OneTouch Ultramini 021-422Recalls & reported problems
LifeScan Europe GmbH · OneTouch Ultramini Blood Glucose Monitoring System with Delica Lancing Device, not for resale or reimbursement. Uses OneTouch Ultra Blue Test strip · Class 2
All models of OneTouch Ultramini (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00353885001311), received between Feb 3, 2017 and Aug 21, 2019 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Incorrect Or Inadequate Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Incorrect Or Inadequate Test Results1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- Itching Sensation1
- Nausea1
- No Consequences Or Impact To Patient1
- Sweating1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 21, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON (B)(6) 2019, THE REPORTER CONTACTED LIFESCAN UNITED STATES, ALLEGING THAT THE SUBJECT METER READ INACCURATELY ERRATIC. THE REPORTER CLAIMED OBTAINING BLOOD GLUCOSE READINGS OF ¿218 AND 334 MG/DL¿ WITH THE SUBJECT METER, PERFORMED WITHIN 20 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN¿S CRITERIA FOR PRECISION. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8914261 ↗InjuryFeb 3, 2017Incorrect Or Inadequate Test Results; Incorrect Or Inadequate Test Results
ON (B)(6) 2017, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) USA, ALLEGING THAT HER ONETOUCH ULTRAMINI METER WAS READING INACCURATELY HIGH COMPARED TO ANOTHER ONETOUCH ULTRAMINI. THE COMPLAINT WAS CLASSIFIED BASED ON INFORMATION OBTAINED FROM THE CUSTOMER SERVICE REPRESENTATIVE (CSR) DOCUMENTATION, SINCE THE PATIENT WAS UNABLE TO BE REACHED BY PHONE FOR ADDITIONAL INFORMATION. THE PATIENT REPORTED THAT THE ALLEGED ISSUE BEGAN AT 2 A.M., ON (B)(6) 2017. THE PATIENT CLAIMED OBTAINING BLOOD GLUCOSE READINGS OF ¿109 MG/DL¿ WITH THE SUBJECT METER AND ¿70 MG/DL¿ WITH THE OTHER DEVICE, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LFS¿S CRITERIA FOR ACCURACY. THE PATIENT ADVISED THAT SHE DOES NOT TAKE ANY MEDICATION TO MANAGE HER DIABETES AND REPORTED THAT APPROXIMATELY 10 MINUTES PRIOR TO OBTAINING THE ALLEGED INACCURATELY HIGH BLOOD GLUCOSE READING, SHE DEVELOPED SYMPTOMS OF ¿ITCHING, SWEATING AND NAUSEA¿. THE PATIENT ADVISED THAT SHE SELF-TREATED HER SYMPTOMS BY DRINKING 3 OZ., OF ORANGE JUICE. AT THE TIME OF TROUBLESHOOTING, THE CSR NOTED THAT THE UNIT OF MEASURE WAS SET CORRECTLY ON THE SUBJECT
Patient problems: Itching Sensation; Nausea; Sweating
MAUDE report 6299232 ↗
Other SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER devices
- OneTouch Verio Flex 023-271LifeScan Europe GmbH · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER3,703 reports
- FreeStyle Libre 2 71953Abbott Diabetes Care Inc · System, test, blood glucose, over the counter2 recalls1 death3,286 reports
- OneTouch Ultra2 021-105LifeScan U.S. LLC · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER1,731 reports
- Omnipod DASH PDM PDM-USA1-D001-MG-USA1Insulet Corporation · System, Test, Blood Glucose, Over The Counter1 recall10 deaths1,129 reports
- Omnipod UST400Insulet Corporation · System, Test, Blood Glucose, Over The Counter12 deaths816 reports
- TRUE METRIX TRUE METRIX SELF MONITORING BLOOD GLUCOSE METER - NOT FOR RETAIL SALETrividia Health, Inc. · System, Test, Blood Glucose, Over The Counter418 reports
- FreeStyle Freedom Lite 70923-70Abbott Diabetes Care Inc · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER378 reports
- FreeStyle Lite 70829-70Abbott Diabetes Care Inc · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER378 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page