CoaguChek XS Patient Self Testing Care Kit 4837738001Recalls & reported problems

Roche Diagnostics GmbH · Class 2

All models of CoaguChek XS Patient Self Testing Care Kit (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Roche CoaguChek XS is a self-testing kit for blood clotting time. Many of the 136 reports describe display errors and hard-to-read screens, and some describe results that changed between tests. The reports include no deaths and 7 injuries, and reports rose from 5 in the previous 12 months to 13 in the latest 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
137
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 137 adverse event reports that list this exact device (DI 00365702125100), received between Aug 2, 2018 and Aug 28, 2026 — classified as 7 injuries, 130 malfunctions.

The most frequently reported specific device problem is “Image Display Error/Artifact” (62 reports).

4 other recalls cover devices cleared under the same FDA submission (K062925); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 3 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 3 reportsMay 2026: 1 reportsJun 2026: 3 reportsJul 2026: 2 reportsAug 2026: 3 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Image Display Error/Artifact62
  • Display Difficult to Read37
  • Non Reproducible Results16
  • Circuit Failure9
  • Device Difficult to Setup or Prepare8
  • High Test Results7
  • Incorrect, Inadequate or Imprecise Result or Readings6
  • Adverse Event Without Identified Device or Use Problem4

Patient problems

  • No Clinical Signs, Symptoms or Conditions121
  • No Consequences Or Impact To Patient7
  • Insufficient Information3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Coagulation Disorder1
  • Fall1
  • Gastrointestinal Hemorrhage1
  • Hematoma1

Recent adverse event reports

The 25 most recent of 137 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 28, 2026High Test Results; Non Reproducible Results

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50502046 ↗
  • MalfunctionAug 21, 2026Non Reproducible Results

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50373275 ↗
  • MalfunctionAug 19, 2026High Test Results; Non Reproducible Results

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50328115 ↗
  • MalfunctionJul 27, 2026High Test Results

    It was alleged that a patient received a questionable result when testing with CoaguChek XS meter serial number (b)(6).\\A sample from the patient was tested with the meter at an unknown time, resulting in a value of 2.1 INR. The patient was not feeling well prior to testing. \\At an unknown time on the same day, a sample from the patient was tested in the laboratory using an unknown method, resulting in a value of 1.06 INR. \\The patient's therapeutic range is 2.0 - 3.0 INR.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 50001112 ↗
  • MalfunctionJul 17, 2026Non Reproducible Results

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25857881 ↗
  • MalfunctionJun 22, 2026Low Test Results

    IT WAS ALLEGED THAT A PATIENT RECEIVED DISCREPANT RESULTS WHEN TESTING WITH COAGUCHEK XS METER SERIAL NUMBER (B)(6). AT 9:05 A.M., A SAMPLE FROM THE PATIENT WAS TESTED USING THE METER, RESULTING IN A VALUE OF 3.0 INR. AT 9:08 A.M., A SAMPLE FROM THE PATIENT WAS TESTED USING THE METER, RESULTING IN A VALUE OF 2.0 INR. THE PATIENT'S THERAPEUTIC RANGE IS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 25581168 ↗
  • MalfunctionJun 9, 2026Circuit Failure; Use of Device Problem

    WE RECEIVED AN ALLEGATION OF A COAGUCHEK XS METER SOFTWARE MALFUNCTION FOR ONE PATIENT. THE CUSTOMER INITIALLY CALLED DUE TO ERROR 3. DURING THE CALL, THE CUSTOMER ALLEGED THAT THE METER GAVE RESULTS USING EXPIRED TEST STRIPS. THE CUSTOMER REPORTED THE FOLLOWING METER RESULTS: ON (B)(6) 2026, THE METER RESULT WAS 2.6 INR. ON (B)(6) 2026, THE METER RESULT WAS 3.3 INR. ON (B)(6) 2026, THE METER RESULT WAS 3.9 INR. THE THERAPEUTIC RANGE IS 2-3 INR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25448812 ↗
  • MalfunctionJun 1, 2026Display Difficult to Read

    THERE WAS AN ALLEGATION OF A DISPLAY ISSUE WITH THE COAGUCHEK XS METER AFFECTING THE INTERPRETATION OF RESULTS. THE CUSTOMER ALLEGED THAT WHILE ATTEMPTING TO TEST, AN ERROR MESSAGE WAS DISPLAYED ON THE SCREEN, WHICH FLUCTUATED BETWEEN SHOWING " ERROR + 5" AND "ERROR + 9". WITH THE ASSISTANCE OF TECHNICAL SUPPORT, THE CUSTOMER PERFORMED A TEST, AND THE RESULT APPEARED TO BE 1.9 INR, BUT THEN THE RESULT CHANGED FROM 1.9 INR TO 1.4 INR. THE THERAPEUTIC RANGE WAS 2.0 INR - 4.0 INR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25347329 ↗
  • MalfunctionMay 11, 2026Image Display Error/Artifact; Lack of Maintenance Documentation or Guidelines

    THERE WAS AN ALLEGATION THAT THE COAGUCHEK XS PST CARE KIT GAVE A RESULT OF 24 INSTEAD OF 2.4 INR. IT IS UNCLEAR IF THERE IS AN ISSUE WITH THE DISPLAY OR IF THE METER IS SET IN THE INCORRECT UNIT OF MEASUREMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25135313 ↗
  • MalfunctionApr 16, 2026Non Reproducible Results

    WE RECEIVED AN ALLEGATION OF QUESTIONABLE INR RESULTS FOR ONE PATIENT TESTED WITH THE COAGUCHEK XS PST CARE KIT METER. THE METER RESULT AT 9:44 AM WAS 7.7 INR. THE METER RESULT AT 9:48 AM WAS 5.7 INR. THE PATIENT'S THERAPEUTIC RANGE IS 2.0 INR - 3.0 INR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24895575 ↗
  • MalfunctionApr 14, 2026Non Reproducible Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM THE COAGUCHEK XS METER. AT 3:13 P.M., THE METER RESULT WAS 2.4 INR. AT 3:16 P.M., THE METER RESULT WAS 1.9 INR. AT 4:57 P.M., THE METER RESULT WAS 3.4 INR. AT 5:42 P.M., THE METER RESULT WAS 1.1 INR. THE PATIENT¿S THERAPEUTIC RANGE IS 2.0 - 3.0 INR. THE PATIENT STOPPED TAKING THEIR WARFARIN MEDICATION FOR 2 DAYS BASED ON THE RESULT OF 1.1 INR AND CONFIRMED NO HARM OR SERIOUS INJURY OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24877030 ↗
  • MalfunctionApr 2, 2026Device Difficult to Setup or Prepare; Non Reproducible Results

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¿S THERAPEUTIC RANGE IS 2.0 INR - 3.0 INR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24765458 ↗
  • MalfunctionJan 12, 2026Non Reproducible Results

    WE RECEIVED AN ALLEGATION OF A QUESTIONABLE HIGH INR RESULT FOR ONE PATIENT TESTED WITH THE COAGUCHEK XS PST METER, WHEN COMPARED WITH ANOTHER COAGUCHEK METER. THE METER RESULT AT THE DOCTOR'S OFFICE WAS 2.5 INR, WHILE THE PATIENT'S METER RESULT WAS 5.0 INR. THE TIME DIFFERENCE BETWEEN THE MEASUREMENTS WAS WITHIN MINUTES. THE PATIENT¿S THERAPEUTIC RANGE WAS NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24036054 ↗
  • MalfunctionJan 8, 2026Adverse Event Without Identified Device or Use Problem

    THERE WAS AN ALLEGATION OF A QUESTIONABLE INR RESULT USING THE COAGUCHEK XS PT TEST 6 STRIP WITH THE COAGUCHEK XS PST CARE KIT FOR ONE PATIENT. THE INITIAL RESULT AT 6:41 PM WAS 4.3 INR, AND ANOTHER RESULT AT 6:55 PM WAS 5.6 INR. THE PATIENT WAS NOT SURE IF HER HANDS WERE COMPLETELY DRY WHEN DOING THE FIRST TEST.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 24009889 ↗
  • MalfunctionAug 14, 2025Image Display Error/Artifact; Image Display Error/Artifact

    THERE WAS AN ALLEGATION OF A DISPLAY ISSUE ON A COAGUCHEK XS METER. THE CUSTOMER STATED THAT THE METER DISPLAY IS FADED AND MISSING SEGMENTS. THEY ALSO ALLEGED THAT THEY RECEIVED A RESULT OF 1.2 INR, BUT THEY WERE NOT SURE AS THE RESULTS WERE FADED. THE CUSTOMER PERFORMED A DISPLAY CHECK AND CONFIRMED THAT THERE WERE MISSING SEGMENTS IN THE NUMBER 8 ON THE DISPLAY SCREEN. THE CUSTOMER ALSO CHANGED THE BATTERIES ON THE METER AND THE SCREEN WAS STILL FADED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22805872 ↗
  • MalfunctionJul 16, 2025Image Display Error/Artifact; Image Display Error/Artifact

    THERE WAS A METER DISPLAY ISSUE WITH COAGUCHEK XS METER, SERIAL NUMBER (B)(6). CUSTOMER ALLEGED THE METER HAD A DISPLAY MALFUNCTION THAT RESULTED IN A MISINTERPRETATION OF RESULTS. THE RESULTS FIELD OF THE METER IS FADED. THE RESULTS FIELD DISPLAYED A RESULT OF 5.4 INR ON THE DAY OF THE EVENT. CUSTOMER DID NOT BELIEVE THIS WAS THE CORRECT RESULT DUE TO THE DISPLAY ISSUE. THE CUSTOMER DID NOT REPORT THE RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22521917 ↗
  • MalfunctionFeb 13, 2025Circuit Failure; Failure to Power Up; Battery Problem; Device Contamination with Chemical or Other Material; Circuit Failure; Failure to Power Up; Battery Problem; Device Contamination with Chemical or Other Material

    THERE WAS AN ALLEGATION OF BURNING ON COAGUCHEK XS METER, SERIAL NUMBER (B)(6). CUSTOMER STATED THE METER WOULD NOT POWER ON AND ONE OF THE BATTERIES HAD CORRODED AND LEAKED IN THE BATTERY COMPARTMENT. THEY ATTEMPTED TO CHANGE THE BATTERIES AND NOTICED THERE WAS A BLACK/BURNT SPOT ON THE BATTERY DOOR AS WELL AS A MELTED SPOT ON THE BOTTOM OF THE BATTERY DOOR. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21375125 ↗
  • InjuryDec 4, 2024Adverse Event Without Identified Device or Use Problem; Non Reproducible Results; Adverse Event Without Identified Device or Use Problem; Non Reproducible Results

    THERE WAS AN ALLEGATION OF A PATIENT EXPERIENCING A FALL, WHICH RESULTED IN AN INFECTED HEMATOMA, AFTER ALLEGEDLY BEING UNABLE TO USE THE COAGUCHEK XS METER DUE TO ERROR MESSAGES. ON (B)(6) 2024 THE METER RESULT WAS REPORTEDLY 2.6 INR. ALLEGEDLY, THE PATIENT HAD BEEN UNABLE TO USE THE METER SINCE (B)(6) 2024 DUE TO RUNNING OUT OF TEST STRIPS AFTER REPEATEDLY RECEIVING ERROR MESSAGE 5 ON THE METER. PRODUCT LABELING INDICATES THAT ERROR 5 IS A BLOOD APPLICATION ERROR. PRODUCT LABELING STATES: "ERROR APPLYING BLOOD TO THE TEST STRIP OF BLOOD SAMPLE SIZE WAS TOO SMALL." THE PATIENT REPORTEDLY TOLD HER PHYSICIAN'S OFFICE SHE WAS OUT OF TEST STRIPS; SHE WAS ALLEGEDLY NOT ADVISED TO DO ANYTHING OR COME IN FOR TESTING. THE PATIENT DID NOT HAVE AN ALTERNATIVE METHOD OF TESTING AVAILABLE. THE PATIENT WAS REPORTEDLY IN THE PROCESS OF TRYING TO OBTAIN ADDITIONAL TESTING STRIPS WHEN SHE FELL ON (B)(6) 2024. ON (B)(6) 2024, THE PATIENT REPORTEDLY VISITED HER GENERAL PRACTITIONER DUE TO HAVING A HEMATOMA. THE HEMATOMA WAS INFECTED AND THE PATIENT WAS ALLEGEDLY ADVISED TO GO TO THE EMERGENCY ROOM. THE PATIENT'S RESULT IN THE EMERGENCY ROOM FROM AN UNKNOWN LABORATORY METHOD WAS ALLEGEDLY 7.0 INR. W

    Patient problems: Hematoma

    MAUDE report 20843130 ↗
  • InjuryNov 8, 2024No Device Output; High Test Results; Adverse Event Without Identified Device or Use Problem; No Device Output; High Test Results; Adverse Event Without Identified Device or Use Problem

    WE RECEIVED AN ALLEGATION OF UNSPECIFIED ERROR MESSAGES ON A COAGUCHEK ® XS SYSTEM THAT MAY HAVE CONTRIBUTED TO A PATIENT'S INTRACRANIAL HEMORRHAGE AND NEED FOR HOSPITALIZATION. THE REPORTER COULD NOT REMEMBER THE SPECIFIC ERROR MESSAGES THAT OCCURRED. THE PATIENT HAD REPORTEDLY BEEN UNABLE TO OBTAIN A RESULT ON THE METER BEFORE THE HOSPITALIZATION ON (B)(6) 2024 DUE TO UNSPECIFIED ERROR MESSAGES AFTER DOSING THE TEST STRIP. IT IS NOT KNOWN WHEN THE PATIENT LAST ATTEMPTED TO TEST ON THE DEVICE BEFORE THE EVENT. ERROR MESSAGES ARE DESIGNED FAIL SAFES TO PREVENT THE DEVICE FROM PRODUCING A RESULT WHEN CERTAIN OPERATING CONDITIONS ARE MET. CLARIFICATION ON THE SPECIFIC ERROR MESSAGES RECEIVED HAS BEEN REQUESTED BUT NOT PROVIDED. THE FOLLOWING RESULTS FROM (B)(6) 2024 WERE PROVIDED FROM THE MEMORY OF THE METER: ON (B)(6) 2024 THE RESULT WAS REPORTEDLY 6.1 INR. ON (B)(6) 2024 THE RESULT WAS REPORTEDLY 4.9 INR. ON (B)(6) 2024 THE RESULT WAS REPORTEDLY 1.2 INR. THE PATIENT¿S THERAPEUTIC RANGE IS REPORTEDLY 2.0 ¿ 3.0 INR WITH AN ALLEGED TESTING FREQUENCY OF ONCE PER WEEK. ON (B)(6) 2024 THE PATIENT ACCIDENTALLY BUMPED HIS HAND ON THE WALL. REPORTEDLY, THE BLEEDING ON THE PATIENT¿S HAND COU

    Patient problems: Coagulation Disorder; Intracranial Hemorrhage

    MAUDE report 20642536 ↗
  • MalfunctionJan 8, 2024Device Difficult to Setup or Prepare; Non Reproducible Results; Device Difficult to Setup or Prepare; Non Reproducible Results

    WE RECEIVED AN ALLEGATION OF AN ISSUE WITH A COAGUCHEK XS METER. THE PATIENT ALLEGED RUNNING A TEST ON THE METER AND RECEIVING AN ERROR 5 MESSAGE AND THEN A RESULT OF 1.6 INR WITH THE SAME TEST STRIP. AN ERROR 5 IS RELATED TO AN ISSUE APPLYING BLOOD TO THE TEST STRIP. THE PATIENT REPORTED THE RESULT OF 1.6 INR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18464259 ↗
  • MalfunctionDec 21, 2023Incorrect, Inadequate or Imprecise Result or Readings; High Test Results; Incorrect, Inadequate or Imprecise Result or Readings; High Test Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM THE COAGUCHEK XS METER. THE INITIAL RESULT WAS ALLEGEDLY 23.8 AND THE PATIENT WENT TO THE ER FOR TESTING. THE LABORATORY RESULT USING AN UNKNOWN REAGENT WAS 1.6 INR. IT WAS DETERMINED THE METER WAS SET TO DISPLAY RESULTS IN SECONDS AND NOT INR. THE RESULT FROM THE METER MEMORY AT 10:29 WAS 1.9 INR (22.8 SEC). THE RESULT FROM THE METER MEMORY AT 10:32 WAS 2.1 INR (23.8 SEC). THE RESULT FROM THE METER MEMORY AT 10:39 WAS 1.9 INR (22.4 SEC). A DIFFERENT FINGER WAS USED FOR EACH TEST. THE THERAPEUTIC RANGE WAS 2.0-3.0 INR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18374545 ↗
  • MalfunctionDec 20, 2023Image Display Error/Artifact; Misassembly by Users; Image Display Error/Artifact; Misassembly by Users

    WE RECEIVED AN ALLEGATION ABOUT A FADED DISPLAY WHEN A PATIENT TESTED WITH A COAGUCHEK XS METER. THE REPORTER ALLEGED THAT SINCE THE METER WAS FLICKERING, HE REPLACED THE BATTERIES. HE ALSO ALLEGED THAT THE METER DID NOT POWER ON AFTER THAT. THE REPORTER WAS ADVISED TO TURN OFF THE METER AND PRESS THE M BUTTON TO ACCESS THE METER MEMORY. THE METER WAS POWERED ON WITH AN ERROR MESSAGE. THE REPORTER ALLEGED THAT HE COULD NOT DETERMINE IF THE ERROR MESSAGE WAS AN ERROR 4 OR AN ERROR 9 AS THE NUMBERS WERE FADED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18365979 ↗
  • MalfunctionDec 19, 2023Calibration Problem; Non Reproducible Results; Calibration Problem; Non Reproducible Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM COAGUCHEK XS METER, SERIAL NUMBER (SN) (B)(6), COMPARED TO ANOTHER COAGUCHEK XS METER, SERIAL NUMBER (B)(6). AT 3:00 P.M., THE RESULT FROM METER SN (B)(6) WAS 3.5 INR. AT 3:05 P.M., THE RESULT FROM METER SN (B)(6) WAS 2.5 INR. THE PATIENT¿S THERAPEUTIC RANGE IS 3.0 - 4.0 INR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18356695 ↗
  • MalfunctionNov 30, 2023Non Reproducible Results; Non Reproducible Results

    THERE WAS AN ALLEGATION OF A METER DISPLAY ISSUE ON COAGUCHEK XS METER, SERIAL NUMBER UNKNOWN. THE PATIENT'S DAUGHTER ALLEGED THAT THE SCREEN APPEARS FOGGY AND CANNOT READ THE RESULTS. SHE ALLEGED THEY REPLACED THE BATTERIES BUT THE ISSUES PERSISTED AND THAT THERE ARE MISSING SEGMENTS IN THE RESULTS FIELD. SHE STATED THAT SHE PERFORMED A TEST ON 05-NOV-2023 AND IT APPEARED TO GIVE A RESULT OF 2.1 INR BUT WAS UNSURE AS THERE WERE SEGMENTS MISSING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18239197 ↗
  • MalfunctionNov 16, 2023Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare

    WE RECEIVED AN ALLEGATION OF "VASTLY DIFFERENT" RESULTS ON A COAGUCHEK XS METER. IT WAS DETERMINED THE PATIENT HAD CHANGED THE BATTERIES IN THE METER ON (B)(6) 2023 AND THE UNITS WERE NOT RESET TO INR; THE RESULTS WERE IN SECONDS (SEC) UNIT OF MEASURE. THE CALL AGENT ASSISTED WITH RESETTING THE METER TO READ IN INR UNIT OF MEASURE. DISCREPANT RESULTS IN INR FROM (B)(6) 2023 WERE IDENTIFIED. THE RESULT AT 12:52 P.M. USING STRIP LOT 68883822 (EXPIRATION 30-SEP-2024) WAS 4.7 INR. THE RESULT AT 3:54 P.M. WITH A DIFFERENT FINGER USING STRIP LOT 66283721 (EXPIRATION 30-JUN-2024) WAS 2.8 INR. THE PATIENT¿S THERAPEUTIC RANGE IS 2 -3 INR WITH A TESTING FREQUENCY OF EVERY 2 WEEKS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18148454 ↗

FDA clearance and classification

Cleared through 510(k) K062925 on Jan 29, 2007 (decision: Substantially Equivalent). Submitted by Roche Diagnostics Corp. as “Coaguchek Xs System”. The FDA review took 123 days. FDA record ↗

The FDA regulates product code GJS (test, time, prothrombin) under 21 CFR 864.7750 as a Class II device. 21 CFR 864.7750 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K062925. One submission often covers many models, so check the product and lot details in each record.

  • Z-0804-2021 · Dec 2, 2020 · Terminated
    CoaguChek XS PST Meters- IVD (Patient Self-Testing) measures blood-clotting time for people who are taking anticoagulation medications such as Coumadin or warfarin. Catalog 04837
  • Z-1581-2013 · May 16, 2013 · Terminated
    CoaguChek XS System (PST) The CoaguChek XS PT test strips are part of the CoaguChek XS System. The CoaguChek XS System measures blood-clotting time for people who are taking antic
  • Z-2003-2009 · Jul 22, 2009 · Terminated
    Roche CoaguChek XS System, CoaguChek XS PST Kit, Roche Diagnostics, Indianapolis, IN; Catalog/Model Number 04837738001. CoaguChek XS System and CoaguChek XS Plus System are poin
  • Z-1594-2009 · May 13, 2009 · Terminated
    Roche Coaguchex XS Systems for Patient Self-Testing, Roche Diagnostics, Indianapolis, IN; REF 04837738001 includes the Roche ACCU-CHEK Softclix Lancing Device, Distributed by Roche

Common questions about the CoaguChek XS Patient Self Testing Care Kit 4837738001

Has the CoaguChek XS Patient Self Testing Care Kit 4837738001 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00365702125100), based on the FDA recall file dated Nov 27, 2026. However, 4 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the CoaguChek XS Patient Self Testing Care Kit 4837738001?

The FDA's MAUDE database lists 137 adverse event reports for this device, received between Aug 2, 2018 and Aug 28, 2026. In those reports, the most common specific device problems are “Image Display Error/Artifact” (62), “Display Difficult to Read” (37) and “Non Reproducible Results” (16); the most common patient problems are “Coagulation Disorder” (1), “Fall” (1) and “Gastrointestinal Hemorrhage” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CoaguChek XS Patient Self Testing Care Kit 4837738001?

None of the reports are classified as deaths, 7 as injuries and 130 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the CoaguChek XS Patient Self Testing Care Kit 4837738001 FDA approved?

It was cleared through 510(k) K062925 on Jan 29, 2007. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the CoaguChek XS Patient Self Testing Care Kit 4837738001, and is it still sold?

The FDA's device database lists Roche Diagnostics GmbH as the company responsible for it (brand CoaguChek XS Patient Self Testing Care Kit, model 4837738001). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.