ACCU-CHEK AVIVA PLUS 06908349001Recalls & reported problems
Roche Diagnostics Operations, Inc. · ACCU-CHEK AVIVA PLUS 50 HN STP BRK 36/CS · Class 2
All models of ACCU-CHEK AVIVA PLUS (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Roche Accu-Chek Aviva Plus blood glucose test strip, a strip used with a glucose meter, has 337 reports, 5 of them injuries and no deaths. The reports mostly describe incorrect or imprecise glucose readings, including results that differed widely in quick succession, and some mention low or high blood sugar in patients. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 337
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 337 adverse event reports that list this exact device (DI 00365702438101), received between Sep 6, 2019 and Feb 7, 2025 — classified as 5 injuries, 332 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (330 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings330
- High Test Results4
- Device Displays Incorrect Message1
- Expiration Date Error1
- Incorrect Measurement1
Patient problems
- No Consequences Or Impact To Patient212
- No Known Impact Or Consequence To Patient74
- No Clinical Signs, Symptoms or Conditions29
- Hypoglycemia18
- Hyperglycemia3
- Loss of consciousness1
- Shaking/Tremors1
Recent adverse event reports
The 25 most recent of 337 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 7, 2025High Test Results; High Test Results
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 480 MG/DL AND 110 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21330992 ↗MalfunctionMay 21, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 487 MG/DL AND 50 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19363009 ↗MalfunctionNov 16, 2023Expiration Date Error; Expiration Date Error
THE CALLER REPORTED THAT THE TEST STRIPS LISTED AN EXPIRATION DATE OF 9/30/2024 ON THE VIAL. THE ALLEGED LABELING IS INCORRECT. THE EXPIRATION DATE OF STRIP LOT.498722 IS 09/30/2021.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18148606 ↗MalfunctionMay 22, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 477 MG/DL, 205 MG/DL, AND 266 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16972968 ↗MalfunctionFeb 6, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 479 MG/DL AND 210 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16305580 ↗MalfunctionJan 23, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 155 MG/DL AND 410 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16224516 ↗MalfunctionOct 5, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 235 MG/DL (AVIVA) AND 99 MG/DL (HOSPITAL METER).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15544921 ↗MalfunctionOct 4, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 261 MG/DL AND 127 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15536265 ↗MalfunctionJun 1, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 384 MG/DL AND 160 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14567669 ↗MalfunctionMay 6, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 486 MG/DL, 324 MG/DL, AND 195 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14308221 ↗MalfunctionApr 28, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CALLER REPORTEDLY RECEIVED THE FOLLOWING RESULTS ON AN AVIVA METER, COMPARED TO A PROFESSIONAL METER, WITHIN 15 MINUTES: 182 MG/DL (AVIVA) AND 89 MG/DL (DOCTOR'S METER). THE TEST STRIPS ARE CURRENTLY EXPIRED, HOWEVER THE CUSTOMER REPORTED THAT THE TEST STRIPS WERE NOT EXPIRED AT THE TIME OF THE COMPARISON.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14241115 ↗MalfunctionFeb 17, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: HI MG/DL (WHICH ON THE SYSTEM INDICATES A RESULT IN EXCESS OF 600 MG/DL) AND 148 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13547600 ↗MalfunctionOct 12, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 211 MG/DL AND 84 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12616624 ↗MalfunctionSep 28, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 44 MG/DL, 200 MG/DL, AND 212 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12540853 ↗MalfunctionSep 24, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: A RESULT IN THE 300'S MG/DL AND A RESULT OF EITHER 75 MG/DL OR 80 MG/DL. THE CUSTOMER COULD NOT RECALL THE EXACT VALUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12526119 ↗MalfunctionJun 29, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 264 MG/DL, 271 MG/DL, LO MG/DL, 287 MG/DL AND 80 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12085573 ↗MalfunctionJun 29, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 264 MG/DL, 271 MG/DL, LO MG/DL, 287 MG/DL AND 80 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12085529 ↗MalfunctionJun 28, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 257 MG/DL AND 281 MG/DL (SYSTEM 1) AND 115 MG/DL AND 131 MG/DL (SYSTEM 2).
Patient problems: Hypoglycemia; Shaking/Tremors
MAUDE report 12076007 ↗MalfunctionJun 21, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 356 MG/DL, 157 MG/DL AND 160 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12036389 ↗InjuryJun 21, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED A HIGH BLOOD GLUCOSE RESULT OF 200 MG/DL WHILE EXPERIENCING AN EPISODE OF HYPOGLYCEMIA. AROUND 4-5PM, THE PATIENT'S SYMPTOMS OF LOW BLOOD SUGAR WERE IMPAIRED SPEECH, NO CONTROL OVER HIS BODY MOVEMENTS, AND HE WAS INCOHERENT. THE PATIENT DID NOT TREAT DURING THIS TIME. BETWEEN 6-7PM, THE CUSTOMER BEGAN TO HAVE A SEIZURE. DUE TO THE IMPACT OF THE SEIZURE, THE CUSTOMER WAS NOT CAPABLE OF PERFORMING A BLOOD TEST ON HIS OWN. THEREFORE, HIS SON PERFORMED A BLOOD TEST FOR HIM AND RECEIVED A RESULT OF 200 MG/DL. THE CUSTOMER'S SON CALLED THE EMTS AROUND 7PM. WHEN THE EMTS ARRIVED THEY PERFORMED A BLOOD TEST ON THEIR METER AND RESULT WAS 50 MG/DL. BASED ON THE RESULT, THE PATIENT WAS TREATED WITH A GEL SUBSTANCE AND WAS TRANSPORTED TO THE MEDICAL CENTER FOR ADDITIONAL TREATMENT. THEY ARRIVED AT 9:59PM TO THE MEDICAL CENTER, WHERE THE PATIENT RECEIVED AN IV, BUT HE DOES NOT KNOW WHAT MEDICATION HE RECEIVED. THE CUSTOMER WAS NOT ADMITTED INTO THE MEDICAL CENTER, BUT WAS KEPT FOR SEVERAL HOURS FOR OBSERVATION. THE CUSTOMER WAS DISCHARGED ABOUT 2:45AM WITH A STEADY BLOOD GLUCOSE LEVEL AND ARRIVED BACK HOME ABOUT 3:00AM ON (B)(6) 2021. THE PATIENT WAS USING E
Patient problems: Hypoglycemia
MAUDE report 12035687 ↗MalfunctionJun 7, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 179 MG/DL, 193 MG/DL, 299 MG/DL, AND 76 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11948063 ↗MalfunctionMay 27, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 497 MG/DL AND 194 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11895239 ↗MalfunctionMay 20, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 316 MG/DL (SYSTEM 1) AND 142 MG/DL (SYSTEM 2).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11857037 ↗MalfunctionMay 10, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 243 M/GDL (AVIVA) AND 126 MG/DL (GUIDE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11797836 ↗MalfunctionMay 7, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES USING TWO BLOOD GLUCOSE MONITORS: 578 MG/DL AND A RESULT IN THE 140'S (MG/DL) WITH THE GUIDE METER (SYSTEM 1) AND 137 MG/DL WITH THE AVIVA METER (SYSTEM 2).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11792458 ↗
Other Glucose Dehydrogenase, Glucose devices
- OneTouch Verio IQ 022-275LifeScan Europe GmbH · Glucose dehydrogenase, glucose6,035 reports
- OneTouch Verio 022-659LifeScan Europe GmbH · Glucose dehydrogenase, glucose4,219 reports
- TRUE METRIX TRUE METRIX SELF MONITORING BLOOD GLUCOSE TEST STRIPS - 50 Test StripsTrividia Health, Inc. · Glucose Dehydrogenase, Glucose2 deaths2,548 reports
- FreeStyle Lite 70805Abbott Diabetes Care Inc · Glucose dehydrogenase, glucose2,366 reports
- FreeStyle Freedom Lite 70920Abbott Diabetes Care Inc · Glucose dehydrogenase, glucose1,037 reports
- FreeStyle Freedom Lite 70914Abbott Diabetes Care Inc · Glucose dehydrogenase, glucose760 reports
- Accu-Chek Guide 08256292001Roche Diagnostics Operations, Inc. · Glucose Dehydrogenase, Glucose476 reports
- ACCU-CHEK AVIVA PLUS 06908268001Roche Diagnostics Operations, Inc. · Glucose Dehydrogenase, Glucose472 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page