FreeStyle Precision Neo 71748Recalls & reported problems
Abbott Diabetes Care Inc · FreeStyle Precision Neo Blood Glucose Test Strips Not For Retail Sale · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00357599000332), received between Feb 3, 2020 and Aug 6, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings2
Patient problems
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 6, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 39 MG/DL AND 389 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 12289904 ↗MalfunctionFeb 3, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF "LO" (LESS THAN 20 MG/DL), AND 113 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9662369 ↗
Other SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER devices
- OneTouch Verio Flex 023-271LifeScan Europe GmbH · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER3,703 reports
- FreeStyle Libre 2 71953Abbott Diabetes Care Inc · System, test, blood glucose, over the counter2 recalls1 death3,286 reports
- OneTouch Ultra2 021-105LifeScan U.S. LLC · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER1,731 reports
- Omnipod DASH PDM PDM-USA1-D001-MG-USA1Insulet Corporation · System, Test, Blood Glucose, Over The Counter1 recall10 deaths1,129 reports
- Omnipod UST400Insulet Corporation · System, Test, Blood Glucose, Over The Counter12 deaths816 reports
- TRUE METRIX TRUE METRIX SELF MONITORING BLOOD GLUCOSE METER - NOT FOR RETAIL SALETrividia Health, Inc. · System, Test, Blood Glucose, Over The Counter418 reports
- FreeStyle Freedom Lite 70923-70Abbott Diabetes Care Inc · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER378 reports
- FreeStyle Lite 70829-70Abbott Diabetes Care Inc · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER378 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.