Skip to content

FreeStyle Libre 14 day 71937Recalls & reported problems

Abbott Diabetes Care Inc · FreeStyle Libre 14 day Reader Kit US · Class 3

All models of FreeStyle Libre 14 day (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Abbott FreeStyle Libre 14 day Reader is a handheld device that displays glucose readings from a FreeStyle Libre sensor. There are 82 adverse event reports describing it, with no deaths and 33 injuries; reports mostly mention the reader becoming too hot to hold while charging, failing to power on, or being unable to get glucose readings. One recall, open since February 2023, covers lithium-ion batteries in the reader that may swell, overheat, or pose a fire hazard.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
82
Deaths reported
0
Injuries reported
33
Recalls naming it
1

Summary

The FDA's MAUDE database contains 82 adverse event reports that list this exact device (DI 00357599000974), received between Apr 18, 2019 and Feb 21, 2025 — classified as 33 injuries, 49 malfunctions.

The most frequently reported specific device problem is “Product Quality Problem” (45 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 13, 2023Open, Classified

    Lithium-ion batteries in glucose monitoring system readers may swell, overheat, or pose fire hazard. Only use the charging cable/power adapter included with readers. Do not use reader if it is damaged, cracked, swelling, too hot to hold, it will no longer hold charge, or it fails Reader Test. Do not use other chargers. User Manual update includes instructions on reader safe storage, charging, use.

    Z-1274-2023 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Product Quality Problem45
  • Failure to Power Up20
  • Unable to Obtain Readings12
  • Incorrect, Inadequate or Imprecise Result or Readings3
  • Battery Problem2
  • Device Displays Incorrect Message2
  • Appropriate Device Problem Term/Code Not Available1
  • Display or Visual Feedback Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions45
  • Hypoglycemia12
  • Hyperglycemia11
  • Loss of consciousness8
  • Dizziness5
  • Elevated Ketones/ Diabetic Ketoacidosis4
  • No Known Impact Or Consequence To Patient4
  • Diaphoresis3

Recent adverse event reports

The 25 most recent of 82 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 21, 2025Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD WHILE CHARGING. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21440291 ↗
  • InjurySep 6, 2024Failure to Power Up; Failure to Power Up

    A BATTERY/NO POWER ISSUE WAS REPORTED WITH THE ABBOTT DIABETES CARE (ADC) DEVICE. THE CUSTOMER WAS UNABLE TO TEST DUE TO THE DEVICE NOT POWERING ON WITH BUTTON PRESS OR TEST STRIP INSERTION. THE CUSTOMER EXPERIENCED WEAKNESS, "COULD NOT STAND AND KEPT SITTING DOWN", AND WAS UNABLE TO SELF-TREAT. THE CUSTOMER WAS ADMITTED IN THE HOSPITAL AND RECEIVED UNSPECIFIED TREATMENT. THERE WAS NO REPORT OF DEATH OR PERMANENT IMPAIRMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: Fatigue; Hyperglycemia

    MAUDE report 20163918 ↗
  • MalfunctionApr 18, 2024Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ABBOTT DIABETES DEVICE IS TOO HOT TO HOLD WHILE CHARGING. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. CUSTOMER REPORTS USING THE CABLE AND ADAPTER SUPPLIED BY ABBOTT DIABETES CARE FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19139793 ↗
  • MalfunctionOct 12, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17917737 ↗
  • MalfunctionSep 29, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-2023. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB2023.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17842788 ↗
  • MalfunctionSep 20, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17779474 ↗
  • MalfunctionSep 14, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17743088 ↗
  • MalfunctionAug 31, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17663481 ↗
  • InjuryAug 16, 2023Unable to Obtain Readings; Unable to Obtain Readings

    A UNSPECIFIED ISSUE WAS REPORTED WITH THE ADC DEVICE. CUSTOMER WAS UNABLE TO OBTAIN READINGS DUE TO THE DEVICE "WAS LOST DUE TO A CAR ACCIDENT". THE CUSTOMER REPORTED UNSPECIFIED TREATMENT FROM THIRD-PARTY DUE TO THIS ISSUE. THERE WAS NO REPORT OF DEATH OR PERMANENT IMPAIRMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 17555767 ↗
  • MalfunctionAug 7, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17468372 ↗
  • MalfunctionAug 1, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17432279 ↗
  • MalfunctionJul 31, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17423040 ↗
  • MalfunctionJul 21, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17369001 ↗
  • MalfunctionJul 11, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17299018 ↗
  • MalfunctionJul 10, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17285689 ↗
  • MalfunctionJul 7, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17276599 ↗
  • MalfunctionJul 5, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17259425 ↗
  • MalfunctionJul 5, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17254885 ↗
  • MalfunctionJul 3, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17246150 ↗
  • MalfunctionJun 30, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER (B)(4). SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION (B)(4). WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17235992 ↗
  • MalfunctionJun 20, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17162795 ↗
  • MalfunctionJun 16, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17143281 ↗
  • MalfunctionJun 13, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17116439 ↗
  • MalfunctionJun 13, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17115930 ↗
  • MalfunctionMay 30, 2023Product Quality Problem; Product Quality Problem

    CUSTOMER REPORTS THEIR ADC DEVICE IS TOO HOT TO HOLD. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME. IT IS UNKNOWN IF THE CUSTOMER'S DEVICE WAS TOO HOT TO HOLD DURING USE OR WHILE CHARGING. THERE WAS NO REPORT OF FIRE, SMOKE, EXPLOSION, OR SHOCK. IT IS UNKNOWN AT THIS TIME IF THE CHARGING CABLE AND ADAPTER USED WAS THE CABLE AND ADAPTER SUPPLIED BY ADC FOR USE WITH THE LIBRE DEVICE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT. ISSUES INVOLVING THERMAL EVENTS OCCURRING WITH ADC FREESTYLE LIBRE AND LIBRE 2 READERS IS ASSOCIATED WITH REPORT OF CORRECTIONS AND REMOVAL NUMBER 2954323-02/09/23-001-C, SUBMITTED TO THE FDA ON 09-FEB-23. ADC FIELD ACTION FA1005-2023 WAS ISSUED TO THE US 09FEB23.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17019062 ↗

Other Sensor, Glucose, Invasive, Non-Adjunctive, Factory-Calibrated, User-Initiated devices

All Sensor, Glucose, Invasive, Non-Adjunctive, Factory-Calibrated, User-Initiated devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page