OneTouch Ultra Blue 020-994Recalls & reported problems
LifeScan U.S. LLC · OneTouch Ultra Test Strips Blue for blood glucose testing with: OneTouch Ultra 2, OneTouch UltraMini, OneTouch Ultra, OneTouch UltraSmart, OneTouch Ping and OneTouch UltraLink Meters · Class 2
All models of OneTouch Ultra Blue (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00353885994255), received between Sep 19, 2017 and Sep 27, 2017 — classified as 2 injuries.
The most frequently reported specific device problem is “Peeled/Delaminated” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Mar 16, 2021Class IIOpen, Classified
Due to shipping delays the devices were exposed to temperature excursions of 31.9F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results.
Z-1938-2021 · Root cause: Environmental control · Recall details · FDA record ↗
Most reported problems
Device problems
- Peeled/Delaminated2
Patient problems
- Fatigue2
- Headache2
- Sweating2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjurySep 27, 2017Peeled/Delaminated; Peeled/Delaminated
ON (B)(6) 2017, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) ALLEGING THAT HER ONETOUCH ULTRA TEST STRIPS WERE PEELING. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE PATIENT REPORTED THAT SHE FIRST NOTICED THAT THE TEST STRIPS WERE PEELING AT 7:00PM ON (B)(6) 2017; THE ALLEGED ISSUE OCCURRED WHEN THE TEST STRIPS WAS INSERTED INTO THE TEST STRIP PORT OF THE METER. THE PATIENT MANAGES HER DIABETES WITH ORAL MEDICATION, DIET AND/OR EXERCISE. SHE DENIED MAKING ANY CHANGES TO HER USUAL DIABETES MANAGEMENT REGIMEN AS A RESULT OF THE ALLEGED ISSUE. SHE REPORTED THAT ONE DAY AFTER THE ISSUE BEGAN, SHE DEVELOPED SYMPTOMS OF ¿HEADACHE, SWEATING AND TIRED¿. SHE DENIED RECEIVING ANY TREATMENT FOR HER SYMPTOMS ABOVE AND BEYOND HER USUAL DIABETES CARE AND MANAGEMENT ROUTINE. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THAT THE PATIENT WAS USING THE CORRECT TESTING STEPS. THE ALLEGED ISSUE OCCURRED AFTER INSERTING THE STRIP INTO THE METER. THE CCA WALKED THE PATIENT THROUGH INSERTING A NEW TEST STRIP AND THE ISSUE WAS RESOLVED. REPLACEMENT TEST STRIPS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT REPORTEDLY DEVELOPED
Patient problems: Fatigue; Headache; Sweating
MAUDE report 6896588 ↗InjurySep 19, 2017Peeled/Delaminated; Peeled/Delaminated
ON (B)(6) 2017, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) ALLEGING THAT HER ONETOUCH ULTRA TEST STRIPS WERE PEELING. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE PATIENT REPORTED THAT SHE FIRST NOTICED THAT THE TEST STRIPS WERE PEELING AT 7:00PM ON (B)(6) 2017; THE ALLEGED ISSUE OCCURRED WHEN THE TEST STRIPS WAS INSERTED INTO THE TEST STRIP PORT OF THE METER. THE PATIENT MANAGES HER DIABETES WITH ORAL MEDICATION, DIET AND/OR EXERCISE. SHE DENIED MAKING ANY CHANGES TO HER USUAL DIABETES MANAGEMENT REGIMEN AS A RESULT OF THE ALLEGED ISSUE. SHE REPORTED THAT ONE DAY AFTER THE ISSUE BEGAN, SHE DEVELOPED SYMPTOMS OF ¿HEADACHE, SWEATING AND TIRED¿. SHE DENIED RECEIVING ANY TREATMENT FOR HER SYMPTOMS ABOVE AND BEYOND HER USUAL DIABETES CARE AND MANAGEMENT ROUTINE. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THAT THE PATIENT WAS USING THE CORRECT TESTING STEPS. THE ALLEGED ISSUE OCCURRED AFTER INSERTING THE STRIP INTO THE METER. THE CCA WALKED THE PATIENT THROUGH INSERTING A NEW TEST STRIP AND THE ISSUE WAS RESOLVED. REPLACEMENT TEST STRIPS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT REPORTEDLY DEVELOPED
Patient problems: Fatigue; Headache; Sweating
MAUDE report 6875034 ↗
Other GLUCOSE OXIDASE, GLUCOSE devices
- TRUEtrack TRUEtrack 50 Blood Glucose Test StripsTrividia Health, Inc. · Glucose Oxidase, Glucose153 reports
- HemoCue® Glucose 201 Microcuvettes 110706Hemocue AB · Glucose Oxidase, Glucose2 recalls0 reports
- VITROS 1707801Ortho-Clinical Diagnostics, Inc. · GLUCOSE OXIDASE, GLUCOSE1 recall48 reports
- Prodigy AutoCode Blood Glucose Monitoring System 051885Prodigy Diabetes Care, LLC · Glucose Oxidase, Glucose93 reports
- OneTouch Ultra Blue 022-895LifeScan U.S. LLC · GLUCOSE OXIDASE, GLUCOSE1 recall7 reports
- Alere Cholestech LDX® 10-991Alere San Diego, Inc. · Glucose Oxidase, Glucose1 recall1 report
- Cholestech LDX™ 97991Alere San Diego, Inc. · Glucose Oxidase, Glucose1 recall0 reports
- Alere Cholestech LDX® 99021Alere San Diego, Inc. · Glucose Oxidase, Glucose1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page