COAGUCHEK XS PT TEST 6 STRIP 4625374160Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This is the Roche CoaguChek XS PT test strip, used to measure blood clotting time. The 1,676 reports often describe results that read too high or too low, could not be repeated, or differ for the same patient. Some mention strokes. The records include 3 deaths, 109 injuries and 1,564 malfunctions, and no recalls name this device. Reports rose to 53 in the last 12 months from 36 in the 12 months before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,682
- Deaths reported
- 3
- Injuries reported
- 109
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,682 adverse event reports that list this exact device (DI 00365702127104), received between Jul 12, 2018 and Sep 29, 2026 — classified as 3 deaths, 109 injuries, 1,570 malfunctions.
The most frequently reported specific device problem is “High Test Results” (1,098 reports).
4 other recalls cover devices cleared under the same FDA submission (K062925); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results1,098
- Non Reproducible Results289
- Incorrect, Inadequate or Imprecise Result or Readings233
- Low Test Results103
- Adverse Event Without Identified Device or Use Problem43
- Device Difficult to Setup or Prepare36
- No Device Output13
- Lack of Maintenance Documentation or Guidelines10
Patient problems
- No Clinical Signs, Symptoms or Conditions1,176
- No Consequences Or Impact To Patient397
- Insufficient Information40
- Ischemia Stroke13
- Stroke/CVA11
- Transient Ischemic Attack11
- Coagulation Disorder9
- Thrombosis/Thrombus9
Recent adverse event reports
The 25 most recent of 1,682 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026High Test Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50937051 ↗MalfunctionSep 29, 2026Non Reproducible Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50932051 ↗MalfunctionSep 25, 2026No Device Output; Adverse Event Without Identified Device or Use Problem
There was an allegation of questionable INR results for 1 patient using the COAGUCHEK XS PT TEST 6 STRIP on a COAGUCHEK XS PST CARE KIT. \\The patient provided the following results:\1.2 INR, \1.0 INR ,\1.1 INR ,\1.3 INR ,\1.6 INR, \2.4 INR .\\The patient stated that the results were obtained before her hospitalization on (b)(6)2026. The specific dates and times of the results were requested, but not provided.\The patient tests on the meter every two days. \The therapeutic range is 2.0 to 3.0 INR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50898164 ↗MalfunctionSep 17, 2026Device Difficult to Setup or Prepare
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50806941 ↗MalfunctionSep 10, 2026Non Reproducible Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50720529 ↗MalfunctionSep 3, 2026Non Reproducible Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50592647 ↗MalfunctionSep 2, 2026Non Reproducible Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50571003 ↗MalfunctionAug 18, 2026Non Reproducible Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50298343 ↗MalfunctionAug 14, 2026Non Reproducible Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50260182 ↗MalfunctionAug 13, 2026Non Reproducible Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50233744 ↗MalfunctionAug 13, 2026High Test Results
Patient problems: Hematoma
MAUDE report 50233550 ↗MalfunctionAug 5, 2026High Test Results; Adverse Event Without Identified Device or Use Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50146347 ↗MalfunctionJul 13, 2026Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM THE COAGUCHEK XS METER.\\THE PATIENT REPORTED TESTING USING HIS OWN METER AT 12:38 AM, WITH A RESULT OF 6.1 INR.\\WITHIN 1.5 HOURS, HE RETESTED AT A CLINIC USING HIS OWN METER WITH CLINIC-PROVIDED STRIPS. THE RESULT WAS 2.6 INR.\\LATER THAT SAME DAY (WITHIN 4 HOURS), HE TESTED USING THE FACILITY'S METER AND STRIPS, AND THE RESULT WAS 2.7 INR. THE FACILITY'S TYPE OF METER WAS NOT KNOWN, BUT THE PATIENT STATED IT WAS IDENTICAL TO THE COAGUCHEK XS METER HE CURRENTLY OWNS.\\THE THERAPEUTIC RANGE WAS 2.5 INR TO 3.5 INR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25794818 ↗MalfunctionJun 26, 2026Device Difficult to Setup or Prepare
WE RECEIVED AN ALLEGATION OF QUESTIONABLE INR RESULTS FOR ONE PATIENT TESTED WITH THE COAGUCHEK XS PST CARE KIT. AT 2:41 PM, THE METER RESULT WAS 2.0 INR. AT 2:46 PM, THE METER RESULT WAS 2.9 INR. THE PATIENT'S THERAPEUTIC RANGE IS 2.0 TO 3.1 INR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25636542 ↗MalfunctionJun 15, 2026Device Difficult to Setup or Prepare; High Test Results; Adverse Event Without Identified Device or Use Problem
IT WAS ALLEGED THAT A PATIENT RECEIVED DISCREPANT RESULTS WHEN TESTING WITH COAGUCHEK XS METER SERIAL NUMBER (B)(6). SAMPLES FROM THE PATIENT WERE MEASURED ON THE METER AT UNKNOWN TIMES ON (B)(6) 2026, RESULTING IN VALUES OF 3.0 INR AND 6.2 INR. AT AN UNKNOWN TIME ON (B)(6) 2026, THE PATIENT HAD A RESULT OF 2.1 INR WHEN TESTING WITH AN UNKNOWN LABORATORY METHOD. THE PATIENT'S THERAPEUTIC RANGE IS 3.0 INR TO 3.5 INR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25503394 ↗InjuryJun 4, 2026High Test Results; Adverse Event Without Identified Device or Use Problem
IT WAS ALLEGED THAT A PATIENT HAD BLOOD CLOTS WHILE USING COAGUCHEK XS METER SERIAL NUMBER (B)(6). STARTING ON (B)(6) 2026 UNTIL (B)(6) 2026, IT WAS ALLEGED THAT THE PATIENT HAD 2 BLOOD CLOTS. THE PATIENT'S DOCTOR ALLEGEDLY TREATED THE PATIENT WITH AN ANTIBIOTIC, CIPROFLOXACIN. THE PATIENT'S THERAPEUTIC RANGE WAS REPORTED AS 2.5 - 3.5 INR. THIS MEDWATCH IS BEING SUBMITTED IN AN ABUNDANCE OF CAUTION.
Patient problems: Thrombosis/Thrombus
MAUDE report 25394384 ↗MalfunctionMay 27, 2026High Test Results
WE RECEIVED AN ALLEGATION OF DISCREPANT HIGH INR RESULTS WITH A COAGUCHEK VANTUS USA METER. THE RESULTS ON THE METER WERE 5.5 INR, 4.5 INR, AND 3.4 INR. THE RESULTS WERE OBTAINED ON THE SAME DAY; HOWEVER, THE SPECIFIC TIMEFRAME BETWEEN TESTS WAS NOT PROVIDED. THE PATIENT¿S THERAPEUTIC RANGE IS 2.0 ¿ 3.0 INR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25304206 ↗MalfunctionMay 22, 2026No Device Output; Non Reproducible Results
THERE WAS AN ALLEGATION OF A QUESTIONABLE INR RESULT USING THE COAGUCHEK XS PT TEST 6 STRIP WITH THE COAGUCHEK XS PST CARE KIT FOR 1 INDIVIDUAL. THE INITIAL RESULT WAS 1.6 INR. A SECOND RESULT IN LESS THAN A 4-HOUR WINDOW WAS 2.5 INR. THE CUSTOMER'S THERAPEUTIC INR RANGE IS 2.0 INR TO 3.0 INR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25257319 ↗InjuryMay 22, 2026Contamination; Adverse Event Without Identified Device or Use Problem
WE RECEIVED AN ALLEGATION THAT THE COAGUCHEK XS PST CARE KIT MAY HAVE CONTRIBUTED TO A PATIENT'S LUNG EMBOLISM. THE PATIENT WAS HAVING DIFFICULTY APPLYING THE BLOOD TO THE TEST STRIPS, AND SHE REPORTEDLY RECEIVED \"ERROR 5\" AND \"ERROR 000\". ON (B)(6) 2026, THE PATIENT WAS ALLEGEDLY HOSPITALIZED FOR A CLOT IN THE LUNG, AND IT WAS DISCOVERED THAT SHE HAD A LUNG EMBOLISM AND WAS GIVEN MEDICATION FOR THIS. THE PATIENT ALLEGED THAT THE METER GIVES HER ERRORS AND THAT THEY UTILIZE MORE THAN ONE TEST STRIP EVERY TIME THEY TRY TO DO THE INR TEST, SINCE THEY COME ACROSS THE ERROR ON THE METER, USUALLY ERROR 5; THEY ARE NOT SURE OF THE OTHER SPECIFIC ERRORS. THERE WERE REPORTEDLY NO CHANGES IN WARFARIN, MEDICATION, OR DIETARY CHANGES BEFORE THE HOSPITALIZATION, AND SHE HAD NO ILLNESS OR SICKNESS BEFORE THE HOSPITALIZATION. BEFORE THE EVENT, THE PATIENT'S PREVIOUS RESULTS WERE REPORTEDLY MOSTLY WITHIN HER THERAPEUTIC RANGE. THE PATIENT'S THERAPEUTIC RANGE IS REPORTEDLY 2 INR - 3 INR. FURTHER CLARIFICATION REGARDING THE FOLLOWING HAS BEEN REQUESTED BUT NOT PROVIDED: IF THE PATIENT ATTEMPTED TO TEST ON THE COAGUCHEK METER BETWEEN (B)(6) 2026 THE PATIENT'S INR IN THE HOSPITAL. IF THE PATIENT
Patient problems: Pulmonary Embolism
MAUDE report 25256310 ↗MalfunctionMay 1, 2026Low Test Results; Non Reproducible Results
IT WAS REPORTED THAT A PATIENT RECEIVED QUESTIONABLE RESULTS WHEN TESTING WITH COAGUCHEK XS METER SERIAL NUMBER (B)(6). PLEASE REFER TO THE ATTACHMENT FOR ALL PATIENT DATA. CLARIFICATIONS HAVE BEEN REQUESTED FOR ALL INFORMATION WRITTEN IN RED.
Patient problems: Insufficient Information
MAUDE report 25056411 ↗MalfunctionApr 30, 2026High Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM A COAGUCHEK XS METER. AT 9:59 AM, THE METER RESULT WAS 2.3 INR. AT 1:57 PM, THE METER RESULT WAS 3.2 INR. THE PATIENT¿S THERAPEUTIC RANGE IS 2.4 INR-2.6INR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25047129 ↗MalfunctionApr 28, 2026High Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM THE COAGUCHEK XS METER. THE INITIAL RESULT WAS 3.9 INR. THE REPEAT RESULT WAS 2.9 INR. THE TIMEFRAME BETWEEN THE RESULTS WAS NOT KNOWN. THE THERAPEUTIC RANGE WAS 2-3 INR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25014937 ↗MalfunctionApr 23, 2026Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM COAGUCHEK XS METER, SERIAL NUMBER (B)(6). THE METER RESULT WAS 5.5 INR. WITHIN 5 MINUTES, THE METER RESULT WAS 7.7 INR. THE LABORATORY RESULT USING AN UNKNOWN METHOD WAS 2.8 INR. THE EXACT TIME OF THE LABORATORY TEST WAS NOT PROVIDED. THE PATIENT¿S THERAPEUTIC RANGE WAS NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24968871 ↗MalfunctionApr 23, 2026Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM COAGUCHEK VANTUS METER, SERIAL NUMBER: (B)(6). AT 8:54 AM THE METER RESULT WAS 3.6 INR. AT 9:10 AM THE METER RESULT WAS 6.0 INR. AT 10:03 AM THE METER RESULT WAS 1.3 INR. THE CUSTOMER'S THERAPEUTIC RANGE IS 2.0-3.0 INR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24968859 ↗MalfunctionApr 20, 2026Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM A COAGUCHEK VANTUS METER. AT 3:58 PM, THE METER RESULT WAS 6.0 INR. AT 4:00 PM, THE METER RESULT WAS 4.2 INR. THE PATIENT¿S THERAPEUTIC RANGE IS 2.0-3.0 INR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24932053 ↗
FDA clearance and classification
Cleared through 510(k) K062925 on Jan 29, 2007 (decision: Substantially Equivalent). Submitted by Roche Diagnostics Corp. as “Coaguchek Xs System”. The FDA review took 123 days. FDA record ↗
The FDA regulates product code GJS (test, time, prothrombin) under 21 CFR 864.7750 as a Class II device. 21 CFR 864.7750 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K062925. One submission often covers many models, so check the product and lot details in each record.
- Z-0804-2021 · Dec 2, 2020 · TerminatedCoaguChek XS PST Meters- IVD (Patient Self-Testing) measures blood-clotting time for people who are taking anticoagulation medications such as Coumadin or warfarin. Catalog 04837
- Z-1581-2013 · May 16, 2013 · TerminatedCoaguChek XS System (PST) The CoaguChek XS PT test strips are part of the CoaguChek XS System. The CoaguChek XS System measures blood-clotting time for people who are taking antic
- Z-2003-2009 · Jul 22, 2009 · TerminatedRoche CoaguChek XS System, CoaguChek XS PST Kit, Roche Diagnostics, Indianapolis, IN; Catalog/Model Number 04837738001. CoaguChek XS System and CoaguChek XS Plus System are poin
- Z-1594-2009 · May 13, 2009 · TerminatedRoche Coaguchex XS Systems for Patient Self-Testing, Roche Diagnostics, Indianapolis, IN; REF 04837738001 includes the Roche ACCU-CHEK Softclix Lancing Device, Distributed by Roche
Common questions about the COAGUCHEK XS PT TEST 6 STRIP 4625374160
Has the COAGUCHEK XS PT TEST 6 STRIP 4625374160 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00365702127104), based on the FDA recall file dated Nov 27, 2026. However, 4 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the COAGUCHEK XS PT TEST 6 STRIP 4625374160?
The FDA's MAUDE database lists 1,682 adverse event reports for this device, received between Jul 12, 2018 and Sep 29, 2026. In those reports, the most common specific device problems are “High Test Results” (1,098), “Non Reproducible Results” (289) and “Incorrect, Inadequate or Imprecise Result or Readings” (233); the most common patient problems are “Ischemia Stroke” (13), “Stroke/CVA” (11) and “Transient Ischemic Attack” (11). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the COAGUCHEK XS PT TEST 6 STRIP 4625374160?
3 reports are classified as deaths, 109 as injuries and 1,570 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the COAGUCHEK XS PT TEST 6 STRIP 4625374160 FDA approved?
It was cleared through 510(k) K062925 on Jan 29, 2007. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the COAGUCHEK XS PT TEST 6 STRIP 4625374160, and is it still sold?
The FDA's device database lists Roche Diagnostics GmbH as the company responsible for it (brand COAGUCHEK XS PT TEST 6 STRIP, model 4625374160). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other TEST, TIME, PROTHROMBIN devices
- CoaguChek XS PT strips 4625315160Roche Diagnostics GmbH · TEST, TIME, PROTHROMBIN604 reports
- CoaguChek XS Patient Self Testing Care Kit 03666794001Roche Diagnostics GmbH · Test, Time, Prothrombin212 reports
- CoaguChek XS Patient Self Testing Care Kit 4837738001Roche Diagnostics GmbH · TEST, TIME, PROTHROMBIN137 reports
- i-STAT 03P89-24Abbott Point Of Care Inc. · TEST, TIME, PROTHROMBIN127 reports
- HemosIL ReadiPlasTin (20 mL) 00020301400Instrumentation Laboratory Company · TEST, TIME, PROTHROMBIN2 recalls2 reports
- Coag-sense 100218Coagusense Inc. · Test, Time, Prothrombin1 recall0 reports
- Coag-Sense Test Strip Kit, 50 ct. 100218Coagusense Inc. · Test, Time, Prothrombin1 recall0 reports
- STA® - Néoplastine® CI 10 00666Diagnostica Stago · TEST, TIME, PROTHROMBIN1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page