FreeStyle Libre 71536Recalls & reported problems
Abbott Diabetes Care Inc · FreeStyle Libre Sensor Kit · Class 3
Quick summary
This Abbott FreeStyle Libre glucose sensor appears in 642 adverse event reports, including 174 that describe injuries and none that describe deaths. Most reports describe glucose readings that differ sharply from meter checks, and some mention low blood sugar, skin irritation, or loss of consciousness. Reports reached 19 in the last 12 months, up from 0 in the 12 months before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 640
- Deaths reported
- 0
- Injuries reported
- 174
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 640 adverse event reports that list this exact device (DI 00357599000196), received between Feb 28, 2018 and Aug 27, 2026 — classified as 174 injuries, 466 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (466 reports).
1 other recall cover devices cleared under the same FDA submission (P160030); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings466
- Appropriate Device Problem Term/Code Not Available75
- Detachment of Device or Device Component25
- Device Displays Incorrect Message23
- High Readings21
- Low Readings19
- Failure to Fire4
- Patient-Device Incompatibility3
Patient problems
- No Clinical Signs, Symptoms or Conditions456
- Hypoglycemia49
- Skin Irritation43
- Hyperglycemia37
- Loss of consciousness35
- Unspecified Infection28
- Itching Sensation25
- Dizziness17
Recent adverse event reports
The 25 most recent of 640 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 27, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 142.00 mg/dL, 161.00 mg/dL, 127.00 mg/dL compared to readings of 240.00 mg/dL, 295.00 mg/dL, 280.00 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 7 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50470839 ↗MalfunctionAug 26, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 491.00 mg/dL compared to glucose values of 129.00 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 11 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50443697 ↗MalfunctionAug 5, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50141491 ↗MalfunctionJul 9, 2026Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 258 MG/DL COMPARED TO READINGS OF 130 MG/DL RESPECTIVELY OBTAINED ON A METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO THE C ZONE, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS 11 DAYS. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25760813 ↗MalfunctionJun 18, 2026Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 2.10 MG/DL COMPARED TO READINGS OF 7.80 MG/DL RESPECTIVELY OBTAINED ON A METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO THE C ZONE, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS 11 DAYS. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25548497 ↗MalfunctionJun 12, 2026Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 400 MG/DL COMPARED TO READINGS OF 170 MG/DL RESPECTIVELY OBTAINED ON A METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO THE C ZONE, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS UNKNOWN DAYS. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25483298 ↗MalfunctionJun 9, 2026Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 63 MG/DL COMPARED TO READINGS OF 171 MG/DL RESPECTIVELY OBTAINED ON A METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO THE C ZONE, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS 7 DAYS. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25443398 ↗MalfunctionJun 3, 2026Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 2.60 MG/DL COMPARED TO READINGS OF 9.00 MG/DL RESPECTIVELY OBTAINED ON A METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO THE C ZONE, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS 6 DAYS. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25375902 ↗MalfunctionJun 2, 2026Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 39MG/DL, 54MG/DL, 59MG/DL COMPARED TO READINGS OF 136MG/DL, 115MG/DL, 120MG/DL RESPECTIVELY OBTAINED ON A METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO C ZONE, SHOWING DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS 6 DAYS. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25358745 ↗MalfunctionMay 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 380 MG/DL COMPARED TO READINGS OF 85 MG/DL RESPECTIVELY OBTAINED ON A METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO THE C ZONE, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS 5 DAYS. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25333360 ↗MalfunctionMay 26, 2026Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 168, 215, 414 MG/DL COMPARED TO READINGS OF 71, 89, 73 MG/DL RESPECTIVELY OBTAINED ON A METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO THE C ZONE, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS 5 DAYS. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25277523 ↗MalfunctionMay 21, 2026Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 62 MG/DL, 54 MG/DL, AND 53 MG/DL COMPARED TO READINGS OF 165 MG/DL, 148 MG/DL, AND 186 MG/DL RESPECTIVELY OBTAINED ON A METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO THE C ZONE, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS 1 DAY. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25241504 ↗MalfunctionMay 21, 2026Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 53 MG/DL COMPARED TO READINGS OF 156 MG/DL RESPECTIVELY OBTAINED ON A METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO THE C ZONE, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS 1 DAY. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25239998 ↗MalfunctionApr 23, 2026Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 3.7 MMOL/L , 3.1 MMOL/L AND 3.2 MMOL/L COMPARED TO READINGS OF 8.20 MMOL/L , 6.4 MMOL/L AND 12.4 MMOL/L RESPECTIVELY OBTAINED ON A METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO THE C ZONE, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS 5 DAYS. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24964132 ↗MalfunctionApr 21, 2026Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 5.40 MMOL/L COMPARED TO READINGS OF 23.40 MMOL/L; RESPECTIVELY OBTAINED ON A METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO THE D ZONE, SHOWING DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS 5 DAYS. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24935850 ↗MalfunctionApr 20, 2026Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 53 MG/DL AND 53 MG/DL COMPARED TO READINGS OF 186 MG/DL AND 180 MG/DL RESPECTIVELY OBTAINED ON A METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO C ZONE, SHOWING DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS 1 DAY. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24926787 ↗MalfunctionApr 17, 2026Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 400 MG/DL COMPARED TO READINGS OF 127 MG/DL RESPECTIVELY OBTAINED ON A METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO THE C ZONE, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS 5 DAYS. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24915916 ↗MalfunctionJan 9, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16120316 ↗MalfunctionAug 11, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15208987 ↗MalfunctionApr 12, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED RECEIVING A LOW READING FROM THEIR ADC DEVICE. CUSTOMER REPORTED A LOW READING OF 40 MG/DL WHICH WAS LOWER THAN LAB READING OF 200 MG/DL. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14084146 ↗InjuryMar 8, 2022Device Displays Incorrect Message; Device Displays Incorrect Message
A CALLER REPORTED THAT THE CUSTOMER OBTAINED AN "INCOMPATIBLE SENSOR" MESSAGE UPON SCANNING THE ADC DEVICE. THE CUSTOMER FELT HYPOGLYCEMIC AND WAS TRANSFERRED TO THE HOSPITAL WHERE HE RECEIVED UNSPECIFIED TREATMENT. NO ADDITIONAL INFORMATION WAS PROVIDED. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: Hypoglycemia
MAUDE report 13689366 ↗MalfunctionJan 12, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13237765 ↗MalfunctionDec 17, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13024315 ↗MalfunctionOct 18, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC FREESTYLE LIBRE SENSOR. CUSTOMER REPORTED RECEIVING READINGS OF 350 MG/DL, 120 MG/DL AND 40 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12652206 ↗InjuryOct 11, 2021Failure to Fire; Failure to Fire
THE WIFE OF CUSTOMER REPORTED THAT THE ADC FREESTYLE LIBRE SENSOR "DID NOT APPLY" AND CUSTOMER WAS UNABLE TO OBTAIN SCAN READINGS. THE CUSTOMER EXPERIENCED A LOSS OF CONSCIOUSNESS, AN AMBULANCE WAS CALLED, AND HE WAS TREATED WITH FAST ACTING INSULIN. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: Hyperglycemia; Loss of consciousness
MAUDE report 12609035 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P160030. One submission often covers many models, so check the product and lot details in each record.
- Z-1273-2023 · Feb 13, 2023 · Open, ClassifiedFreeStyle Libre Reader, REF: 71525-01, 71701-01, 71739-01 a component of the FreeStyle Libre Flash Glucose Monitoring System
Other Sensor, glucose, invasive, non-adjunctive, factory-calibrated, user-initiated devices
- FreeStyle Libre 14 day 71940Abbott Diabetes Care Inc · Sensor, glucose, invasive, non-adjunctive, factory-calibrated, user-initiated1 death16,002 reports
- FreeStyle Libre 14 day 71938Abbott Diabetes Care Inc · Sensor, Glucose, Invasive, Non-Adjunctive, Factory-Calibrated, User-Initiated1 recall578 reports
- FreeStyle Libre 14 day 71937Abbott Diabetes Care Inc · Sensor, Glucose, Invasive, Non-Adjunctive, Factory-Calibrated, User-Initiated1 recall82 reports
- FreeStyle LibreLink 71732-01Abbott Diabetes Care Inc · Sensor, glucose, invasive, non-adjunctive, factory-calibrated, user-initiated111 reports
- FreeStyle LibreLink 71733-01Abbott Diabetes Care Inc · Sensor, glucose, invasive, non-adjunctive, factory-calibrated, user-initiated92 reports
- FreeStyle Libre 14 day 71939Abbott Diabetes Care Inc · Sensor, glucose, invasive, non-adjunctive, factory-calibrated, user-initiated78 reports
- FreeStyle Libre 71525Abbott Diabetes Care Inc · Sensor, glucose, invasive, non-adjunctive, factory-calibrated, user-initiated1 recall18 reports
- FreeStyle Libre 14 day 71936Abbott Diabetes Care Inc · Sensor, glucose, invasive, non-adjunctive, factory-calibrated, user-initiated1 recall9 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.