ACCU-CHEK INFORM II 05942861001Recalls & reported problems
Roche Diagnostics Operations, Inc. · ACCU-CHEK INFORM II US 50CT 36/CS F2 · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Roche Accu-Chek Inform II glucose test is used to measure blood glucose. Of its 352 reports, 346 describe malfunctions and 4 describe injuries, while 2 mention deaths. Many describe results that read too high or too low, results that were imprecise, or readings that could not be repeated. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 354
- Deaths reported
- 2
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 354 adverse event reports that list this exact device (DI 00365702428102), received between Jul 13, 2018 and Sep 23, 2026 — classified as 2 deaths, 4 injuries, 348 malfunctions.
The most frequently reported specific device problem is “High Test Results” (126 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results126
- Incorrect, Inadequate or Imprecise Result or Readings102
- Non Reproducible Results96
- Low Test Results59
- Circuit Failure2
- Low Readings2
- No Apparent Adverse Event2
- Use of Device Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions328
- No Consequences Or Impact To Patient21
- Hypoglycemia5
- Hyperglycemia1
- Loss of consciousness1
Recent adverse event reports
The 25 most recent of 354 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026Non Reproducible Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50870602 ↗MalfunctionSep 23, 2026Non Reproducible Results
An allegation of questionable blood glucose results was received for a single patient tested consecutively on an Accu-Chek Inform II meter + RF meter. Testing performed within 15 minutes yielded 323 mg/dL and 145 mg/dL. The lower result was deemed correct.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50865753 ↗MalfunctionSep 22, 2026Non Reproducible Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50859945 ↗MalfunctionAug 19, 2026Non Reproducible Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50327395 ↗MalfunctionApr 20, 2026Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE GLUCOSE RESULTS FROM THE ACCUCHEK INFORM II METER. AT 11:57 A.M., THE METER RESULT WAS 206 MG/DL. AT 11:58 A.M., THE METER RESULT WAS 144 MG/DL. AT 11:59 A.M., THE METER RESULT WAS 117 MG/DL. THE INITIAL RESULT WAS DEEMED CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24932055 ↗MalfunctionApr 2, 2026Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE GLUCOSE RESULTS FOR TWO PATIENTS TESTED WITH ACCU-CHEK INFORM II METER SERIAL NUMBER: (B)(6). PATIENT 1: ON (B)(6) 2026, AT 10:18, THE METER RESULT WAS 521 MG/DL. AT 10:20, THE METER RESULT WAS 285 MG/DL. THIS RESULT WAS BELIEVED TO BE CORRECT. AT 10:21, THE METER RESULT WAS 277 MG/DL. PATIENT 2: ON (B)(6) 2026, AT 23:32, THE METER RESULT WAS "HI" (REPRESENTED AS 601 MG/DL). AT 23:34, THE METER RESULT WAS 149 MG/DL. THIS RESULT WAS BELIEVED TO BE CORRECT. ON (B)(6) 2026, AT 4:32, THE LABORATORY RESULT WAS 117 MG/DL. THE CUSTOMER DID NOT KNOW IF DIFFERENT FINGERS WERE USED TO COLLECT SAMPLES FOR EACH TEST. THE CUSTOMER HAS A POLICY OF REPEATING TESTING IF THE METER RESULT IS > 400 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24764372 ↗MalfunctionApr 1, 2026Non Reproducible Results
THE INITIAL REPORTER COMPLAINED OF DISCREPANT GLUCOSE RESULTS FOR 1 PATIENT TESTED ON AN ACCU-CHEK INFORM II METER. THE INITIAL RESULT FROM THE METER WAS 383 MG/DL. THE PATIENT WAS RETESTED WITHIN 15 MINUTES, AND THE RESULT WAS 61 MG/DL. THE RESULT OF 61 MG/DL WAS BELIEVED TO BE CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24757831 ↗MalfunctionMar 30, 2026Non Reproducible Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT GLUCOSE RESULTS FOR ONE PATIENT TESTED WITH ACCU-CHEK INFORM II METER SERIAL NUMBER (B)(6). A SAMPLE FROM THE PATIENT WAS TESTED USING THE METER, RESULTING IN A GLUCOSE VALUE OF 90 MG/DL. THE FACILITY HAS A POLICY TO REPEAT TESTING IF THE OPERATOR DOES NOT BELIEVE THE RESULT OR IF THE RESULT IS > 400 MG/DL. METER TESTING WAS REPEATED ON THE PATIENT TWICE, RESULTING IN GLUCOSE VALUES OF 450 MG/DL AND 469 MG/DL. ALL MEASUREMENTS WERE PERFORMED WITHIN 15 MINUTES. THE CUSTOMER DEEMED THE REPEAT VALUE OF 450 MG/DL TO BE CORRECT AND THE PATIENT WAS TREATED BASED ON THIS RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24731169 ↗MalfunctionMar 30, 2026Non Reproducible Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT GLUCOSE RESULTS FOR TWO PATIENTS TESTED WITH ACCU-CHEK INFORM II METER SERIAL NUMBER (B)(6). THE CUSTOMER HAS A POLICY OF REPEATING TESTING IF THE METER RESULT IS > 400 MG/DL. A SAMPLE FROM THE FIRST PATIENT WAS TESTED USING THE METER, RESULTING IN A GLUCOSE VALUE OF "HI" (> 600 MG/DL). THE TEST WAS REPEATED ON THE PATIENT USING THE METER, RESULTING IN A VALUE OF 218 MG/DL. ANOTHER MEASUREMENT OF THE PATIENT WAS PERFORMED USING THE PATIENT'S DEXCOM CONTINUOUS GLUCOSE MONITOR AND THE RESULT WAS 216 MG/DL. ALL MEASUREMENTS WERE PERFORMED WITHIN 15 MINUTES. THE SECOND RESULT OF 218 MG/DL WAS BELIEVED TO BE CORRECT AND THE PATIENT WAS TREATED BASED ON THAT RESULT. THE CUSTOMER DID NOT KNOW IF DIFFERENT FINGERS WERE USED TO COLLECT SAMPLES FOR EACH TEST. A SAMPLE FROM THE SECOND PATIENT WAS TESTED USING THE METER, RESULTING IN A GLUCOSE VALUE OF 461 MG/DL. TESTING WAS REPEATED ON THE PATIENT AND THE METER RESULT WAS 253 MG/DL. ALL MEASUREMENTS WERE PERFORMED WITHIN 15 MINUTES. THE SECOND RESULT OF 253 MG/DL WAS BELIEVED TO BE CORRECT AND THE PATIENT WAS TREATED BASED ON THAT RESULT. THE CUSTOMER DID NOT KNOW IF DIFFERENT FINGERS WERE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24731163 ↗MalfunctionMar 30, 2026Non Reproducible Results
THE INITIAL REPORTER COMPLAINED OF DISCREPANT GLUCOSE RESULTS FOR 1 PATIENT TESTED ON AN ACCU-CHEK INFORM II METER. THE RESULT AT 12:26 P.M. WAS 442 MG/DL. THE RESULT AT 12:29 P.M. WAS 171 MG/DL. THE RESULT AT 12:31 P.M. WAS 173 MG/DL. THE PATIENT WAS ASYMPTOMATIC. THE RESULTS OF 171 MG/DL AND 173 MG/DL WERE BELIEVED TO BE CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24716470 ↗MalfunctionMar 30, 2026Non Reproducible Results
THE INITIAL REPORTER COMPLAINED OF A DISCREPANT HIGH GLUCOSE RESULT FOR 1 PATIENT TESTED ON AN ACCU-CHEK INFORM II METER. THE RESULT AT 12:12 P.M. WAS HI (RESULT > 600 MG/DL). THE RESULT AT 12:14 P.M. WAS 226 MG/DL. THE PATIENT WAS ASYMPTOMATIC. THE RESULT OF 226 MG/DL WAS BELIEVED TO BE CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24716146 ↗MalfunctionMar 19, 2026Non Reproducible Results
WE RECEIVED AN ALLEGATION OF A QUESTIONABLE GLUCOSE RESULT FROM ONE PATIENT TESTED WITH TWO ACCU-CHEK INFORM II METER + RF METERS (METER 1 AND METER 2). AT 8:44 AM, METER 1'S RESULT WAS 513 MG/DL. AT 8:57 AM, METER 2'S RESULT WAS 187 MG/DL. METER 2'S RESULT WAS DEEMED CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24641215 ↗MalfunctionMar 13, 2026Incorrect, Inadequate or Imprecise Result or Readings; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM TWO ACCU-CHEK INFORM II METERS WITH SERIAL NUMBERS (B)(6). AT 5:16 A.M., METER (B)(6) GAVE A RESULT OF 224 MG/DL. AT 5:20 A.M., METER (B)(6) GAVE A RESULT OF 124 MG/DL. AT 4:11 P.M., METER (B)(6) GAVE A RESULT OF >600 MG/DL. AT 4:20 P.M., METER (B)(6) GAVE A RESULT OF 274 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24598360 ↗MalfunctionMar 6, 2026High Test Results
THE INITIAL REPORTER COMPLAINED OF DISCREPANT HIGH RESULTS FOR 1 PATIENT TESTED ON AN ACCU-CHEK INFORM II METER + RF COMPARED TO THE LABORATORY. AT 7:48 A.M., THE RESULT FROM THE METER FROM A FINGER STICK WAS > 600 MG/DL. AT 7:50 A.M., THE RESULT FROM THE METER USING A DIFFERENT FINGER STICK WAS > 600 MG/DL. A SAMPLE WAS DRAWN FOR THE LABORATORY AT 8:03 A.M. WITH A RESULT OF 83 MG/DL. THE RESULT FROM THE LABORATORY WAS BELIEVED TO BE CORRECT. THE PATIENT WAS ASYMPTOMATIC.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24529585 ↗MalfunctionFeb 18, 2026Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE GLUCOSE RESULTS AND ERROR MESSAGES USING THE ACCU-CHEK INFORM II TEST STRIPS WITH THE ACCU-CHEK INFORM II METER + RF. RESULTS WERE PROVIDED FOR ONE INDIVIDUAL. THE CUSTOMER TESTED THEMSELVES TO TRY TO RECREATE THE ISSUE. THE INITIAL RESULT AT 18:55 WAS 152 MG/DL. THE SECOND RESULT AT 18:56 WAS 111 MG/DL. THE SAME FINGER AND FINGERSTICK FROM THE FIRST RESULT WERE USED FOR THE SECOND RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24376240 ↗MalfunctionJan 29, 2026Low Test Results
THERE WAS AN ALLEGATION OF A QUESTIONABLE GLUCOSE RESULT USING THE ACCU-CHEK INFORM II TEST STRIPS FROM TWO ACCU-CHEK INFORM II METERS + RF FOR ONE PATIENT. THE INITIAL RESULT FROM METER A AT 12:58 WAS 67 MG/DL. THE SECOND RESULT FROM METER A AT 12:59 WAS 60 MG/DL. THE THIRD RESULT FROM METER A AT 14:01 WAS 27 MG/DL. THE FOURTH RESULT FROM METER A AT 14:03 WAS "LO", STATING THAT THE RESULT WAS LESS THAN 20 MG/DL. A RESULT FROM METER B AT 14:09 WAS 108 MG/DL. A SECOND RESULT FROM METER B AT 14:10 WAS 120 MG/DL. A VENOUS SAMPLE WAS COLLECTED AT 14:50, AND THE RESULT WAS 142 MG/DL. A SECOND VENOUS SAMPLE WAS COLLECTED AT 15:13, AND THE RESULT WAS 132 MG/DL. THE PATIENT WAS GIVEN D50 AS A RESULT OF THE INITIAL QUESTIONABLE RESULT OF 67 MG/DL. IT WAS VERIFIED THAT THE PATIENT WAS NOT ADVERSELY AFFECTED BY THE TREATMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24210550 ↗MalfunctionJan 28, 2026High Test Results
THERE WAS AN ALLEGATION OF A QUESTIONABLE GLUCOSE RESULT USING THE ACCU-CHEK INFORM II TEST STRIPS WITH TWO OF THE ACCU-CHEK INFORM II METERS + RF FOR ONE PATIENT. THE INITIAL RESULT WITH METER A AT 6:05 AM WAS 464 MG/DL. THE RESULT WITH METER B AT 6:15 AM WAS "HI", WHICH IS EQUIVALENT TO A RESULT THAT MAY BE GREATER THAN 600 MG/DL. A VENOUS SAMPLE WAS COLLECTED AT 6:29 AM, AND THE RESULT WAS 123 MG/DL. THE RESULT OF 123 MG/DL WAS BELIEVED TO BE CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24201446 ↗MalfunctionJan 22, 2026Non Reproducible Results
THE REPORTER COMPLAINED OF DISCREPANT GLUCOSE RESULTS FOR 1 PATIENT TESTED ON AN ACCU-CHEK INFORM II METER (METER 1) COMPARED TO A DIFFERENT ACCU-CHEK INFORM II METER (METER 2). THE PATIENT WAS INITIALLY TESTED WITH METER 1. AT 8:21 PM, METER 1 RESULT WAS 496 MG/DL, AND AT 8:23 PM, METER 1 RESULT WAS 234 MG/DL, WHILE AT 8:28 PM, METER 2 RESULT WAS 187 MG/DL. THE METER 2 RESULT OF 187 MG/DL WAS BELIEVED TO BE CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24143371 ↗MalfunctionJan 20, 2026High Test Results
THERE WAS AN ALLEGATION OF A QUESTIONABLE GLUCOSE RESULT USING THE ACCU-CHEK INFORM II TEST STRIPS WITH TWO OF THE ACCU-CHEK INFORM II METERS + RF FOR ONE PATIENT. THE INITIAL RESULT AT 11:42 AM WITH METER A WAS 225 MG/DL. THE SECOND RESULT AT 11:44 AM WITH METER A WAS 151 MG/DL. THE THIRD RESULT AT 11:50 AM WITH METER B WAS 202 MG/DL. THE RESULTS OF 225 MG/DL AND 202 MG/DL WERE CONSIDERED TO BE CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24115496 ↗MalfunctionJan 2, 2026Non Reproducible Results
THE INITIAL REPORTER COMPLAINED OF DISCREPANT GLUCOSE RESULTS FOR 1 PATIENT TESTED ON 2 ACCU-CHEK INFORM II METERS WITH SERIAL NUMBERS (B)(6) (METER A) AND (B)(6) (METER B). METER A WAS ALSO DISCREPANT WITH THE LABORATORY. AT 10:04 P.M., THE RESULT FROM METER A WAS 35 MG/DL. THE RESULT AT 10:10 P.M. WAS 37 MG/DL. AT 10:19 P.M., THE RESULT FROM THE LABORATORY WAS 121 MG/DL. AT 10:40 P.M., THE RESULT FROM METER A WAS 59 MG/DL. AT 10:46 P.M., THE RESULT FROM METER B WAS 83 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23942638 ↗MalfunctionDec 29, 2025Non Reproducible Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM ACCUCHEK INFORM II METER SERIAL NUMBER (B)(6). AT 5:24 A.M., THE METER RESULT WAS 186 MG/DL. AT 7:17 A.M., THE METER RESULT WAS 148 MG/DL. AT 7:54 A.M., THE METER RESULT WAS 426 MG/DL. AT 7:56 A.M., THE METER RESULT WAS HI (>600 MG/DL). AT 8:16 A.M., THE METER RESULT WAS 198 MG/DL USING A NEW VIAL OF TESTING STRIPS. THE RESULT OF 198 MG/DL WAS DEEMED CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23909692 ↗MalfunctionDec 23, 2025High Test Results; High Test Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT GLUCOSE RESULTS FOR ONE PATIENT TESTED WITH ACCU-CHEK INFORM II METER SERIAL NUMBER (B)(6). AT 6:52 A.M., A SAMPLE FROM THE PATIENT WAS TESTED ON THE METER, RESULTING IN A GLUCOSE VALUE OF 597 MG/DL. AT 6:54 A.M., A SAMPLE FROM THE PATIENT WAS TESTED ON THE METER, RESULTING IN A GLUCOSE VALUE OF 116 MG/DL. AT 6:55 A.M., A SAMPLE FROM THE PATIENT WAS TESTED ON THE METER, RESULTING IN A GLUCOSE VALUE OF 108 MG/DL. THE CUSTOMER WAS NOT SURE IF THE SAME FINGERSTICK WAS USED TO COLLECT SAMPLE FOR EACH MEASUREMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23881976 ↗MalfunctionDec 22, 2025High Test Results; High Test Results
THE INITIAL REPORTER COMPLAINED OF A HIGH GLUCOSE RESULT NOT CORRESPONDING TO THE CLINICAL STATUS OF 1 PATIENT TESTED ON AN ACCU-CHEK INFORM II METER + RF. THE RESULT ON THE METER WAS "HI" (NUMERIC RESULT > 600 MG/DL). THIS RESULT DID NOT FIT THE PATIENT¿S HEALTH STATUS, AS THEY WERE ASYMPTOMATIC.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23865716 ↗MalfunctionDec 22, 2025High Test Results; High Test Results
THERE WAS AN ALLEGATION OF A QUESTIONABLE GLUCOSE RESULT USING THE ACCU-CHEK INFORM II TEST STRIPS WITH THE ACCU-CHEK INFORM II METER + RF FOR ONE PATIENT. THE INITIAL RESULT AT 12:05 AM WAS 365 MG/DL. THE RESULT AT 12:07 AM ON AN UNKNOWN METER WAS 174 MG/DL. THE RESULT OF 174 MG/DL IS BELIEVED TO BE CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23863145 ↗MalfunctionDec 19, 2025No Device Output; High Test Results; No Device Output; High Test Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT GLUCOSE RESULTS FOR ONE PATIENT TESTED WITH ACCUCHEK INFORM II METER SERIAL NUMBERS (B)(6). AT 15:18, A SAMPLE FROM THE PATIENT WAS TESTED WITH METER SERIAL NUMBER (B)(6), RESULTING IN A GLUCOSE VALUE OF 46 MG/DL. AT 15:20, A SAMPLE FROM THE PATIENT WAS TESTED WITH METER SERIAL NUMBER (B)(6), RESULTING IN A GLUCOSE VALUE OF 90 MG/DL. AT 15:24, A SAMPLE FROM THE PATIENT WAS TESTED WITH METER SERIAL NUMBER (B)(6), RESULTING IN A GLUCOSE VALUE OF 56 MG/DL. AT 15:24, A SAMPLE FROM THE PATIENT WAS TESTED WITH METER SERIAL NUMBER (B)(6), RESULTING IN A GLUCOSE VALUE OF 101 MG/DL. THE CUSTOMER BELIEVED THE VALUE OF 101 MG/DL TO BE CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23850945 ↗
Other Glucose Dehydrogenase, Glucose devices
- OneTouch Verio IQ 022-275LifeScan Europe GmbH · Glucose dehydrogenase, glucose6,035 reports
- OneTouch Verio 022-659LifeScan Europe GmbH · Glucose dehydrogenase, glucose4,219 reports
- TRUE METRIX TRUE METRIX SELF MONITORING BLOOD GLUCOSE TEST STRIPS - 50 Test StripsTrividia Health, Inc. · Glucose Dehydrogenase, Glucose2 deaths2,548 reports
- FreeStyle Lite 70805Abbott Diabetes Care Inc · Glucose dehydrogenase, glucose2,366 reports
- FreeStyle Freedom Lite 70920Abbott Diabetes Care Inc · Glucose dehydrogenase, glucose1,037 reports
- FreeStyle Freedom Lite 70914Abbott Diabetes Care Inc · Glucose dehydrogenase, glucose760 reports
- Accu-Chek Guide 08256292001Roche Diagnostics Operations, Inc. · Glucose Dehydrogenase, Glucose476 reports
- ACCU-CHEK AVIVA PLUS 06908268001Roche Diagnostics Operations, Inc. · Glucose Dehydrogenase, Glucose472 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page