CONSTELLATION® HYPERVIT® TOTAL PLUS® 8065000096Recalls & reported problems
Alcon Laboratories, Inc. · 27+®Ga HYPERVIT® Bevel 20K CPM TOTAL PLUS® Combined Procedure Pak · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Alcon Constellation Hypervit is a cutter used in eye surgery. Its 556 reports mostly describe the cutter failing to cut or suck, and several say the product was swapped during surgery. The reports include no deaths and 11 injuries. There is 1 recall, open, for some lots damaged during manufacturing that may affect sterility. Reports fell from 226 in the earlier 12 months to 106 in the latest 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 554
- Deaths reported
- 0
- Injuries reported
- 11
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 554 adverse event reports that list this exact device (DI 00380650000965), received between Mar 9, 2021 and Sep 30, 2026 — classified as 11 injuries, 543 malfunctions.
The most frequently reported specific device problem is “Failure to Cut” (368 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Apr 28, 2025Open, Classified
Due to the potential that some units within specific lots were damaged during the manufacturing process. The area of damage is within the same area of the lidding. While the damage in the lidding is visibly noticeable, due to risk of sterility being compromised.
Z-2108-2025 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Failure to Cut368
- Suction Problem86
- Suction Failure57
- Break22
- Ambient Noise Problem18
- Inability to Irrigate17
- Material Twisted/Bent14
- Fluid/Blood Leak9
Patient problems
- No Clinical Signs, Symptoms or Conditions458
- Insufficient Information87
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
- Endophthalmitis2
- Eye Injury2
- Hemorrhage/Blood Loss/Bleeding2
- Intraocular Pressure Decreased2
- Retinal Tear2
Recent adverse event reports
The 25 most recent of 554 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Incomplete or Inadequate Connection
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949933 ↗MalfunctionSep 30, 2026Failure to Cut
A healthcare professional reported that the vitrector stopped cutting during vitrectomy surgery. In order to continue the operation, the doctor ordered to open another cassette, In total, there were two with the failure and one to be able to complete the surgery. There was no information about patient impact.
Patient problems: Insufficient Information
MAUDE report 50943065 ↗MalfunctionSep 29, 2026Failure to Cut
A physician reported actuation problem of the cutter occurred during surgery. The procedure type was unknown. The surgery was completed after replacing it with another product. There was no patient impact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50929610 ↗MalfunctionSep 29, 2026Suction Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50928546 ↗MalfunctionSep 25, 2026Use of Device Problem
A nurse reported that the cutter was defective at unknown timing of the surgery. The procedure type was unknown. The surgery completed ion same day, but the procedure completion details were unknown. There was no information about patient impact.
Patient problems: Insufficient Information
MAUDE report 50898508 ↗MalfunctionSep 23, 2026Failure to Cut
A physician reported that the cutting failure of cutter occurred during surgery during vitrectomy. The surgery was completed on same day, after replacing it with new one. There was a patient contact but no impact on the patient.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50864986 ↗MalfunctionAug 25, 2026Suction Problem
A physician reported that during surgery, aspiration failure of irrigation/ aspiration (I/A) occurred. The procedural type was unknown. The surgery was completed on the same day after replacing the product with another one. There was contact with patient but no impact to the patient.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50436801 ↗MalfunctionAug 20, 2026Failure to Cut
A physician reported that a cutting failure of the probe occurred during surgery. The surgery was completed on the same day after replacing the product with another one. The procedure type was unknown. There was patient contact with the suspect product, but no impact occurred to the patient.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50339859 ↗MalfunctionAug 10, 2026Suction Problem; Failure to Cut
A physician reported that the cutter did not aspirate and cut during surgery. The procedure type was unknown. The procedure was completed on same day after replacing the product with another one. There was no patient impact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50189566 ↗MalfunctionApr 1, 2026Failure to Cut
A PHYSICIAN REPORTED THAT CUTTER DID NOT ACTUATE DURING THE SURGERY. THE SURGERY TYPE WAS UNKNOWN. THE SURGERY WAS COMPLETED ON SAME DAY BY REPLACING THE PRODUCT. THE PRODUCT WAS NOT CONTACTED WITH PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24757217 ↗MalfunctionFeb 20, 2026Suction Problem; Failure to Cut
A NURSE REPORTED THAT CUTTING PROCEDURE WAS ONLY POSSIBLE WITH CONSTANT INTERRUPTIONS, ASPIRATION WAS LIMITED AND A VACUUM WAS PRESENT, BUT ONLY VERY SLOWLY AND WITH SIGNIFICANTLY REDUCED ASPIRATION POWER DURING THE VITRECTOMY SURGERY. THE DEFECTIVE VITRECTOMY PROBE WAS REPLACED, AND FIVE HYPERVIT BEVEL VITRECTOMY PROBES WERE USED UNTIL THE END OF THE FIRST OPERATION OF THE DAY. THE PATIENT EXPERIENCED IMPAIRMENT DUE TO THE EXTENDED SURGERY AND ANESTHESIA TIME. THE CURRENT PATIENT CONDITION WAS CAPTURED AS RESOLVED SINCE THE REPORTED IMPAIRMENT WAS REFERRING TO EXTENDED SURGERY AND ANESTHESIA TIME, NO ADDITIONAL IMPAIRMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24398981 ↗MalfunctionFeb 20, 2026Suction Problem; Failure to Cut
A HEALTHCARE PROFESSIONAL REPORTED THAT THE VITRECTOR STOPPED FUNCTIONING, AN ATTEMPT WAS MADE TO CHECK THE CUTTER, SIGNIFICANTLY REDUCED ASPIRATION WAS OBSERVED ALONG THE ENTIRE LENGTH OF THE TUBING, AS WELL AS SLOWED, IMPROPER CUTTER PERFORMANCE DURING VITRECTOMY SURGERY. THE VITRECTOR WAS REPLACED USING AN ADDITIONAL CASSETTE; HOWEVER, THE PROBLEM WITH PROPER CUTTER FUNCTION OCCURRED AGAIN. THE SURGERY WAS COMPLETED ON THE SAME DAY WITH A SECOND SURGICAL SET WAS OPENED, AND THE CASSETTE TOGETHER WITH THE CUTTER WAS REPLACED. THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24398461 ↗MalfunctionFeb 19, 2026Suction Problem
A HEALTHCARE PROFESSIONAL REPORTED THAT CUTTER REDUCED ASPIRATION WAS OBSERVED ALONG THE ENTIRE LENGTH OF THE TUBING AND SLOWED DURING THE VITRECTOMY SURGERY. THE CUTTER WAS REPLACED USING AN ADDITIONAL CASSETTE. HOWEVER, THE PROBLEM WITH PROPER CUTTER FUNCTION OCCURRED AGAIN. THE SURGERY WAS COMPLETED ON SAME DAY BY USING THE SECOND CASSETTE TOGETHER WITH THE CUTTER. THE PRODUCT WAS CONTACTED WITH PATIENT, BUT THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24386476 ↗MalfunctionFeb 16, 2026Failure to Cut
A CUSTOMER REPORTED THAT PACK FAILED THIS WEEK STOPPED CUTTING THE VITRECTOR, SO THEY DECIDED TO OPEN ANOTHER PACK. AFTER REPLACING IT, EVERYTHING WORKED PERFECTLY DURING SURGERY. THE PROCEDURE TYPE WAS UNKNOWN. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.
Patient problems: Insufficient Information
MAUDE report 24343411 ↗MalfunctionFeb 12, 2026Suction Problem; Failure to Cut
A HEALTH CARE PROFESSIONAL REPORTED THAT A SIGNIFICANT REDUCTION OF ASPIRATION WAS OBSERVED ALONG THE ENTIRE LENGTH OF THE TUBING, AS WELL AS SLOWED AND IMPROPER CUTTER PERFORMANCE DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS PATIENT CONTACT WITH THE PRODUCT, BUT NO IMPACT OCCURRED TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24319910 ↗MalfunctionFeb 9, 2026Failure to Cut
A HEALTHCARE PROFESSIONAL REPORTED THAT THE VITRECTOMY PROBE COULD NOT CUT DURING VITRECTOMY SURGERY. THE PROCEDURE WAS COMPLETED ON SAME DAY BY REPLACING THE PRODUCT WITH NEW ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24298253 ↗MalfunctionFeb 5, 2026Failure to Cut
A PHYSICIAN REPORTED THAT POOR ACTUATION OF CUTTER WAS CONFIRMED DURING THE SURGERY. THE PROCEDURAL TYPE WAS UNKNOWN. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24265974 ↗MalfunctionFeb 5, 2026Suction Problem; Failure to Cut
A HEALTHCARE PROFESSIONAL REPORTED THAT THE CUTTER ACTUATION WAS POOR, AND THE CUTTER COULD NOT ASPIRATE DURING SURGERY. THE PROCEDURE TYPE WAS UNKNOWN. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24265319 ↗MalfunctionJan 30, 2026Suction Problem
A PHYSICIAN REPORTED THAT ASPIRATION PROBLEM OF CUTTER OCCURRED DURING SURGERY. THE PROCEDURE TYPE WAS UNKNOWN. THE PROCEDURE WAS COMPLETED AFTER REPLACEMENT OF PRODUCT WITH NEW ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24219525 ↗MalfunctionJan 29, 2026Suction Problem
A PHYSICIAN REPORTED THAT THE CUTTER COULD NOT ASPIRATE DURING SURGERY. THE PROCEDURE TYPE WAS UNKNOWN. THE PROCEDURE WAS COMPLETED BY REPLACING THE PRODUCT WITH NEW ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24208224 ↗MalfunctionJan 14, 2026Infusion or Flow Problem
AN OPHTHALMIC SURGEON REPORTED THAT THE PRODUCT (CASSETTE) COULD NOT BE USED DUE TO IRRIGATION ERROR DURING CATARACT AND VITRECTOMY COMBINATION SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS PATIENT CONTACT WITH SUSPECT PRODUCT BUT NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24070032 ↗MalfunctionJan 14, 2026Failure to Cut
A CUSTOMER REPORTED THAT THE VITRECTOMY TIP FAILED DURING CUTTING OF THE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON SAME DAY AFTER REPLACING WITH A NEW TIP. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.
Patient problems: Insufficient Information
MAUDE report 24065487 ↗MalfunctionJan 12, 2026Failure to Cut
A PHYSICIAN REPORTED THAT THERE WAS ACTUATION ISSUE OF THE CUTTER DURING SURGERY. THE PROCEDURAL TYPE WAS UNKNOWN. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24034008 ↗MalfunctionJan 12, 2026Break
A PHYSICIAN REPORTED THAT NEEDLE OF CUTTER PROBE WAS FOUND BROKEN DURING CALIBRATION. THE SURGERY WAS PERFORMED AND COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.
Patient problems: Insufficient Information
MAUDE report 24033536 ↗MalfunctionJan 9, 2026Suction Problem
A PHYSICIAN REPORTED THAT A CUTTER COULD NOT ASPIRATE DURING CATARACT AND VITRECTOMY COMBINATION SURGERY. THE CONDITION OF ACTUATION AND CUTTING OF THE CUTTER WAS UNKNOWN. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24020303 ↗
Other Vitrectomy, instrument cutter devices
- Constellation, EDGE PLUS, TOTAL PLUS, ULTRAVIT 8065752450Alcon Laboratories, Inc. · Vitrectomy, instrument cutter2 recalls1,160 reports
- Constellation, EDGE PLUS, TOTAL PLUS, ULTRAVIT 8065752437Alcon Laboratories, Inc. · Vitrectomy, instrument cutter2 recalls1,023 reports
- CONSTELLATION® HYPERVIT® TOTAL PLUS® 8065000095Alcon Laboratories, Inc. · Vitrectomy, instrument cutter2 recalls616 reports
- CONSTELLATION® HYPERVIT® TOTAL PLUS® 8065753106Alcon Laboratories, Inc. · Vitrectomy, instrument cutter2 recalls546 reports
- Constellation, EDGE PLUS, TOTAL PLUS, ULTRAVIT 8065752451Alcon Laboratories, Inc. · Vitrectomy, instrument cutter2 recalls258 reports
- Constellation, EDGE PLUS, TOTAL PLUS, ULTRAVIT 8065752438Alcon Laboratories, Inc. · Vitrectomy, instrument cutter2 recalls203 reports
- CONSTELLATION® HYPERVIT® TOTAL PLUS® 8065753109Alcon Laboratories, Inc. · Vitrectomy, instrument cutter1 recall223 reports
- CONSTELLATION® HYPERVIT® TOTAL PLUS® 8065830077Alcon Laboratories, Inc. · Vitrectomy, instrument cutter2 recalls159 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.