LibreLinkUp 72154-01Recalls & reported problems
Abbott Diabetes Care Inc · LibreLinkUp · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00357599824006), received between Jun 19, 2024 and Nov 25, 2025 — classified as 4 injuries.
The most frequently reported specific device problem is “Application Program Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Application Program Problem4
- Device Alarm System1
Patient problems
- Hypoglycemia3
- Loss of consciousness3
- Convulsion/Seizure2
- Brain Injury1
- Confusion/ Disorientation1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryNov 25, 2025Application Program Problem; Application Program Problem
A COMPLAINT VIA EMAIL WAS RECEIVED. AN APPLICATION ISSUE WAS REPORTED WITH THE ABBOTT DIABETES CARE (ADC) DEVICE IN USE WITH AN UNKNOWN PHONE WITH AN IOS OPERATING SYSTEM VERSION. THE CUSTOMER INDICATED THAT THE LIBRELINKUP APPLICATION SEEMS TO NOT BE WORKING AT ALL ON IOS. IT APPEARS THAT THIS ISSUE STARTED RIGHT AFTER UPGRADING THE PHONE AND WATCH TO VERSION 26.1. AS A RESULT, THE CUSTOMER WAS UNABLE TO MONITOR GLUCOSE USING SENSOR READINGS AND EXPERIENCED A LOSS OF CONSCIOUSNESS. THERE WAS NO REPORT OF DEATH OR PERMANENT IMPAIRMENT ASSOCIATED WITH THIS EVENT.
Patient problems: Loss of consciousness
MAUDE report 23646770 ↗InjuryNov 15, 2024Application Program Problem; Application Program Problem
AN UNEXPECTED APPLICATION ISSUE WAS REPORTED WITH THE ABBOTT DIABETES CARE (ADC) APPLICATION, IN USE WITH IPHONE XR WITH IOS OPERATING SYSTEM VERSION 17.6.1. THE CUSTOMER'S FREESTYLE LIBRELINKUP APPLICATION WAS "NOT RECEIVING INFORMATION" FROM FREESTYLE LIBRE 3 APPLICATION. AS A RESULT, THE CAREGIVER WAS NOT NOTIFIED VIA THE FREESTYLE LIBRELINKUP APPLICATION OF CUSTOMER'S GLYCEMIC STATUS. THE CUSTOMER EXPERIENCED CONFUSION, CONVULSIONS, SEIZURE AND WAS UNABLE TO SELF-TREAT. THE CUSTOMER RECEIVED ORAL GLUCOSE FROM A FAMILY MEMBER AND JUICE FROM A HEALTHCARE PROFESSIONAL AS TREATMENT. THERE WAS NO REPORT OF DEATH OR PERMANENT IMPAIRMENT ASSOCIATED WITH THIS EVENT.
Patient problems: Confusion/ Disorientation; Convulsion/Seizure; Hypoglycemia
MAUDE report 20695895 ↗InjuryJun 25, 2024Device Alarm System; Application Program Problem; Device Alarm System; Application Program Problem
AN ALARM ISSUE WAS REPORTED WITH THE ABBOTT DIABETES CARE (ADC) DEVICE IN USE WITH IPHONE PHONE WITH IOS OPERATING SYSTEM VERSION 17.2.1. THE LOW AND HIGH GLUCOSE ALARMS DID NOT SOUND AND CUSTOMER WAS NOT ALERTED OF CHANGES IN GLUCOSE LEVEL. AS A RESULT, THE CUSTOMER EXPERIENCED "SCREAMING, SLEEPING", A SEIZURE AND A LOSS OF CONSCIOUSNESS. THE CUSTOMER RECEIVED TREATMENT OF HONEY, SUGAR AND GLUCAGON INJECTION PROVIDED BY A NON-HEALTHCARE PROFESSIONAL (CAREGIVER). THE CAREGIVER REPORTED CALLING EMERGENCY SERVICES BUT IT IS UNKNOWN IF THEY ADMINISTERED ANY ADDITIONAL MEDICAL TREATMENT. THERE WAS NO REPORT OF DEATH OR PERMANENT IMPAIRMENT ASSOCIATED WITH THIS EVENT.
Patient problems: Convulsion/Seizure; Hypoglycemia; Loss of consciousness
MAUDE report 19605595 ↗InjuryJun 19, 2024Application Program Problem; Application Program Problem
AN EMAIL COMPLAINT WAS RECEIVED IN WHICH AN UNEXPECTED APPLICATION ISSUE WAS REPORTED WITH THE ABBOTT DIABETES CARE (ADC) APPLICATION, IN USE WITH NOTHING PHONE 1 WITH ANDROID OPERATING SYSTEM VERSION 13. THE CUSTOMER'S FREESTYLE LIBRELINKUP APPLICATION WAS SHOWING "UNEXPECTED APPLICATION ERROR" MESSAGE. THE CAREGIVER WAS UNSURE WHETHER THE OTHER USER OF THE SAME ACCOUNT TURNED OFF THEIR NOTIFICATIONS OR ACKNOWLEDGED THE ALERT WITHOUT TAKING ACTION, AND IF THIS AFFECTED THE MAIN USER'S APPLICATION. AS A RESULT, THE CAREGIVER WAS NOT NOTIFIED VIA THE LINKUP APPLICATION OF CUSTOMER'S GLYCEMIC STATUS. THE CUSTOMER EXPERIENCED HYPOGLYCEMIA AND LOSS OF CONSCIOUSNESS. THIS LED THEM BEING TAKEN TO A HOSPITAL, WHERE THEY WERE DIAGNOSED WITH OBSTRUCTIVE SLEEP APNEA, REPORTEDLY RESULTED "BRAIN DAMAGE" AND RECEIVED UNSPECIFIED TREATMENT BY A HEALTHCARE PROFESSIONAL (HCP). HCP SUSPECTED THAT THIS MIGHT HAVE BEEN CAUSED BY THE HYPOGLYCEMIA, BUT THEY COULD NOT CONFIRM A DIRECT LINK. AT THE SAME TIME, THEY WERE UNABLE TO DETERMINE ANY OTHER SPECIFIC CAUSE FOR THE SLEEP APNEA AND "BRAIN DAMAGE". THERE WAS NO REPORT OF DEATH OR PERMANENT IMPAIRMENT ASSOCIATED WITH THIS EVENT.
Patient problems: Brain Injury; Hypoglycemia; Loss of consciousness
MAUDE report 19570435 ↗
FDA clearance and classification
The FDA regulates product code PJT (continuous glucose monitor secondary display) under 21 CFR 862.1350 as a Class II device. 21 CFR 862.1350 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
FDA definition of this device type: The purpose of the continuous glucose monitor secondary display is to notify another person, the follower, of the patient's continuous glucose monitoring system sensor glucose information in real time. The secondary display is intended for providing secondary notification of a continuous glucose monitoring system and does not replace primary real time continuous glucose monitoring or standard home blood glucose monitoring
Common questions about the LibreLinkUp 72154-01
Has the LibreLinkUp 72154-01 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00357599824006), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the LibreLinkUp 72154-01?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jun 19, 2024 and Nov 25, 2025. In those reports, the most common specific device problems are “Application Program Problem” (4) and “Device Alarm System” (1); the most common patient problems are “Hypoglycemia” (3), “Loss of consciousness” (3) and “Convulsion/Seizure” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the LibreLinkUp 72154-01?
None of the reports are classified as deaths, 4 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the LibreLinkUp 72154-01 FDA approved?
Devices of this type (product code PJT) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Abbott Diabetes Care Inc.
Who makes the LibreLinkUp 72154-01, and is it still sold?
The FDA's device database lists Abbott Diabetes Care Inc as the company responsible for it (brand LibreLinkUp, model 72154-01). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Continuous glucose monitor secondary display devices
- t:connect Mobile Application (iOS) 1004128Tandem Diabetes Care, Inc. · Continuous Glucose Monitor Secondary Display1 recall0 reports
- Dexcom Follow App for iOS SW10497Dexcom, Inc. · Continuous Glucose Monitor Secondary Display35 reports
- Dexcom Follow 5 Android Mobile SW14256Dexcom, Inc. · Continuous Glucose Monitor Secondary Display7 reports
- Dexcom Follow App for Android SW10583Dexcom, Inc. · Continuous Glucose Monitor Secondary Display6 reports
- t:connect Mobile Application, Mobile Bolus Enabled (iOS) 1007111Tandem Diabetes Care, Inc. · Continuous Glucose Monitor Secondary Display3 reports
- Dexcom Follow 5 iOS Mobile SW14255Dexcom, Inc. · Continuous Glucose Monitor Secondary Display2 reports
- t:connect Mobile Application (Android) 1004127Tandem Diabetes Care, Inc. · Continuous Glucose Monitor Secondary Display1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page