Portex 4917M-19Recalls & reported problems
Icu Medical, Inc. · Class 2
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 00351688047574), received between May 11, 2018 and Aug 27, 2024 — classified as 1 injury, 18 malfunctions.
The most frequently reported specific device problem is “Output Problem” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Output Problem10
- Break5
- Leak/Splash5
- Material Puncture/Hole3
- Device Markings/Labelling Problem1
- Fluid/Blood Leak1
- Insufficient Device Problem Information1
- Material Fragmentation1
Patient problems
- Blood Loss10
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient3
- Foreign Body In Patient1
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 27, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT WHEN THE CATHETER WAS REMOVED, BLACK COLOR WAS NOTED UNDER THE DRESSING. WHEN THE CATHETER WAS PULLED, IT WAS FURTHER NOTED THAT THE CM MARKINGS HAD ¿RUBBED OFF¿ THE CATHETER AND WERE NO LONGER VISIBLE ON THE CATHETER BUT ON THE SKIN AND MAYBE IN THE PATIENT ALONG THE TRACT. THERE WAS PATIENT INVOLVEMENT, AND UNKNOW PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 20081157 ↗MalfunctionAug 29, 2022Break; Leak/Splash; Break; Leak/Splash
IT WAS REPORTED THAT THE CATHETER DISCONNECTED FROM THE EPIDURAL CATHETER DURING LABOR. FEW OTHER INSTANCES THE CATHETER APPEARED TO HAVE A HOLE ALLOWING INFUSION SOLUTION TO LEAK. ONE RECENT WHERE THE CATHETER BROKE OFF. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15316102 ↗MalfunctionAug 29, 2022Break; Leak/Splash; Break; Leak/Splash
IT WAS REPORTED THAT CATHETER DISCONNECTED FROM THE EPIDURAL CATHETER DURING LABOR. FEW OTHER INSTANCES THE CATHETER APPEARED TO HAVE A HOLE ALLOWING INFUSION SOLUTION TO LEAK. ONE RECENT WHERE THE CATHETER BROKE OFF. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15315919 ↗MalfunctionAug 29, 2022Break; Fluid/Blood Leak; Material Puncture/Hole; Break; Fluid/Blood Leak; Material Puncture/Hole
IT WAS REPORTED THAT THE CATHETER DISCONNECTED FROM THE EPIDURAL CATHETER DURING LABOR, CATHETER HAD A HOLE ALLOWING INFUSION SOLUTION TO LEAK, AND CATHETER BROKE OFF. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15315664 ↗MalfunctionAug 29, 2022Break; Leak/Splash; Break; Leak/Splash
IT WAS REPORTED THAT CATHETER DISCONNECTED FROM THE EPIDURAL CATHETER DURING LABOR. FEW OTHER INSTANCES THE CATHETER APPEARED TO HAVE A HOLE ALLOWING INFUSION SOLUTION TO LEAK. ONE RECENT WHERE THE CATHETER BROKE OFF. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15315302 ↗InjuryApr 4, 2021Break; Material Fragmentation; Break; Material Fragmentation
IT WAS REPORTED THAT A SMITHS MEDICAL CATHETER BROKE OFF WHILE INSERTED IN THE PATIENT. WHEN THE ANESTHESIOLOGIST WENT TO REMOVE THE CATHETER AND REPOSITION IT, THE TIP WAS DISCONNECTED. CT SCAN WAS PERFORMED AND CONFIRMED IT REMAINED INSIDE OF PATIENTS BACK. PATIENT CURRENTLY HAS NO NEUROLOGIC SYMPTOMS AND NEUROSURGERY WAS CONSULTED AND RECOMMENDED NO CURRENT NEED FOR SURGICAL INTERVENTION. NO FURTHER MEDICAL INTERVENTION WAS NECESSARY. THE PATIENT WAS DISCHARGED STABLE, AND ANESTHESIA WILL CONTINUE TO FOLLOW UP WITH THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11611915 ↗MalfunctionFeb 26, 2020Leak/Splash; Material Puncture/Hole; Leak/Splash; Material Puncture/Hole
INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL PORTEX® EPIDURAL CATHETER BECAME DISCONNECTED FROM THE DURAFLEX CATHETER. A HOLE WAS REPORTED IN WHICH THE INFUSION LEAKED. THERE WERE NO REPORTED ADVERSE EFFECTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9753982 ↗MalfunctionFeb 26, 2020Leak/Splash; Material Puncture/Hole; Leak/Splash; Material Puncture/Hole
INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL PORTEX® EPIDURAL CATHETER BECAME DISCONNECTED FROM THE DURAFLEX CATHETER. A HOLE WAS REPORTED IN WHICH THE INFUSION LEAKED. THERE WERE NO REPORTED ADVERSE EFFECTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9753980 ↗MalfunctionOct 2, 2019Output Problem; Output Problem
THE REPORTER STATED THE USER ATTEMPTED TO INSERT THE CATHETER IN PATIENT EPIDURAL SPACE BUT THE CATHETER WAS "STIFF" AND THIS ISSUE CAUSED THE USER TO "HIT VASCULAR VESSELS" WHICH CAUSED BLEEDING. THE PROCEDURE WAS NOT COMPLETED. NO ADVERSE EFFECTS TO PATIENT REPORTED.
Patient problems: Blood Loss
MAUDE report 9145295 ↗MalfunctionAug 19, 2019Output Problem; Output Problem
THE REPORTER STATED THE USER ATTEMPTED TO INSERT THE CATHETER IN PATIENT EPIDURAL SPACE BUT THE CATHETER WAS "STIFF" AND THIS ISSUE CAUSED THE USER TO "HIT VASCULAR VESSELS" WHICH CAUSED BLEEDING. THE PROCEDURE WAS NOT COMPLETED. NO ADVERSE EFFECTS TO PATIENT REPORTED.
Patient problems: Blood Loss
MAUDE report 8906417 ↗MalfunctionJun 21, 2019Output Problem; Output Problem
THE REPORTER STATED THE USER ATTEMPTED TO INSERT THE CATHETER IN PATIENT EPIDURAL SPACE BUT THE CATHETER WAS "STIFF" AND THIS ISSUE CAUSED THE USER TO "HIT VASCULAR VESSELS" WHICH CAUSED BLEEDING. THE PROCEDURE WAS NOT COMPLETED. NO ADVERSE EFFECTS TO PATIENT REPORTED.
Patient problems: Blood Loss
MAUDE report 8722823 ↗MalfunctionJun 21, 2019Output Problem; Output Problem
THE REPORTER STATED THE USER ATTEMPTED TO INSERT THE CATHETER IN PATIENT EPIDURAL SPACE BUT THE CATHETER WAS "STIFF" AND THIS ISSUE CAUSED THE USER TO "HIT VASCULAR VESSELS" WHICH CAUSED BLEEDING. THE PROCEDURE WAS NOT COMPLETED. NO ADVERSE EFFECTS TO PATIENT REPORTED.
Patient problems: Blood Loss
MAUDE report 8722487 ↗MalfunctionJun 20, 2019Output Problem; Output Problem
THE REPORTER STATED THE USER ATTEMPTED TO INSERT THE CATHETER IN PATIENT EPIDURAL SPACE BUT THE CATHETER WAS "STIFF" AND THIS ISSUE CAUSED THE USER TO "HIT VASCULAR VESSELS" WHICH CAUSED BLEEDING. THE PROCEDURE WAS NOT COMPLETED. NO ADVERSE EFFECTS TO PATIENT REPORTED.
Patient problems: Blood Loss
MAUDE report 8718432 ↗MalfunctionJun 20, 2019Output Problem; Output Problem
THE REPORTER STATED THE USER ATTEMPTED TO INSERT THE CATHETER IN PATIENT EPIDURAL SPACE BUT THE CATHETER WAS "STIFF" AND THIS ISSUE CAUSED THE USER TO "HIT VASCULAR VESSELS" WHICH CAUSED BLEEDING. THE PROCEDURE WAS NOT COMPLETED. NO ADVERSE EFFECTS TO PATIENT REPORTED.
Patient problems: Blood Loss
MAUDE report 8718431 ↗MalfunctionJun 20, 2019Output Problem; Output Problem
THE REPORTER STATED THE USER ATTEMPTED TO INSERT THE CATHETER IN PATIENT EPIDURAL SPACE BUT THE CATHETER WAS "STIFF" AND THIS ISSUE CAUSED THE USER TO "HIT VASCULAR VESSELS" WHICH CAUSED BLEEDING. THE PROCEDURE WAS NOT COMPLETED. NO ADVERSE EFFECTS TO PATIENT REPORTED.
Patient problems: Blood Loss
MAUDE report 8718430 ↗MalfunctionJun 20, 2019Output Problem; Output Problem
THE REPORTER STATED THE USER ATTEMPTED TO INSERT THE CATHETER IN PATIENT EPIDURAL SPACE BUT THE CATHETER WAS "STIFF" AND THIS ISSUE CAUSED THE USER TO "HIT VASCULAR VESSELS" WHICH CAUSED BLEEDING. THE PROCEDURE WAS NOT COMPLETED. NO ADVERSE EFFECTS TO PATIENT REPORTED.
Patient problems: Blood Loss
MAUDE report 8718429 ↗MalfunctionJun 20, 2019Output Problem; Output Problem
THE REPORTER STATED THE USER ATTEMPTED TO INSERT THE CATHETER IN PATIENT EPIDURAL SPACE BUT THE CATHETER WAS "STIFF" AND THIS ISSUE CAUSED THE USER TO "HIT VASCULAR VESSELS" WHICH CAUSED BLEEDING. THE PROCEDURE WAS NOT COMPLETED. NO ADVERSE EFFECTS TO PATIENT REPORTED.
Patient problems: Blood Loss
MAUDE report 8718428 ↗MalfunctionJun 20, 2019Output Problem; Output Problem
THE REPORTER STATED THE USER ATTEMPTED TO INSERT THE CATHETER IN PATIENT EPIDURAL SPACE BUT THE CATHETER WAS "STIFF" AND THIS ISSUE CAUSED THE USER TO "HIT VASCULAR VESSELS" WHICH CAUSED BLEEDING. THE PROCEDURE WAS NOT COMPLETED. NO ADVERSE EFFECTS TO PATIENT REPORTED.
Patient problems: Blood Loss
MAUDE report 8718425 ↗MalfunctionMay 11, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT FELT A WET SPOT ON BACK. WHEN THE NURSING STAFF ASSESSED, THEY SAW THE TIP OF THE CATHETER THAT CONNECTS TO THE IV LINE HAD DISCONNECTED FROM THE CATHETER LINE, AND FLUID WAS LEAKING OUT AS A RESULT. BROKEN CATHETER HAD TO BE REMOVED, NEW EPIDURAL CATHETER HAD TO BE INSERTED. NO PATIENT HARM REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7507711 ↗
Other Anesthesia conduction kit devices
- BD Luer-Lok, BD Glaspak 405671Becton, Dickinson And Company · Anesthesia conduction kit1 recall68 reports
- ARROW IPN922678Teleflex Incorporated · Anesthesia conduction kit2 recalls0 reports
- BD Luer-Lok, BD Glaspak 405672Becton, Dickinson And Company · Anesthesia conduction kit1 recall27 reports
- BD Luer-Lok, BD Glaspak 405673Becton, Dickinson And Company · Anesthesia conduction kit1 recall10 reports
- BD Luer-Lok, BD Glaspak 405637Becton, Dickinson And Company · Anesthesia conduction kit1 recall7 reports
- BD Luer-Lok, BD Glaspak 405707Becton, Dickinson And Company · Anesthesia conduction kit1 recall4 reports
- BD Luer-Lok, BD Glaspak 405609Becton, Dickinson And Company · Anesthesia conduction kit1 recall3 reports
- BD Luer-Lok, BD Glaspak 405735Becton, Dickinson And Company · Anesthesia conduction kit1 recall3 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.