PULLING THE DILATORS BACK OUT, IT WAS DIFFICULT AND THE INNERMOST ONE DID NOT CATCH ON THE SECOND ONE AS IT SHOULD HAVE AND THE SECOND ONE ADVANCED INTO THE PATIENT. UPON REMOVAL OF THE DILATORS, THE PHYSICIAN COULD BARELY SEE THE ONE THAT WAS IN THE STOMACH, ATTEMPTED TO REMOVE IT VIA SNARE (WAS UNSUCCESSFUL). THE G-TUBE WAS PLACED AND THE PATIENT WENT TO THE GI (GASTRO INTESTINAL) LAB TO HAVE THIS FOREIGN BODY REMOVED." THE BROKEN PIECE WAS REMOVED VIA ENDOSCOPY ON (B)(6) 2021. THE PATIENT WAS DOING WELL, NO COMPLICATIONS.
Patient problems: Foreign Body In Patient
MAUDE report 12781609 ↗
FDA clearance and classification
Cleared through Special 510(k) K093312 on Dec 8, 2009 (decision: Substantially Equivalent). Submitted by Kimberly-Clark Corp. as “Kimberly-Clark Gastrointestinal Anchor Set With Saf-T-Pexy Fasteners, Enternal Access Dilation System”. The FDA review took 47 days. FDA record ↗
The FDA regulates product code KGC (tube, gastro-enterostomy) under 21 CFR 876.5980 as a Class II device. 21 CFR 876.5980 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the AVANOS 98705
Has the AVANOS 98705 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00350770987057), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the AVANOS 98705?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Nov 9, 2021. That report lists device problem “Break” (1); patient problem “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the AVANOS 98705?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the AVANOS 98705 FDA approved?
It was cleared through Special 510(k) K093312 on Dec 8, 2009. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the AVANOS 98705 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the AVANOS 98705, and is it still sold?
The FDA's device database lists Avanos Medical, Inc. as the company responsible for it (brand AVANOS, model 98705). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other TUBE, GASTRO-ENTEROSTOMY devices
- AVANOS 98431Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY93 reports
- AVANOS 98432Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY72 reports
- Gastrointestinal Anchor Set, SAF-T-PEXY* T-Fasteners 98701Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY1 death61 reports
- Halyard 98701Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY1 death40 reports
- Halyard 98431Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY35 reports
- AVANOS 98433Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY34 reports
- AVANOS 98430Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY1 death29 reports
- Halyard 98433Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY27 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.