OneTouch Ultra Blue 022-895Recalls & reported problems
LifeScan U.S. LLC · OneTouch Ultra Test Strips Blue for blood glucose testing with: OneTouch Ultra 2, OneTouch UltraMini, OneTouch Ultra, OneTouch UltraSmart, OneTouch Ping and OneTouch UltraLink Meters · Class 2
All models of OneTouch Ultra Blue (3)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00353885009713), received between Nov 3, 2017 and Aug 11, 2022 — classified as 6 injuries, 1 malfunction.
The most frequently reported specific device problem is “Peeled/Delaminated” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Mar 16, 2021Class IIOpen, Classified
Due to shipping delays the devices were exposed to temperature excursions of 31.9F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results.
Z-1936-2021 · Root cause: Environmental control · Recall details · FDA record ↗
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available2
- Peeled/Delaminated2
- Bent1
- Defective Component1
- Physical Resistance/Sticking1
Patient problems
- Shaking/Tremors3
- Fatigue2
- Hyperglycemia2
- Nausea2
- Anxiety1
- Dizziness1
- Hypoglycemia1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryAug 11, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ON (B)(6) 2022, THE LAY-USER/PATIENT CONTACTED LIFESCAN (LFS) USA, ALLEGING THAT HER ONE TOUCH ULTRA TEST TRIPS WERE ¿MESSED UP¿ AND ¿DIDN T REGISTER¿. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE AGENT (CCA) DOCUMENTATION AND ON ADDITIONAL INFORMATION OBTAINED DURING A FOLLOW-UP CALL WITH THE PATIENT. THE PATIENT DID NOT REPORT WHEN THE ALLEGED STRIP ISSUE STARTED. DURING THE FOLLOW-UP CALL THE PATIENT STATED THAT SHE MANAGES HER DIABETES WITH AN INSULIN PUMP AND PILLS (TYPE AND DOSE UNSPECIFIED) AND INDICATED THAT SHE DID NOT MADE ANY CHANGES TO HER DIABETES MANAGEMENT REGIMEN IN RESPONSE TO THE ALLEGED ISSUE. THE PATIENT INFORMED THE AGENT THAT DUE TO THE ALLEGED STRIP ISSUE HER BLOOD SUGAR LEVELS WENT UP TO 500 MG/DL FOR THREE DAYS. DURING THE FOLLOW-UP CALL THE PATIENT CONFIRMED THAT SHE WAS ABLE TO RECEIVE THESE READINGS WITH THE SUBJECT TEST STRIPS. THE PATIENT DID NOT REPORT ANY SYMPTOMS AND DENIED RECEIVING ANY MEDICAL TREATMENT DUE TO THE ALLEGED ISSUE. THE CCA WAS NOT ABLE TO TROUBLESHOOT THE PROBLEM DURING THE INITIAL CALL AND THE PATIENT WAS UNABLE TO READ THE LOT NUMBER. REPLACEMENT TEST STRIPS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECA
Patient problems: Hyperglycemia
MAUDE report 15214728 ↗InjuryAug 11, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ON (B)(6) 2022, THE LAY-USER/PATIENT CONTACTED LIFESCAN (LFS) USA, ALLEGING THAT HER ONETOUCH ULTRA TEST TRIPS WERE ¿MESSED UP¿ AND ¿DIDN¿T REGISTER¿. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE AGENT (CCA) DOCUMENTATION AND ON ADDITIONAL INFORMATION OBTAINED DURING A FOLLOW-UP CALL WITH THE PATIENT. THE PATIENT DID NOT REPORT WHEN THE ALLEGED STRIP ISSUE STARTED. DURING THE FOLLOW-UP CALL THE PATIENT STATED THAT SHE MANAGES HER DIABETES WITH AN INSULIN PUMP AND PILLS (TYPE AND DOSE UNSPECIFIED) AND INDICATED THAT SHE DID NOT MADE ANY CHANGES TO HER DIABETES MANAGEMENT REGIMEN IN RESPONSE TO THE ALLEGED ISSUE. THE PATIENT INFORMED THE AGENT THAT DUE TO THE ALLEGED STRIP ISSUE HER BLOOD SUGAR LEVELS WENT UP TO 500 MG/DL FOR THREE DAYS. DURING THE FOLLOW-UP CALL THE PATIENT CONFIRMED THAT SHE WAS ABLE TO RECEIVE THESE READINGS WITH THE SUBJECT TEST STRIPS. THE PATIENT DID NOT REPORT ANY SYMPTOMS AND DENIED RECEIVING ANY MEDICAL TREATMENT DUE TO THE ALLEGED ISSUE. THE CCA WAS NOT ABLE TO TROUBLESHOOT THE PROBLEM DURING THE INITIAL CALL AND THE PATIENT WAS UNABLE TO READ THE LOT NUMBER. REPLACEMENT TEST STRIPS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAU
Patient problems: Hyperglycemia
MAUDE report 15214727 ↗MalfunctionAug 26, 2019Physical Resistance/Sticking; Physical Resistance/Sticking
THE LAY USER/PATIENTS TEST STRIPS HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE TEST STRIPS FAILED TESTING. THE TEST STRIPS APPEAR TO HAVE BEEN CONTAMINATED WITH EXCESS ADHESIVE. COMPLAINTS RELATING TO THE REPORTED PRODUCT(S) WERE EVALUATED. IT WAS CONCLUDED THAT THE NUMBER OF COMPLAINTS FOR THE PRODUCT(S) DID NOT BREACH THRESHOLDS INDICATIVE OF A SYSTEMIC ISSUE. ON (B)(6) 2018, THE LAY USER/PATIENT CONTACTED LIFESCAN USA, ALLEGING THAT THE SUBJECT OT ULTRA BLUE TEST STRIPS WERE STICKING TOGETHER. THIS COMPLAINT WAS INITIALLY RULED OUT BECAUSE DURING CUSTOMER SERVICES TROUBLESHOOTING, THERE WAS NO INDICATION TO REASONABLY SUGGEST THAT THE PRODUCT MALFUNCTIONED AND THERE WAS ALSO NO ALLEGATION OF AN ADVERSE EVENT AS A RESULT OF THE REPORTED ISSUE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8929425 ↗InjuryJul 15, 2019Peeled/Delaminated; Peeled/Delaminated
ON (B)(6) 2019, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) USA, ALLEGING THAT HIS ONETOUCH ULTRA BLUE TEST STRIPS WERE PEELING AND STICKING TOGETHER. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION OF THE CALL. THE PATIENT REPORTED THAT THE TEST STRIPS WERE PEELING AND STICKING TOGETHER AROUND 9-10 AM ON (B)(6) 2019. THE PATIENT MANAGES HIS DIABETES WITH HUMULIN INSULIN (NO ADJUSTMENTS). HE INDICATED THAT BECAUSE HE WAS UNABLE TO OBTAIN A READING, HE ATE SUGAR TO INCREASE HIS BLOOD GLUCOSE LEVEL ABOUT 30-45 MINUTES LATER. HE REPORTED THAT ABOUT AN HOUR LATER, HE FELT SYMPTOMS OF "NAUSEA, FATIGUE, SHAKING AND ANXIETY". HE DENIED RECEIVING ANY TREATMENT FOR HIS SYMPTOMS ABOVE OR BEYOND THE USUAL ROUTINE OF DIABETES CARE AND MANAGEMENT. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THAT ALLEGED ISSUE OCCURRED BEFORE INSERTING THE STRIP INTO THE METER AND MORE THAN ONE TEST STRIP FROM THE VIAL WAS AFFECTED. THE ISSUE REMAINED UNRESOLVED. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT REPORTEDLY DEVELOPED SYMPTOMS SUGGESTIVE OF A SERIOUS INJURY ADVERSE EVENT, AFTER THE ALLEGED ISSUE WITH PEELING
Patient problems: Anxiety; Fatigue; Nausea; Shaking/Tremors
MAUDE report 8789109 ↗InjuryJul 15, 2019Peeled/Delaminated; Peeled/Delaminated
ON (B)(6) 2019, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) USA, ALLEGING THAT HIS ONETOUCH ULTRA BLUE TEST STRIPS WERE PEELING AND STICKING TOGETHER. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION OF THE CALL. THE PATIENT REPORTED THAT THE TEST STRIPS WERE PEELING AND STICKING TOGETHER AROUND 9-10 AM ON (B)(6) 2019. THE PATIENT MANAGES HIS DIABETES WITH HUMULIN INSULIN (NO ADJUSTMENTS). HE INDICATED THAT BECAUSE HE WAS UNABLE TO OBTAIN A READING, HE ATE SUGAR TO INCREASE HIS BLOOD GLUCOSE LEVEL ABOUT 30-45 MINUTES LATER. HE REPORTED THAT ABOUT AN HOUR LATER, HE FELT SYMPTOMS OF ¿NAUSEA, FATIGUE, SHAKING AND ANXIETY¿. HE DENIED RECEIVING ANY TREATMENT FOR HIS SYMPTOMS ABOVE OR BEYOND THE USUAL ROUTINE OF DIABETES CARE AND MANAGEMENT. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THAT ALLEGED ISSUE OCCURRED BEFORE INSERTING THE STRIP INTO THE METER AND MORE THAN ONE TEST STRIP FROM THE VIAL WAS AFFECTED. THE ISSUE REMAINED UNRESOLVED. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT REPORTEDLY DEVELOPED SYMPTOMS SUGGESTIVE OF A SERIOUS INJURY ADVERSE EVENT, AFTER THE ALLEGED ISSUE WITH PEELING
Patient problems: Fatigue; Nausea; Shaking/Tremors
MAUDE report 8788399 ↗InjuryJan 26, 2018Defective Component; Defective Component
ON (B)(6) 2018, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) USA, ALLEGING AN ISSUE WITH HER ONETOUCH ULTRA TEST STRIP VIAL. THE PATIENT CLAIMED THAT THE VIAL INSIDE A SEALED CONTAINER WAS OPEN. THE COMPLAINT WAS CLASSIFIED BASED ON FURTHER INFORMATION OBTAINED BY THE MEDICAL SURVEILLANCE SPECIALIST AFTER REVIEWING THE CALL RECORDING SINCE THE PATIENT WAS UNABLE TO BE CONTACTED BY PHONE FOR ADDITIONAL INFORMATION. THE PATIENT REPORTED THAT SHE BECAME AWARE OF THE ALLEGED OPEN VIAL ON (B)(6) 2018 AT 8:00 AM. THE PATIENT INFORMED THE CSR THAT SHE MANAGES HER DIABETES WITH A COMBINATION OF ORAL MEDICATION (METFORMIN) AND INSULIN (HUMULIN). THE PATIENT DENIED MAKING ANY CHANGES TO HER USUAL DIABETES MANAGEMENT REGIMEN WHEN SHE WAS UNABLE TO TEST HER BLOOD GLUCOSE DUE TO THE ALLEGED ISSUE. DURING THE CALL, THE PATIENT REPORTED DEVELOPING ¿LOW BLOOD SUGAR¿ AS A RESULT OF THE ALLEGED ISSUE. THE PATIENT CLAIMED SHE PROCEEDED TO THE URGENT CARE CLINIC ON (B)(6) 2018 AT 10:00 AM WHERE HER BLOOD GLUCOSE WAS MEASURED AT ¿48 MG/DL¿ ON THE CLINIC DEVICE. DURING THE CALL, THE PATIENT CLAIMED SHE RECEIVED UNSPECIFIED TREATMENT AND WAS REPORTEDLY ADVISED TO INCREASE HER DAILY DOSE OF METFORMIN MED
Patient problems: Hypoglycemia
MAUDE report 7220952 ↗InjuryNov 3, 2017Bent; Bent
ON (B)(6) 2017 THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) ALLEGING AN ISSUE WITH HER ONETOUCH ULTRA BLUE TEST STRIPS. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER SERVICE REPRESENTATIVE (CSR) DOCUMENTATION AND ADDITIONAL INFORMATION OBTAINED WHEN THE MEDICAL SURVEILLANCE SPECIALIST LISTENED AGAIN TO THE CALL RECORDING. THE PATIENT STATED THAT THE ALLEGED ISSUE BEGAN AROUND 1 MONTH AGO, HOWEVER COULD NOT REMEMBER THE SPECIFIC DATE OR TIME WHEN SHE NOTED THAT APPROXIMATELY HALF OF THE TEST STRIPS WERE BENT IN THE VIAL SHE RECEIVED FROM THE PHARMACY. THE PATIENT MANAGES HER DIABETES USING LANTUS INSULIN (SELF-ADJUSTS) AND SHE REPORTEDLY DECREASED HER FOOD/DRINK CONSUMPTION AND TOOK INSULIN AT THE TIME THE PRODUCT ISSUE OCCURRED. THE PATIENT CLAIMED EXPERIENCING SYMPTOMS OF ¿SHAKING, DIZZY AND LIGHT-HEADED¿ AN UNSPECIFIED AMOUNT OF TIME AFTER THE ALLEGED ISSUE BEGAN, AND REPORTEDLY VISITED THE HOSPITAL WHERE SHE RECEIVED HCP TREATMENT OF ¿SWEET FOOD AND APPLE JUICE¿ IN RESPONSE TO THE REPORTED ISSUE. THE PATIENT CONFIRMED THAT HER BLOOD GLUCOSE WAS NOT MEASURED USING ANY OTHER DEVICE AT THE TIME. THE CSR WAS UNABLE TO RESOLVE THE ISSUE DURING TROUBLESHOOTING. REPLACEMENT PROD
Patient problems: Dizziness; Shaking/Tremors
MAUDE report 6998549 ↗
FDA clearance and classification
The FDA regulates product code CGA (glucose oxidase, glucose) under 21 CFR 862.1345 as a Class II device. 21 CFR 862.1345 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the OneTouch Ultra Blue 022-895
Has the OneTouch Ultra Blue 022-895 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1936-2021 (Class II), began on Mar 16, 2021: Due to shipping delays the devices were exposed to temperature excursions of 31.9F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the OneTouch Ultra Blue 022-895?
The FDA's MAUDE database lists 7 adverse event reports for this device, received between Nov 3, 2017 and Aug 11, 2022. In those reports, the most common specific device problems are “Peeled/Delaminated” (2), “Bent” (1) and “Defective Component” (1); the most common patient problems are “Shaking/Tremors” (3), “Fatigue” (2) and “Hyperglycemia” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the OneTouch Ultra Blue 022-895?
None of the reports are classified as deaths, 6 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the OneTouch Ultra Blue 022-895, and is it still sold?
The FDA's device database lists LifeScan U.S. LLC as the company responsible for it (brand OneTouch Ultra Blue, model 022-895). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other GLUCOSE OXIDASE, GLUCOSE devices
- TRUEtrack TRUEtrack 50 Blood Glucose Test StripsTrividia Health, Inc. · Glucose Oxidase, Glucose153 reports
- HemoCue® Glucose 201 Microcuvettes 110706Hemocue AB · Glucose Oxidase, Glucose2 recalls0 reports
- VITROS 1707801Ortho-Clinical Diagnostics, Inc. · GLUCOSE OXIDASE, GLUCOSE1 recall48 reports
- Prodigy AutoCode Blood Glucose Monitoring System 051885Prodigy Diabetes Care, LLC · Glucose Oxidase, Glucose93 reports
- OneTouch Ultra Blue 020-994LifeScan U.S. LLC · GLUCOSE OXIDASE, GLUCOSE1 recall2 reports
- Alere Cholestech LDX® 10-991Alere San Diego, Inc. · Glucose Oxidase, Glucose1 recall1 report
- Alere Cholestech LDX® 99021Alere San Diego, Inc. · Glucose Oxidase, Glucose1 recall0 reports
- Alere Cholestech LDX® 14-531Alere San Diego, Inc. · Glucose Oxidase, Glucose1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page