ACCU-CHEK AVIVA PLUS 06908217001Recalls & reported problems

Roche Diagnostics Operations, Inc. · ACCU-CHEK AVIVA PLUS 50CT STRIP BRK 36/C · Class 2

All models of ACCU-CHEK AVIVA PLUS (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Roche Accu-Chek Aviva Plus blood glucose test strip is used to measure blood sugar. Its 335 reports mostly describe glucose readings that are wrong or inconsistent, with some patients getting very different results in quick succession. The reports mention 11 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
335
Deaths reported
0
Injuries reported
11
Recalls naming it
0

Summary

The FDA's MAUDE database contains 335 adverse event reports that list this exact device (DI 00365702407107), received between Sep 5, 2019 and Jul 23, 2025 — classified as 11 injuries, 324 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (317 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 3 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings317
  • Imprecise or Erratic Results11
  • High Test Results3
  • Adverse Event Without Identified Device or Use Problem2
  • Low Readings2

Patient problems

  • No Consequences Or Impact To Patient130
  • No Clinical Signs, Symptoms or Conditions125
  • No Known Impact Or Consequence To Patient45
  • Hypoglycemia23
  • Hyperglycemia6
  • Shaking/Tremors4
  • Coma2
  • Dizziness2

Recent adverse event reports

The 25 most recent of 335 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 23, 2025High Test Results; High Test Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 414 MG/DL AND 179 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22575172 ↗
  • MalfunctionMay 19, 2025Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 64 MG/DL, 280 MG/DL AND 364 MG/DL.

    Patient problems: Numbness

    MAUDE report 22044402 ↗
  • MalfunctionMay 6, 2025Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 300 MG/DL AND 40 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21970817 ↗
  • MalfunctionMar 27, 2025Imprecise or Erratic Results; Imprecise or Erratic Results

    ON (B)(6) 2025 THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 454 MG/DL, 121 MG/DL, AND 82 MG/DL. ON (B)(6) 2025, THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 450 MG/DL, 120 MG/DL, AND 80 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21710161 ↗
  • MalfunctionMar 18, 2025Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 70'S (MG/DL) AND 180'S (MG/DL).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21635568 ↗
  • MalfunctionFeb 25, 2025High Test Results; High Test Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 311 MG/DL (SYSTEM 1) AND 153 MG/DL (SYSTEM 2).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21460343 ↗
  • MalfunctionFeb 11, 2025Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 243 MG/DL AND A RESULT IN THE 100 MG/DL RANGE. AT ANOTHER TIME, THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 256 MG/DL AND A RESULT IN THE 100 MG/DL RANGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21356333 ↗
  • MalfunctionJan 15, 2025High Test Results; High Test Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 328 MG/DL AND 160 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21158416 ↗
  • MalfunctionNov 27, 2024Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 322 MG/DL, 158 MG/DL, 137 MG/DL, 83 MG/DL, AND 126 MG/DL. THE CUSTOMER WAS FEELING LOW BLOOD SUGAR SYMPTOMS AND WAS ABLE TO SELF-TREAT.

    Patient problems: Shaking/Tremors

    MAUDE report 20797207 ↗
  • MalfunctionOct 15, 2024Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 120 MG/DL AND 60 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20448959 ↗
  • MalfunctionSep 16, 2024Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 91 MG/DL, 180 MG/DL AND 93 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20226514 ↗
  • MalfunctionSep 16, 2024Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 162 MG/DL AND 86 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20226473 ↗
  • MalfunctionAug 28, 2024Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 455 MG/DL AND 149 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20084199 ↗
  • MalfunctionJul 23, 2024Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING SETS OF RESULTS WITHIN 15 MINUTES: SYSTEM 1: AT 9:58AM A READING OF 357 MG/DL. AT 9:59AM A READING OF 226 MG/DL. AT 10:02AM A READING OF 135 MG/DL. AT 10:34AM A READING OF 354 MG/DL. AT 10:35AM A READING OF 203 MG/DL. AT 10:36 AM A READING OF 164 MG/DL. SYSTEM 2: AT 10:50AM A READING OF 355 MG/DL. AT 10:59AM A READING OF 202 MG/DL. AT 11:00AM A READING OF 117 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19812002 ↗
  • MalfunctionJul 3, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 688 MG/DL AND 180 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19665550 ↗
  • MalfunctionApr 15, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 114 MG/DL AND 305 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19112262 ↗
  • MalfunctionApr 15, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 123 MG/DL AND 262 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19110925 ↗
  • MalfunctionMar 18, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 230 MG/DL AND 115 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18927026 ↗
  • MalfunctionMar 13, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 33 MG/DL (SYSTEM 1) AND 98 MG/DL (SYSTEM 2).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18893757 ↗
  • MalfunctionMar 13, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 33 MG/DL (SYSTEM 1) AND 98 MG/DL (SYSTEM 2).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18893749 ↗
  • MalfunctionFeb 19, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 331 MG/DL, 83 MG/DL, AND 321 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18730607 ↗
  • MalfunctionDec 21, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 97 MG/DL AND 402 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18377709 ↗
  • MalfunctionNov 30, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 230 MG/DL, 98 MG/DL, AND 98 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18237569 ↗
  • MalfunctionNov 30, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 40 MG/DL (SYSTEM 1) AND 300 MG/DL (SYSTEM 2).

    Patient problems: Hypoglycemia; Shaking/Tremors

    MAUDE report 18237046 ↗
  • MalfunctionNov 29, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 70 MG/DL (PATIENT´S METER) AND 21 MG/DL (PROFESSIONAL´S METER).

    Patient problems: Hypoglycemia

    MAUDE report 18232755 ↗

Other Glucose Dehydrogenase, Glucose devices

All Glucose Dehydrogenase, Glucose devices →